三维定量乳腺超声分析用于鉴别小于1厘米的纤维腺瘤与癌
本研究纳入112个经活检的乳腺实性病灶(小于1cm,34个纤维腺瘤、78个浸润性导管癌),基于三维超声手动勾画三个感兴趣区并提取8个特征,用逻辑回归分类,受试者工作特征曲线下面积两位阅片者均为0.86,在100%敏感度下特异度分别为26%和50%,第二阅片者若以第一阅片者结果为训练集则特异度40%、良性病灶活检减少40%且敏感度不降低。
要点
- 三维定量乳腺超声对小于1cm的实性病灶,受试者工作特征曲线下面积两位阅片者均为0.86。
- 在100%敏感度下,特异度阅片者1为26%、阅片者2为50%。
- 第二阅片者以第一阅片者结果为训练集时,特异度40%,良性病灶活检减少40%且敏感度不降低。
核心结论
- 三维定量乳腺超声分析似乎是有前景的技术,可将可疑乳腺病灶分类为良性,可能避免不必要的活检。
研究设计与样本
- 纳入112个经活检的乳腺实性病灶(小于1cm,34个纤维腺瘤、78个浸润性导管癌)的三维超声研究,手动勾画病灶并定义三个感兴趣区,经自动增益校正后计算回声水平均值和标准差及残余衰减系数,共8个特征用逻辑回归分类,以留一法交叉验证评估分类准确性,构建受试者工作特征曲线评估性能,全部病灶由两位阅片者勾画并比较结果以评估手动勾画的影响。
诊断标准与关键数值
- 受试者工作特征曲线下面积两位阅片者均为0.86;在100%敏感度下,特异度阅片者1为26%、阅片者2为50%;第二阅片者以第一阅片者结果为训练集时,受试者工作特征曲线下面积为0.86,敏感度100%、特异度40%。
临床应用与阅片要点
- 作者认为三维定量乳腺超声分析是有前景的技术,可将可疑乳腺病灶分类为良性,可能避免不必要的活检。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: In (3D) ultrasound, accurate discrimination of small solid masses is difficult, resulting in a high frequency of biopsies for benign lesions. In this study, we investigate whether 3D quantitative breast ultrasound (3DQBUS) analysis can be used for improving non-invasive discrimination between benign and malignant lesions.
METHODS AND MATERIALS: 3D US studies of 112 biopsied solid breast lesions (size <1cm), were included (34 fibroadenomas and 78 invasive ductal carcinomas). The lesions were manually delineated and, based on sonographic criteria used by radiologists, 3 regions of interest were defined in 3D for analysis: ROI (ellipsoid covering the inside of the lesion), PER (peritumoural surrounding: 0.5mm around the lesion), and POS (posterior-tumoural acoustic phenomena: region below the lesion with the same size as delineated for the lesion). After automatic gain correction (AGC), the mean and standard deviation of the echo level within the regions were calculated. For the ROI and POS also the residual attenuation coefficient was estimated in decibel per cm [dB/cm]. The resulting eight features were used for classification of the lesions by a logistic regression analysis. The classification accuracy was evaluated by leave-one-out cross-validation. Receiver operating characteristic (ROC) curves were constructed to assess the performance of the classification. All lesions were delineated by two readers and results were compared to assess the effect of the manual delineation.
RESULTS: The area under the ROC curve was 0.86 for both readers. At 100% sensitivity, a specificity of 26% and 50% was achieved for reader 1 and 2, respectively. Inter-reader variability in lesion delineation was marginal and did not affect the accuracy of the technique. The area under the ROC curve of 0.86 was reached for the second reader when the results of the first reader were used as training set yielding a sensitivity of 100% and a specificity of 40%. Consequently, 3DQBUS would have achieved a 40% reduction in biopsies for benign lesions for reader 2, without a decrease in sensitivity.
CONCLUSION: This study shows that 3DQBUS is a promising technique to classify suspicious breast lesions as benign, potentially preventing unnecessary biopsies.
引用
- Meel-van den Abeelen AS, Weijers G, van Zelst JC, et al. 3D quantitative breast ultrasound analysis for differentiating fibroadenomas and carcinomas smaller than 1cm. Eur J Radiol. 2017. PMID: 28189199. doi:10.1016/j.ejrad.2017.01.006
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。