文献速递腹部泌尿生殖外周血管CT

肾脏CT血管成像:常规CT与最新宝石能谱成像联合不同噪声指数在图像质量和辐射剂量上的比较

CT angiography of the kidney using routine CT and the latest Gemstone Spectral Imaging combination of different noise indexes: image quality and radiation dose.
PubMed 文献 · Radiol Med · 2017-02-15

研究纳入120例行肾脏CTA的患者,分为常规CT组(A组)与GSI不同噪声指数组(B、C、D组),每组30例,均用ASiR 2.0重建。A组SNR、CNR和主观图像质量显著低于B、C、D组(P<0.01),B、C、D组间无显著差异;D组有效剂量较A组降低47.81%、较B组降低18.59%(P<0.01)。作者认为最新GSI联合不同噪声指数可在保持诊断图像质量的同时更有效降低辐射剂量。

要点

  • GSI联合不同噪声指数组的SNR、CNR和主观图像质量均显著优于常规CT组(P<0.01)。
  • B、C、D三组GSI方案之间SNR、CNR和主观图像质量无显著差异。
  • D组有效剂量较A组降低47.81%、较B组降低18.59%(P<0.01)。

核心结论

  • 最新GSI联合不同噪声指数值用于肾脏CTA时,较常规CT扫描模式能更有效降低辐射剂量,同时保持诊断图像质量。

研究设计与样本

  • 共120例接受肾脏CTA的患者分为A、B、C、D四组,每组30例;A组行常规CT检查,B、C、D组行GSI检查并使用不同噪声指数;所有图像采用ASiR 2.0重建;由两名独立放射科医师采用5分法进行主观图像评价;记录CTDIvol和DLP并计算均值,DLP转换为有效剂量(ED);所有数据采用单因素方差分析比较。

诊断标准与关键数值

  • 主观图像评价采用5分评分法,由两名独立放射科医师完成。
  • A组SNR、CNR和主观图像质量显著低于B、C、D组(P<0.01)。
  • B、C、D组间SNR、CNR和主观图像质量无显著差异。
  • D组ED较A组降低47.81%、较B组降低18.59%(P<0.01)。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: The objective of our study was to compare the image quality and radiation dose of computed tomography angiography (CTA) of the kidney in patients with different body mass indexes using routine CT and the latest Gemstone Spectral Imaging (GSI) combination of different noise indexes (NI) with the adaptive statistical iterative reconstruction 2.0 algorithm (ASiR 2.0).

METHODS: A total of 120 patients who had undergone a CTA of the kidney were divided into four groups (A, B, C and D), with 30 patients in each group. Group A underwent a routine CT examination, while groups B, C and D underwent GSI with different noise indexes. All images were restructured using ASiR 2.0. The signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) of all images were calculated when the kidney CTA was completed. Each subjective image evaluation used a 5-point scoring method and was conducted by two independent radiologists. The CT dose index of volume (CTDIvol) and dose-length product (DLP) were recorded, and the mean value was calculated. The DLP was converted to the effective dose (ED). All data were compared with one-way ANOVA.

RESULTS: The SNR, CNR and subjective image quality in group A were significantly lower than in groups B, C and D (P < 0.01). There were no significant differences in SNR, CNR and subjective image quality among groups B, C and D. The ED of group D decreased by 47.81 and 18.59% relative to groups A and B, respectively (P < 0.01).

CONCLUSION: The latest GSI with different NI values can more effectively reduce the radiation dose than can the routine CT scan mode for a kidney CTA while still maintaining diagnostic image quality.

来源:Radiol Med,查看原文。版权归原作者所有。

引用

  1. Zhao Y, Wu Y, Zuo Z, et al. CT angiography of the kidney using routine CT and the latest Gemstone Spectral Imaging combination of different noise indexes: image quality and radiation dose. Radiol Med. 2017. PMID: 28197873. doi:10.1007/s11547-017-0739-5

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。