快速千伏峰值切换双能CT多材料分解算法定量肝脏脂肪含量的临床评估

Multimaterial Decomposition Algorithm for the Quantification of Liver Fat Content by Using Fast-Kilovolt-Peak Switching Dual-Energy CT: Clinical Evaluation.
PubMed 文献 · Radiology · 2017-02-17

这项前瞻性研究纳入33例疑似肝脂肪变性患者,在肝活检前30天内行双能CT(80和140 kVp)及单体素质子MR波谱,用多材料分解(MMD)算法生成脂肪体积分数(FVF)图像。双能CT FVF与MR波谱脂肪分数均随组织学脂肪变性分级增加而升高(趋势检验P均<.001),两者Bland-Altman分析显示比例偏倚(P<.001);真实平扫与三期增强FVF差异的95%一致性界限不超过约2%。

要点

  • 双能CT MMD算法生成的FVF随组织学脂肪变性分级(0-3级)升高而增加(趋势检验P<.001)。
  • 双能CT FVF与MR波谱FF的Bland-Altman分析显示比例偏倚(回归斜率P<.001)。
  • 真实平扫与增强各期FVF差异的95%一致性界限不超过约2%。

核心结论

  • MMD算法在双能CT图像上定量肝脏脂肪准确,且在各成像期之间可重复。

研究设计与样本

  • 前瞻性研究,经机构伦理委员会批准并获患者书面知情同意,纳入33例疑似肝脂肪变性患者,行双能CT(80和140 kVp)平扫及三期动态增强扫描、单体素质子MR波谱,均于肝活检前30天内完成,以肝活检组织学脂肪变性分级(0-3级,按非酒精性脂肪性肝病活动度评分系统定义)为参考标准。

诊断标准与关键数值

  • 双能CT FVF与MR波谱FF均随组织学脂肪变性分级升高而增加(趋势检验P均<.001);双能CT FVF与MR波谱FF的Bland-Altman分析显示比例偏倚(差值对均值回归线斜率显著,P<.001);真实平扫与增强FVF差异的95%一致性界限不超过约2%。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

Purpose To assess the clinical accuracy and reproducibility of liver fat quantification with the multimaterial decomposition (MMD) algorithm, comparing the performance of MMD with that of magnetic resonance (MR) spectroscopy by using liver biopsy as the reference standard. Materials and Methods This prospective study was approved by the institutional ethics committee, and patients provided written informed consent. Thirty-three patients suspected of having hepatic steatosis underwent non-contrast material-enhanced and triple-phase dynamic contrast-enhanced dual-energy computed tomography (CT) (80 and 140 kVp) and single-voxel proton MR spectroscopy within 30 days before liver biopsy. Percentage fat volume fraction (FVF) images were generated by using the MMD algorithm on dual-energy CT data to measure hepatic fat content. FVFs determined by using dual-energy CT and percentage fat fractions (FFs) determined by using MR spectroscopy were compared with histologic steatosis grade (0-3, as defined by the nonalcoholic fatty liver disease activity score system) by using Jonckheere-Terpstra trend tests and were compared with each other by using Bland-Altman analysis. Real non-contrast-enhanced FVFs were compared with triple-phase contrast-enhanced FVFs to determine the reproducibility of MMD by using Bland-Altman analyses. Results Both dual-energy CT FVF and MR spectroscopy FF increased with increasing histologic steatosis grade (trend test, P < .001 for each). The Bland-Altman plot of dual-energy CT FVF and MR spectroscopy FF revealed a proportional bias, as indicated by the significant positive slope of the line regressing the difference on the average (P < .001). The 95% limits of agreement for the differences between real non-contrast-enhanced and contrast-enhanced FVFs were not greater than about 2%. Conclusion The MMD algorithm quantifying hepatic fat in dual-energy CT images is accurate and reproducible across imaging phases. © RSNA, 2017 Online supplemental material is available for this article.

来源:Radiology,查看原文。版权归原作者所有。

引用

  1. Hyodo T, Yada N, Hori M, et al. Multimaterial Decomposition Algorithm for the Quantification of Liver Fat Content by Using Fast-Kilovolt-Peak Switching Dual-Energy CT: Clinical Evaluation. Radiology. 2017. PMID: 28212047. doi:10.1148/radiol.2017160130

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。