多排CT中辐射剂量降低与重建算法对图像噪声、对比度、分辨率及细微低密度肝脏病灶可检测性的影响:滤波反投影与商用基于模型的迭代重建算法的比较
这项前瞻性单中心研究将21例患者CT投影数据重建为6个剂量水平,并植入105个虚拟低密度肝病灶,比较FBP与SAFIRE两种重建算法。SAFIRE使噪声降低53%±5、病灶对比度降低12%±4、锐利度降低13%±10,但CNR提高89%±19;检测准确率平均提高2%(P=.03),估计可降低辐射剂量16%±13。
要点
- SAFIRE较FBP使图像噪声平均降低53%±5,CNR提高89%±19。
- SAFIRE使病灶对比度降低12%±4、边界锐利度降低13%±10。
- SAFIRE检测准确率较FBP平均高2%(P=.03),估计可降低剂量16%±13。
核心结论
- 在给定辐射剂量下SAFIRE提高检测能力(检测准确率约提高2%),并允许以降低的辐射剂量(16%±13)成像,同时保持对细微低密度局灶性肝脏病灶的低对比度检测能力。
研究设计与样本
- 前瞻性、单中心、符合HIPAA的研究,经机构审查委员会批准,使用双源CT系统基于两次CT采集将21例患者的CT投影数据重建为6个辐射剂量水平(12.5%、25%、37.5%、50%、75%和100%)。
关键影像表现/征象
- 每例患者植入5个虚拟肝脏病灶(共105个),病灶与肝脏的预重建对比度为-15 HU,直径12 mm。
- 图像分别用滤波反投影(FBP,B31f核)和正弦图确认迭代重建(SAFIRE,I31f-5核)重建。
- 比较了图像噪声(像素标准差)、重建后病灶对比度、病灶边界锐利度(病灶边界平均归一化梯度)和对比噪声比(CNR)。
诊断标准与关键数值
- 与FBP相比,SAFIRE使噪声平均降低53%±5、病灶对比度降低12%±4、病灶锐利度降低13%±10,但CNR提高89%±19。
- SAFIRE的检测准确率平均较FBP高2%(P=.03),对应的估计辐射剂量降低潜力(±95%置信区间)为16%±13。
临床应用与阅片要点
- 采用双备择强制选择感知实验(16名读者:6名放射科医生、10名医学物理师)比较FBP与SAFIRE的检测准确率。
- 使用线性混合效应统计模型比较FBP与SAFIRE的检测准确率并估计SAFIRE的辐射剂量降低潜力。
局限性
- 估计的剂量降低幅度略小于既往研究提示的幅度。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Purpose To determine the effect of radiation dose and iterative reconstruction (IR) on noise, contrast, resolution, and observer-based detectability of subtle hypoattenuating liver lesions and to estimate the dose reduction potential of the IR algorithm in question. Materials and Methods This prospective, single-center, HIPAA-compliant study was approved by the institutional review board. A dual-source computed tomography (CT) system was used to reconstruct CT projection data from 21 patients into six radiation dose levels (12.5%, 25%, 37.5%, 50%, 75%, and 100%) on the basis of two CT acquisitions. A series of virtual liver lesions (five per patient, 105 total, lesion-to-liver prereconstruction contrast of -15 HU, 12-mm diameter) were inserted into the raw CT projection data and images were reconstructed with filtered back projection (FBP) (B31f kernel) and sinogram-affirmed IR (SAFIRE) (I31f-5 kernel). Image noise (pixel standard deviation), lesion contrast (after reconstruction), lesion boundary sharpness (average normalized gradient at lesion boundary), and contrast-to-noise ratio (CNR) were compared. Next, a two-alternative forced choice perception experiment was performed (16 readers [six radiologists, 10 medical physicists]). A linear mixed-effects statistical model was used to compare detection accuracy between FBP and SAFIRE and to estimate the radiation dose reduction potential of SAFIRE. Results Compared with FBP, SAFIRE reduced noise by a mean of 53% ± 5, lesion contrast by 12% ± 4, and lesion sharpness by 13% ± 10 but increased CNR by 89% ± 19. Detection accuracy was 2% higher on average with SAFIRE than with FBP (P = .03), which translated into an estimated radiation dose reduction potential (±95% confidence interval) of 16% ± 13. Conclusion SAFIRE increases detectability at a given radiation dose (approximately 2% increase in detection accuracy) and allows for imaging at reduced radiation dose (16% ± 13), while maintaining low-contrast detectability of subtle hypoattenuating focal liver lesions. This estimated dose reduction is somewhat smaller than that suggested by past studies. © RSNA, 2017 Online supplemental material is available for this article.
引用
- Solomon J, Marin D, Roy Choudhury K, et al. Effect of Radiation Dose Reduction and Reconstruction Algorithm on Image Noise, Contrast, Resolution, and Detectability of Subtle Hypoattenuating Liver Lesions at Multidetector CT: Filtered Back Projection versus a Commercial Model-based Iterative Reconstruction Algorithm. Radiology. 2017. PMID: 28170300. doi:10.1148/radiol.2017161736