在哥本哈根普通人群研究中采用两种不同多排探测器计算机断层扫描仪以校准质量评分评估冠状动脉钙化
该研究通过体模实验评估多排CT测量校准质量评分的有效性,并比较两种320排CT在人群中的扫描仪间变异。体模测量的质量评分与物理钙质量相关性R²=0.99,人群研究中两种扫描仪的总质量评分和Agatston评分分布无显著差异(p=0.88和p=0.64)。结论认为不同代320排CT测量的钙化质量评分在体模中准确性很高,在人群研究中似乎可以合并使用。
要点
- 体模中校准质量评分与物理钙质量相关性R²=0.99,两种扫描仪测量均值接近物理值。
- 人群研究中两种扫描仪的总质量评分中位数分别为16.8mg和15.8mg,无显著差异(p=0.88)。
- 两种扫描仪的总Agatston评分中位数分别为92和89,无显著差异(p=0.64)。
核心结论
- 不同代320排CT在体模中测量校准钙化质量评分准确性很高,在人群研究中似乎可以合并不同320排CT获得的钙化评分。
研究设计与样本
- 体模研究加人群研究;体模用第一代(A)和第二代(B)320排CT重复扫描;人群样本来自哥本哈根普通人群研究,扫描仪A组N=254,扫描仪B组N=253;参考标准为体模已知物理钙质量。
诊断标准与关键数值
- 体模平均总质量评分:扫描仪A为32.9±0.8mg,扫描仪B为33.1±0.9mg(p=0.43),物理钙质量为34.0mg;两种扫描仪测量质量评分与物理钙质量相关性均为R²=0.99。人群研究总质量评分中位数:扫描仪A为16.8mg(IQR 3.5-81.1),扫描仪B为15.8mg(IQR 3.8-63.4)(p=0.88);总Agatston评分中位数:扫描仪A为92(IQR 23-471),扫描仪B为89(IQR 40-384)(p=0.64)。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: Population studies have shown coronary calcium score to improve risk stratification in subjects suspected for cardiovascular disease. The aim of this work was to assess the validity of multidetector computed tomography (MDCT) for measurement of calibrated mass scores (MS) in a phantom study, and to investigate inter-scanner variability for MS and Agaston score (AS) recorded in a population study on two different high-end MDCT scanners.
MATERIALS AND METHODS: A calcium phantom was scanned by a first (A) and second (B) generation 320-MDCT. MS was measured for each calcium deposit from repeated measurements in each scanner and compared to known physical phantom mass. Random samples of human subjects from the Copenhagen General Population Study were scanned with scanner A (N=254) and scanner B (N=253) where MS and AS distributions of these two groups were compared.
RESULTS: The mean total MS of the phantom was 32.9±0.8mg and 33.1±0.9mg (p=0.43) assessed by scanner A and B respectively - the physical calcium mass was 34.0mg. Correlation between measured MS and physical calcium mass was R2=0.99 in both scanners. In the population study the median total MS was 16.8mg (interquartile range (IQR): 3.5-81.1) and 15.8mg (IQR: 3.8-63.4) in scanner A and B (p=0.88). The corresponding median total AS were 92 (IQR: 23-471) and 89 (IQR: 40-384) (p=0.64).
CONCLUSION: Calibrated calcium mass score may be assessed with very high accuracy in a calcium phantom by different generations of 320-MDCT scanners. In population studies, it appears acceptable to pool calcium scores acquired on different 320-MDCT scanners.
引用
- Fuchs A, Groen JM, Arnold BA, et al. Assessment of coronary calcification using calibrated mass score with two different multidetector computed tomography scanners in the Copenhagen General Population Study. Eur J Radiol. 2016. PMID: 28189204. doi:10.1016/j.ejrad.2016.12.033