基于模型的迭代重建在冠状动脉CT血管造影中评估冠状动脉支架的研究
本研究比较了23例患者32枚冠状动脉支架在冠状动脉CT血管造影中采用前向投影基于模型的迭代重建解决方案(FIRST)与自适应迭代剂量降低3D(AIDR 3D)重建的图像质量。结果显示FIRST图像的边缘上升斜率更高、可见支架管腔直径比率更高、支架管腔衰减增加比率更低,主观图像质量评分也更高,提示FIRST图像质量优于AIDR 3D。
要点
- FIRST重建的支架边缘上升斜率高于AIDR 3D(325.2 HU/mm vs 224.4 HU/mm,P<0.01)。
- FIRST的可见支架管腔直径比率高于AIDR 3D(51.4% vs 47.3%,P<0.01)。
- FIRST的支架管腔衰减增加比率低于AIDR 3D(0.19 vs 0.30,P<0.01)。
核心结论
- 冠状动脉支架扫描在FIRST重建上的图像质量优于AIDR 3D重建。
研究设计与样本
- 研究纳入23例患者的32枚冠状动脉支架,行冠状动脉CT血管造影,图像分别用AIDR 3D和FIRST重建,并测量边缘上升距离、边缘上升斜率、支架管腔衰减增加比率和可见支架管腔直径,由两名放射科医师采用4分制(1=差,4=优)主观评价支架图像质量。
关键影像表现/征象
- FIRST图像边缘上升斜率高于AIDR 3D图像(325.2 HU/mm vs 224.4 HU/mm,P<0.01)。
- FIRST图像可见支架管腔直径与真实直径的比率高于AIDR 3D图像(51.4% vs 47.3%,P<0.01)。
- FIRST图像支架管腔衰减增加比率低于AIDR 3D图像(0.19 vs 0.30,P<0.01)。
- AIDR 3D与FIRST图像的平均主观图像质量评分分别为3.18和3.63,差异有统计学意义(P<0.01)。
诊断标准与关键数值
- 图像质量主观评分采用4分制:1=差,4=优。
- 支架管腔衰减增加比率(SAIR)FIRST为0.19,AIDR 3D为0.30(P<0.01)。
- 可见支架管腔直径比率FIRST为51.4%,AIDR 3D为47.3%(P<0.01)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
RATIONALE AND OBJECTIVES: This study aims to compare the image quality of coronary artery stent scans on computed tomography images reconstructed with forward projected model-based iterative reconstruction solution (FIRST) and adaptive iterative dose reduction 3D (AIDR 3D).
MATERIALS AND METHODS: Coronary computed tomography angiography scans of 23 patients with 32 coronary stents were used. The images were reconstructed with AIDR 3D and FIRST. We generated computed tomography attenuation profiles across the stents and measured the width of the edge rise distance and the edge rise slope (ERS). We also calculated the stent lumen attenuation increase ratio (SAIR) and measured visible stent lumen diameters. Two radiologists visually evaluated the image quality of the stents using a 4-point scale (1 = poor, 4 = excellent).
RESULTS: There was no significant difference in the edge rise distance between the two reconstruction methods (P = 0.36). The ERS on FIRST images was greater than the ERS on AIDR 3D images (325.2 HU/mm vs 224.4 HU/mm; P <0.01). The rate of the visible stent lumen diameter compared to the true diameter on FIRST images was higher than that on AIDR 3D images (51.4% vs 47.3%, P <0.01). The SAIR on FIRST images was lower than the SAIR on AIDR 3D images (0.19 vs 0.30, P <0.01). The mean image quality scores for AIDR 3D and FIRST images were 3.18 and 3.63, respectively; the difference was also significant (P <0.01).
CONCLUSION: The image quality of coronary artery stent scans is better on FIRST than on AIDR 3D images.
引用
- Tatsugami F, Higaki T, Sakane H, et al. Coronary Artery Stent Evaluation with Model-based Iterative Reconstruction at Coronary CT Angiography. Acad Radiol. 2017. PMID: 28214228. doi:10.1016/j.acra.2016.12.020
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。