十名观察者使用迭代重建算法的胸部数字断层合成对磨玻璃结节检出能力的体模研究:与辐射剂量和结节特征的关系

A phantom study for ground-glass nodule detectability using chest digital tomosynthesis with iterative reconstruction algorithm by ten observers: association with radiation dose and nodular characteristics.
PubMed 文献 · Br J Radiol · 2017-02-17

该体模研究让10名放射科医生在12组胸部数字断层合成图像上检出模拟磨玻璃结节,比较6种辐射剂量与2种重建算法组合的检出能力,并分析其与结节大小和密度的关系;结果显示迭代重建可改善较大、衰减较低结节的检出。

要点

  • 迭代重建可提升10 mm、-800 HU磨玻璃结节的检出敏感度。
  • 120 kV-160 mA联合迭代重建与120 kV-80 mA联合迭代重建检出能力相似。
  • 较大且衰减较高的磨玻璃结节(>8 mm)在最低剂量0.17 mSv下仍可检出。

核心结论

  • 胸部数字断层合成对较大、衰减较高的模拟磨玻璃结节(>8 mm)在最低辐射剂量0.17 mSv下仍显示足够检出能力,迭代重建可改善直径10 mm、衰减较低结节的检出能力。

研究设计与样本

  • 体模研究:在胸部体模中按14种结节分布模式放置74个模拟磨玻璃结节,直径5、8、10 mm,密度-630和-800 HU。
  • 采用胸部数字断层合成在6种辐射剂量下(120 kV-10 mA、20 mA、80 mA、160 mA,100 kV-80 mA和80 kV-320 mA)联合或不联合迭代重建获取12组冠状面图像。
  • 由10名放射科医生以连续分布评分记录磨玻璃结节的存在和位置,采用受试者工作特征分析比较12组图像的检出能力。

诊断标准与关键数值

  • 120 kV-160 mA联合迭代重建与120 kV-80 mA联合迭代重建(0.614 mSv)的受试者工作特征曲线下面积分别为0.798 ± 0.024和0.788 ± 0.025,两者相似,且高于120 kV下获取的6组图像(p < 0.05)。
  • 对于-630 HU、8 mm结节,120 kV-10 mA不联合迭代重建的检出敏感度为73.5 ± 6.0%,与其他11种数据采集方法相似(p = 0.157)。
  • 对于-800 HU、10 mm结节,120 kV-80 mA和120 kV-160 mA不联合迭代重建的检出敏感度均为56.3 ± 11.9%,联合迭代重建后检出敏感度提高(p < 0.05)。

临床应用与阅片要点

  • 对于较大、衰减较高的磨玻璃结节(>8 mm),胸部数字断层合成在低至0.17 mSv的辐射剂量下仍可检出。
  • 对于直径10 mm、衰减较低的磨玻璃结节,在亚毫西弗剂量下联合迭代重建可改善检出能力。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: To compare detectability of simulated ground-glass nodules (GGNs) on chest digital tomosynthesis (CDT) among 12 images obtained at 6 radiation doses using 2 reconstruction algorithms and to analyze its association with nodular size and density.

METHODS: 74 simulated GGNs [5, 8 and 10 mm in diameter/-630 and -800 Hounsfield units (HU) in density] were placed in a chest phantom in 14 nodular distribution patterns. 12 sets of coronal images were obtained using CDT at 6 radiation doses: 120 kV-10 mA/20 mA/80 mA/160 mA, 100 kV-80 mA and 80 kV-320 mA with and without iterative reconstruction (IR). 10 radiologists recorded GGN presence and locations by continuously distributed rating. GGN detectability was compared by receiver operating characteristic analysis among 12 images and detection sensitivities (DS) were compared among 12 images in subgroups classified by nodular diameters and densities.

RESULTS: GGN detectability at 120 kV-160 mA with IR was similar to that at 120 kV-80 mA with IR (0.614 mSv), as area under receiver operating characteristic curve was 0.798 ± 0.024 and 0.788 ± 0.025, respectively, and higher than six images acquired at 120 kV (p < 0.05). For nodules of -630 HU/8 mm, DS at 120 kV-10 mA without IR was 73.5 ± 6.0% and was similar to that by the other 11 data acquisition methods (p = 0.157). For nodules of -800 HU/10 mm, DS both at 120 kV-80 mA and 120 kV-160 mA without IR was improved by IR (56.3 ± 11.9%) (p < 0.05).

CONCLUSION: CDT demonstrated sufficient detectability for larger more-attenuated GGNs (>8 mm) even in the lowest radiation dose (0.17 mSv) and improved detectability for less-attenuated GGNs with the diameter of 10 mm at submillisievert with IR. Advances in knowledge: IR improved detectability for larger less-attenuated simulated GGNs on CDT.

来源:Br J Radiol,查看原文。版权归原作者所有。

引用

  1. Miyata K, Nagatani Y, Ikeda M, et al. A phantom study for ground-glass nodule detectability using chest digital tomosynthesis with iterative reconstruction algorithm by ten observers: association with radiation dose and nodular characteristics. Br J Radiol. 2017. PMID: 28102693. doi:10.1259/bjr.20160555