锡预滤过光谱成形降低鼻旁窦CT辐射剂量
前瞻性随机对照研究纳入100例连续患者,比较100 kV锡预滤过(200 mAs)与100 kV无光谱成形低剂量方案(25 mAs)行鼻旁窦CT的图像质量与辐射剂量。两组图像噪声及CNR无显著差异,研究组辐射暴露与有效剂量显著降低,作者结论为锡预滤过可在维持诊断图像质量下进一步降低17%辐射剂量。
要点
- 锡预滤过鼻旁窦CT方案(100 kV、200 mAs)图像噪声与常规低剂量方案(100 kV、25 mAs)相当。
- 研究组容积CT剂量指数1.7 mGy、有效剂量0.055 mSv,均显著低于对照组。
- 所有扫描均达诊断图像质量,锡预滤过可在低剂量基础上进一步降低17%辐射剂量。
核心结论
- 对于低剂量鼻旁窦CT,采用锡预滤过可在维持诊断图像质量的前提下将辐射剂量进一步降低17%。
研究设计与样本
- 前瞻性随机对照研究,连续纳入100例患者并随机分配至研究组或对照组,使用第三代双源CT,研究组方案为100 kV管电压联合锡预滤过(200 mAs),对照组为100 kV无光谱成形低剂量方案(25 mAs)。
关键影像表现/征象
- 所有扫描图像均达诊断质量。
- 空气、球后脂肪及眼球内图像噪声在两组间相当(均p>0.05)。
- 眼球/空气对比噪声比(CNR)两组间无显著差异(p=0.7)。
诊断标准与关键数值
- 研究组容积CT剂量指数1.7 mGy vs 对照组2.1 mGy(p<0.01),研究组降低。
- 研究组有效剂量0.055 mSv vs 对照组0.066 mSv(p<0.01),研究组降低。
- 作者结论为锡预滤过使低剂量鼻旁窦CT辐射剂量进一步降低17%。
临床应用与阅片要点
- 鼻旁窦CT采用100 kV联合锡预滤过(200 mAs)可在维持诊断图像质量的同时降低辐射剂量。
- 阅片时两组图像噪声及CNR无显著差异,锡预滤过方案图像可满足诊断需求。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
INTRODUCTION: Spectral shaping aims to narrow the X-ray spectrum of clinical CT. The aim of this study was to determine the image quality and the extent of radiation dose reduction that can be achieved by tin prefiltration for parasinus CT.
METHODS: All scans were performed with a third generation dual-source CT scanner. A study protocol was designed using 100 kV tube voltage with tin prefiltration (200 mAs) that provides image noise levels comparable to a low-dose reference protocol using 100 kV without spectral shaping (25 mAs). One hundred consecutive patients were prospectively enrolled and randomly assigned to the study or control group. All patients signed written informed consent. The study protocol was approved by the local Institutional Review Board and applies to the HIPAA. Subjective and objective image quality (attenuation values, image noise, and contrast-to-noise ratio (CNR)) were assessed. Radiation exposure was assessed as volumetric CT dose index, and effective dose was estimated. Mann-Whitney U test was performed for radiation exposure and for image noise comparison.
RESULTS: All scans were of diagnostic image quality. Image noise in air, in the retrobulbar fat, and in the eye globe was comparable between both groups (all p > 0.05). CNReye globe/air did not differ significantly between both groups (p = 0.7). Radiation exposure (1.7 vs. 2.1 mGy, p < 0.01) and effective dose (0.055 vs. 0.066 mSv, p < 0.01) were significantly reduced in the study group.
CONCLUSION: Radiation dose can be further reduced by 17% for low-dose parasinus CT by tin prefiltration maintaining diagnostic image quality.
引用
- May MS, Brand M, Lell MM, et al. Radiation dose reduction in parasinus CT by spectral shaping. Neuroradiology. 2017. PMID: 28091696. doi:10.1007/s00234-016-1780-0
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。