血管内超声与数字减影血管造影:儿童及青少年影像中管腔内径测量的直接比较
回顾性分析2014年10月至2016年3月20例患儿22次血管内介入操作的56对IVUS与DSA管径测量,IVUS平均较DSA大0.6±2.1 mm(95%CI -0.01至1.12,P=0.06);排除静脉压迫综合征后IVUS较DSA大0.7±1.6 mm(95%CI 0.14至1.18,P=0.01);静脉压迫综合征亚组IVUS较DSA大0.3±3.0 mm(95%CI -1.16至1.82,P=0.65)。
要点
- 儿童及青少年中IVUS测得的绝对管径总体略大于DSA。
- 排除静脉压迫综合征后IVUS较DSA大0.7±1.6 mm,差异有统计学意义(P=0.01)。
- 静脉压迫综合征中IVUS可能比DSA更准确反映血管受压与管径。
核心结论
- 在所有数据子集中IVUS测量值均略大于DSA;排除静脉压迫综合征后差异有统计学意义,静脉压迫综合征中IVUS可能更准确反映血管受压与管径。
研究设计与样本
- 回顾性 review,纳入2014年10月至2016年3月所有同时使用IVUS和DSA的儿童及青少年血管内介入操作,共20例患者22次操作、56对测量(DSA=56,IVUS=56),用SAS软件配对t检验比较血管直径测量值。
诊断标准与关键数值
- 全部数据:IVUS较DSA平均大0.6±2.1 mm(95%CI -0.01至1.12,P=0.06,r=0.90)。
- 排除静脉压迫综合征(May-Thurner、胡桃夹、上腔静脉综合征):IVUS较DSA平均大0.7±1.6 mm(95%CI 0.14至1.18,P=0.01,r=0.93)。
- 静脉压迫综合征单独分析:IVUS较DSA平均大0.3±3.0 mm(95%CI -1.16至1.82,P=0.65,r=0.45)。
临床应用与阅片要点
- 儿童及青少年血管内介入中,IVUS与DSA管径测量总体一致性较好(r=0.90),但IVUS测值略大。
- 排除静脉压迫综合征后IVUS测值较DSA大0.7±1.6 mm且差异有统计学意义,报告管径时需注意测量方法差异。
- 静脉压迫综合征中IVUS与DSA相关性较低(r=0.45),作者认为IVUS可能更准确反映血管受压与管径。
局限性
- 本文为回顾性 review。
- 静脉压迫综合征亚组相关性较低(r=0.45),95%CI较宽(-1.16至1.82)。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND: Intravascular ultrasound (IVUS) allows intraluminal imaging of blood vessels rather than the one-dimensional luminal outline depicted by digital subtraction angiography (DSA). Despite extensive literature in multiple adult vascular diseases, IVUS has not been directly compared to DSA in pediatric and adolescent vascular pathologies.
OBJECTIVE: The purpose of this manuscript is to compare absolute luminal diameter measurements obtained via IVUS and DSA during a variety of pediatric endovascular procedures.
MATERIALS AND METHODS: We conducted a retrospective review of all pediatric and adolescent endovascular procedures from October 2014 to March 2016 in which IVUS and DSA were used. We compared the vessel diameter measurements and analyzed them using SAS software with a paired t-test.
RESULTS: There were 102 total measurements (DSA = 56; IVUS = 56; 22 procedures; 20 patients). On average, IVUS measured 0.6 ± 2.1 mm larger than DSA (95% confidence interval [CI] -0.01 to 1.12; P = 0.06; r = 0.90). When venous compression syndrome (May-Thurner, Nutcracker, superior vena cava syndrome) measurements were excluded, IVUS measured 0.7 ± 1.6 mm larger than DSA (95% CI 0.14 to 1.18; P = 0.01; r = 0.93). When venous compression syndrome measurements were evaluated separately, IVUS measured 0.3 ± 3.0 mm larger than DSA (95% CI -1.16 to 1.82; P = 0.65; r = 0.45).
CONCLUSION: Overall, IVUS measurements were slightly larger than DSA measurements in all data subsets. Absolute vessel diameter measurements obtained with IVUS in the pediatric and adolescent population are statistically significantly larger than those obtained using DSA when excluding venous compression syndromes. In venous compression syndromes, IVUS might provide a more accurate representation of vessel compression and diameter than DSA.
引用
- Gill AE, Ciszak T, Braun H, et al. Intravascular ultrasound versus digital subtraction angiography: direct comparison of intraluminal diameter measurements in pediatric and adolescent imaging. Pediatr Radiol. 2017. PMID: 28102453. doi:10.1007/s00247-016-3771-z
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。