乳腺组织标记物在专用锥形束乳腺CT中所致伪影与全视野数字乳腺X线摄影的比较

Artifacts Caused by Breast Tissue Markers in a Dedicated Cone-beam Breast CT in Comparison to Full-field Digital Mammography.
PubMed 文献 · Acad Radiol · 2017-01-24

体外研究用29种市售乳腺组织标记物置于猪碎肉模体中,比较专用锥形束乳腺CT(CBBCT)与全视野数字乳腺X线摄影的伪影。所有标记物在乳腺X线摄影上均可见且无伪影,但在CBBCT上均产生伪影,中位长度7.2 mm(范围0–48.3 mm),轴位最大(中位12.6 mm)。作者结论为标记物伪影可显著影响CBBCT图像质量,从而限制乳腺癌原发诊断与随访。

要点

  • 所有29种标记物在CBBCT上均产生伪影,而在全视野数字乳腺X线摄影上均无伪影。
  • CBBCT伪影中位长度7.2 mm(范围0–48.3 mm,四分位距4.3–11.4 mm),轴位最大(中位12.6 mm)。
  • 伪影大小取决于标记物成分、大小、形状、重量和方向以及重建图像平面。

核心结论

  • 标记物所致伪影可显著影响CBBCT图像质量,从而限制乳腺癌的原发诊断与随访。
  • 伪影大小取决于标记物特征、方向及重建图像平面。

研究设计与样本

  • 体外研究,使用猪碎肉自制组织模体。
  • 评估29种市售标记物,大小、成分和形状各异。
  • CBBCT采集参数为49 kVp、50 mA和100 mA,标记物方向平行和垂直于扫描方向。
  • 图像在矢状位、轴位和冠状位以0.5 mm层厚评估,并与全视野数字乳腺X线摄影对照测量。

关键影像表现/征象

  • 所有标记物在全视野数字乳腺X线摄影上均可见且无任何伪影。
  • 所有标记物在CBBCT上均产生伪影,以条状伪影长度测量。
  • CBBCT伪影中位长度7.2 mm,范围0–48.3 mm(四分位距4.3–11.4 mm)。
  • 轴位伪影最大,中位12.6 mm,范围0–48.3 mm,相对伪影长度(伪影毫米数/标记物实际物理长度)为4.1(四分位距2.3–6.1,范围0–8.7)。

临床应用与阅片要点

  • 原文结论指出标记物伪影可显著影响CBBCT图像质量,从而限制乳腺癌原发诊断与随访。
  • 伪影长度与标记物成分、大小、形状、重量和方向有关,轴位重建时伪影最大。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

RATIONALE AND OBJECTIVES: The purpose of this ex vivo study was to investigate artifacts in a cone-beam breast computed tomography (CBBCT) caused by breast tissue markers.

MATERIALS AND METHODS: Breast phantoms with self-made tissue pork mincemeat were created. Twenty-nine different, commercially available markers with varying marker size, composition, and shape were evaluated. A dedicated CBBCT evaluation of all phantoms was performed with 49 kVp, 50 and 100 mA, and marker orientation parallel and orthogonal to the scan direction. The resultant images were evaluated in sagittal, axial, and coronal view with a slice thickness of 0.5 mm. Additionally, measurements of all markers in the same directions were done with full-field digital mammography.

RESULTS: All markers were visible in full-field digital mammography without any artifacts. However, all markers caused artifacts on a CBBCT. Artifacts were measured as the length of the resulting streakings. Median length of artifacts was 7.2 mm with a wide range from 0 to 48.3 mm (interquartile range 4.3-11.4 mm) dependent on composition, size, shape, weight, and orientation of the markers. The largest artifacts occurred in axial view with a median size of 12.6 mm, with a range from 0 to 48.3 mm, resulting in a relative artifact length (quotient artifact in mm/real physical length of the marker itself) of 4.1 (interquartile range 2.3-6.1, range 0-8.7).

CONCLUSIONS: Artifacts caused by markers can significantly influence image quality in a CBBCT, thus limiting primary diagnostics and follow-up in breast cancer. The size of the artifacts depends on the marker characteristics, orientation, and the image plane of reconstruction.

来源:Acad Radiol,查看原文。版权归原作者所有。