标准剂量与合成数字乳腺X线摄影自动乳腺密度估计的一致性:大型筛查队列结果
本研究回顾性分析3668例阴性数字乳腺断层合成筛查检查,比较标准剂量与合成乳腺X线摄影的自动乳腺密度估计一致性。合成与标准剂量乳腺密度高度相关(r=0.92,P<.001),Bland-Altman 95%一致性界限为-6.4%至9.9%;合成图像密度平均高1.7%(P<.001),致密乳腺组织中差异更大1.56%(P<.0001)。结论认为合成图像自动密度估计与标准剂量总体可比。
要点
- 合成与标准剂量乳腺密度高度相关(r=0.92,P<.001),平均差异1.7%。
- Bland-Altman 95%一致性界限为-6.4%至9.9%。
- 致密乳腺组织中不一致性更大1.56%(P<.0001)。
核心结论
- 合成乳腺X线摄影的全自动乳腺密度估计与标准剂量乳腺X线摄影总体可比,在数字乳腺断层合成筛查中标准二维图像逐渐被合成图像替代以降低辐射剂量的背景下可能具有重要意义。
研究设计与样本
- 回顾性分析一个机构4个月期间连续进行的3668例阴性(BI-RADS 1或2类)数字乳腺断层合成筛查检查,所有图像使用Selenia Dimensions系统(Hologic,Bedford,Mass)采集,合成图像由FDA批准的C-View软件模块生成,标准剂量For Presentation图像与合成图像均用全自动算法分析。
诊断标准与关键数值
- 合成与标准剂量乳腺密度估计高度相关(r=0.92,P<.001),Bland-Altman 95%一致性界限为-6.4%至9.9%;合成图像密度估计平均比标准剂量高1.7%(P<.001),致密乳腺组织中不一致性更大1.56%(P<.0001)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Purpose To evaluate agreement between automated estimates of breast density made from standard-dose versus synthetic digital mammograms in a large cohort of women undergoing screening. Materials and Methods This study received institutional review board approval with waiver of consent. A total of 3668 negative (Breast Imaging Reporting and Data System category 1 or 2) digital breast tomosynthesis (DBT) screening examinations consecutively performed over a 4-month period at one institution for which both standard-dose and synthetic mammograms were available for analysis were retrospectively analyzed. All mammograms were acquired with a Selenia Dimensions system (Hologic, Bedford, Mass), and synthetic mammograms were generated by using the U.S. Food and Drug Administration-approved "C-View" software module. The "For Presentation" standard-dose mammograms and synthetic images were analyzed by using a fully automated algorithm. Agreement between density estimates was assessed by using Pearson correlation, linear regression, and Bland-Altman analysis. Differences were evaluated by using the paired Student t test. Results Breast percentage density (PD) estimates from synthetic and standard-dose mammograms were highly correlated (r = 0.92, P < .001), and the 95% Bland-Altman limits of agreement between PD estimates were -6.4% to 9.9%. Synthetic mammograms had PD estimates by an average of 1.7% higher than standard-dose mammograms (P < .001), with a larger disagreement by 1.56% in women with highly dense breast tissue (P < .0001). Conclusion Fully automated estimates of breast density made from synthetic mammograms are generally comparable to those made from standard-dose mammograms. This may be important, as standard two-dimensional mammographic images are increasingly being replaced by synthetic mammograms in DBT screening in an attempt to reduce radiation dose. © RSNA, 2017 Online supplemental material is available for this article.
引用
- Conant EF, Keller BM, Pantalone L, et al. Agreement between Breast Percentage Density Estimations from Standard-Dose versus Synthetic Digital Mammograms: Results from a Large Screening Cohort Using Automated Measures. Radiology. 2017. PMID: 28121523. doi:10.1148/radiol.2016161286
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。