联合BI-RADS分析与超声回声Nakagami统计在乳腺病变诊断中的价值
研究纳入107个乳腺病变(32恶性、75良性),用Nakagami形状参数m及其与周围组织的差值结合BI-RADS分类鉴别肿瘤。BI-RADS的AUC为0.966,敏感度100%、特异度54.67%;联合B模式与mLmin参数AUC升至0.978(p=0.088),较单用BI-RADS无显著提高,但可能有助于识别BI-RADS 4a中以良性病变为主、恶性怀疑度低的患者组。
要点
- BI-RADS分类诊断乳腺病变AUC为0.966,敏感度100%、特异度54.67%,34个良性病变被不必要活检。
- Nakagami参数中Δmavg敏感度最高达90%,32个恶性病变中仅3个假阴性,这3个均为BI-RADS 4c类。
- B模式联合mLmin参数使AUC升至0.978(p=0.088),较单用BI-RADS无显著提高。
核心结论
- 参数成像与BI-RADS评估联合并未显著提高乳腺病变的鉴别能力,但可能有助于更好地识别BI-RADS 4a评分、恶性怀疑度低且以良性病变为主的患者组。
研究设计与样本
- 研究纳入107个乳腺病变(32个恶性、75个良性)及其周围组织,记录射频回波信号原始数据和B模式图像,评估Nakagami形状参数m的三种特征值(mLmax、mLmin、mLavg)及病变与周围组织的m参数差值(Δmmax、Δmmin、Δmavg);以BI-RADS 3/4为界值,3类视为良性、4类视为恶性。
诊断标准与关键数值
- BI-RADS的AUC为0.966,敏感度100%、特异度54.67%;Nakagami统计中mLavg的敏感度与特异度之和最佳(62.5%和93.33%);Δmavg敏感度最高为90%(32个病变中仅3个假阴性);B模式联合mLmin参数AUC为0.978(p=0.088),单用BI-RADS为0.966。
临床应用与阅片要点
- 作者认为参数成像与BI-RADS评估联合并未显著提高乳腺病变的鉴别能力,但可能有助于更好地识别BI-RADS 4a评分、恶性怀疑度低且以良性病变为主的患者组。
局限性
- 联合B模式与mLmin参数较单用BI-RADS的AUC提高未达统计学显著(p=0.088)。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
AIM: To develop a method combining the statistics of the ultrasound backscatter and the Breast Imaging-Reporting and Data System (BI-RADS) classification to enhance the differentiation of breast tumours.
MATERIALS AND METHODS: The Nakagami shape parameter m was used to characterise the scatter properties of breast tumours. Raw data from the radiofrequency (RF) echo-signal and B-mode images from 107 (32 malignant and 75 benign) lesions and their surrounding tissue were recorded. Three different characteristic values of the shape parameters of m (maximum [mLmax], minimum [mLmin] and average [mLavg]) and differences between m parameters (Δmmax, Δmmin, Δmavg) of the lesions and their surrounding tissues were assessed. A lesion with a BI-RADS score of 3 was considered benign, while a lesion with a score of 4 was considered malignant (a cut-off of BI-RADS 3/4 was set for all patients).
RESULTS: The area under the receiver operating characteristic (ROC) curve (AUC) was equal to 0.966 for BI-RADS, with 100% sensitivity and 54.67% specificity. All malignant lesions were diagnosed correctly, whereas 34 benign lesions were biopsied unnecessarily. In assessing the Nakagami statistics, the sum of the sensitivity and specificity was the best for mLavg (62.5% and 93.33%, respectively). Only four of 20 lesions were found over the cut-off value in BI-RADS of 4a. When comparing the differences in m parameters, Δmavg had the highest sensitivity of 90% (only three of 32 lesions were false negative). These three lesions were classified as BI-RADS category 4c. The combined use of B-mode and mLmin parameter improve the AUC up to 0.978 (p=0.088), compared to BI-RADS alone.
CONCLUSION: The combination of the parametric imaging and the BI-RADS assessment does not significantly improve the differentiation of breast lesions, but it has the potential to better identify the group of patients with mainly benign lesions that have a low level of suspicion for malignancy with a BI-RADS score of 4a.