对比增强能谱乳腺摄影与二维全视野数字乳腺摄影及三维断层摄影的乳腺辐射剂量比较

Breast Radiation Dose With CESM Compared With 2D FFDM and 3D Tomosynthesis Mammography.
PubMed 文献 · AJR Am J Roentgenol · 2017-01-01

该回顾性研究在体模和患者中比较了对比增强能谱乳腺摄影(CESM)、二维全视野数字乳腺摄影(FFDM)和三维断层摄影的辐射剂量。CESM在平均乳腺厚度63 mm时平均腺体剂量(AGD)为3.0±1.1 mGy,较二维FFDM高约0.9 mGy、较三维断层摄影高约0.5 mGy,但总体仍低于乳腺摄影质量标准法案规定的3 mGy剂量限值。

要点

  • CESM在平均乳腺厚度63 mm时AGD为3.0±1.1 mGy,二维FFDM为2.1 mGy,三维断层摄影为2.5 mGy。
  • CESM提供类似二维FFDM的标准图像,该图像由低能投影获得。
  • CESM总体AGD低于乳腺摄影质量标准法案规定的3 mGy剂量限值。

核心结论

  • CESM在平均乳腺厚度63 mm时AGD较二维FFDM增加约0.9 mGy、较三维断层摄影增加约0.5 mGy,总体AGD低于乳腺摄影质量标准法案规定的3 mGy剂量限值。

研究设计与样本

  • 回顾性研究,纳入6214例接受二维FFDM、3662例接受三维断层摄影、173例接受CESM的患者,均选取单侧左头尾位投影确定剂量;另在模拟非致密与致密乳腺组织的体模中测量剂量。

诊断标准与关键数值

  • CESM在平均乳腺厚度63 mm时AGD为3.0±1.1 mGy,同厚度下二维FFDM为2.1 mGy、三维断层摄影为2.5 mGy;非致密体模平均AGD二维FFDM 1.0 mGy、三维断层摄影1.3 mGy、CESM 1.6 mGy;致密体模平均AGD二维FFDM 1.3 mGy、三维断层摄影1.4 mGy、CESM 2.1 mGy;压缩厚度4.5 cm时致密体模CESM辐射暴露较非致密体模高约25%;致密体模压缩厚度6 cm时剂量较非致密体模压缩厚度4.5 cm时高约42%。

临床应用与阅片要点

  • 原文结论指出CESM提供类似二维FFDM的标准图像,该图像由低能投影获得,且总体AGD低于乳腺摄影质量标准法案规定的3 mGy剂量限值。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: We aimed to compare radiation dose received during contrast-enhanced spectral mammography (CESM) using high- and low-energy projections with radiation dose received during 2D full field digital mammography (FFDM) and 3D tomosynthesis on phantoms and patients with varying breast thickness and density.

MATERIALS AND METHODS: A single left craniocaudal projection was chosen to determine the doses for 6214 patients who underwent 2D FFDM, 3662 patients who underwent 3D tomosynthesis, and 173 patients who underwent CESM in this retrospective study. Dose measurements were also collected in phantoms with composition mimicking nondense and dense breast tissue.

RESULTS: Average glandular dose (AGD) ± SD was 3.0 ± 1.1 mGy for CESM exposures at a mean breast thickness of 63 mm. At this thickness, the dose was 2.1 mGy from 2D FFDM and 2.5 mGy from 3D tomosynthesis. The nondense phantom had a mean AGD of 1.0 mGy with 2D FFDM, 1.3 mGy with 3D tomosynthesis, and 1.6 mGy with CESM. The dense breast phantom had a mean AGD of 1.3 mGy with 2D FFDM, 1.4 mGy with 3D tomosynthesis, and 2.1 mGy with CESM. At a compressed thickness of 4.5 cm, radiation exposure from CESM was approximately 25% higher in dense breast phantoms than in nondense breast phantoms. The dose in the dense phantom at a compressed thickness of 6 cm was approximately 42% higher than the dose in the nondense phantom at a compressed thickness of 4.5 cm.

CONCLUSION: CESM was found to increase AGD at a mean breast thickness of 63 mm by approximately 0.9 mGy and 0.5 mGy compared with 2D FFDM and 3D tomosynthesis, respectively. Of note, CESM provides a standard image (similar to 2D FFDM) that is obtained using the low-energy projection. Overall, the AGD from CESM falls below the dose limit of 3 mGy set by Mammography Quality Standards Act regulations.

来源:AJR Am J Roentgenol,查看原文。版权归原作者所有。