挪威乳腺癌筛查项目中使用的压迫力
研究纳入2014年1-3月在挪威乳腺癌筛查项目14个乳腺中心接受全视野数字乳腺摄影的17,951名随机选取女性,分析左侧乳房头尾位和内外斜位压迫力。所有乳腺摄影平均压迫力116 N,各中心间范围91-147 N,约59%的乳腺摄影符合推荐范围11-18 kg(108-177 N);中心间差异大于中心内设备间和放射技师间差异,提示需要基于证据的压迫力建议。
要点
- 所有乳腺摄影平均压迫力116 N,各中心间范围91-147 N。
- 约59%的乳腺摄影符合推荐压迫力范围11-18 kg(108-177 N)。
- 中心间压迫力差异大于中心内设备间(137-155 N)和放射技师间(95-143 N)差异。
核心结论
- 挪威乳腺癌筛查项目中各乳腺中心应用的压迫力存在广泛差异,提示需要基于证据的压迫力建议以优化图像质量并个体化乳房压迫。
研究设计与样本
- 研究纳入2014年1月至3月在挪威乳腺癌筛查项目14个乳腺中心接受全视野数字乳腺摄影的17,951名随机选取女性,分析左侧乳房头尾位和内外斜位压迫力。
诊断标准与关键数值
- 全视野数字乳腺摄影推荐压迫力范围为11-18 kg(108-177 N)。
- 所有乳腺摄影平均压迫力为116 N,各乳腺中心间范围为91 N至147 N。
- 约59%的乳腺摄影符合推荐压迫力范围。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: Compression is used in mammography to reduce breast thickness, which is claimed to improve image quality and reduce radiation dose. In the Norwegian Breast Cancer Screening Program (NBCSP), the recommended range of compression force for full-field digital mammography (FFDM) is 11-18 kg (108-177 N). This is the first study to investigate the compression force used in the programme.
METHODS: The study included information from 17,951 randomly selected females screened with FFDM at 14 breast centres in the NBCSP, during January-March 2014. We investigated the applied compression force on the left breast in craniocaudal and mediolateral oblique views for breast centres, mammography machines within the breast centres and for the radiographers.
RESULTS: The mean compression force for all mammograms in the study was 116 N and ranged from 91 N to 147 N between the breast centres. The variation in compression force was wider between the breast centres than that between mammography machines (range 137-155 N) and radiographers (95-143 N) within one breast centre. Approximately 59% of the mammograms in the study complied with the recommended range of compression force.
CONCLUSION: A wide variation in applied compression force was observed between the breast centres in the NBCSP. This variation indicates a need for evidence-based recommendations for compression force aimed at optimizing the image quality and individualizing breast compression. Advances in knowledge: There was a wide variation in applied compression force between the breast centres in the NBCSP. The variation was wider between the breast centres than that between mammography machines and radiographers within one breast centre.
引用
- Waade GG, Moshina N, Sebuødegård S, et al. Compression forces used in the Norwegian Breast Cancer Screening Program. Br J Radiol. 2017. PMID: 28102696. doi:10.1259/bjr.20160770
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。