24G侧孔与22G端孔外周静脉导管在MDCT中的选择:一项前瞻性随机研究
前瞻性随机单中心研究纳入180例患者,比较24G侧孔导管与22G端孔导管在MDCT静脉注射对比剂时的安全性、注射压力和对比增强。可评估174例,总体外渗率1.1%;24G侧孔组1例(1.2%),22G端孔组1例(1.1%),差异0.1%(95%CI -3.17%至3.28%,非劣效P=1)。24G侧孔组平均最大压力高于22G端孔组(8.16±0.95 kg/cm² vs 4.79±0.63 kg/cm²,P<0.001),腹主动脉、腹腔干、肠系膜上动脉及胰腺实质强化两组无显著差异。
要点
- 24G侧孔导管与22G端孔导管外渗率无显著差异,非劣效检验成立。
- 24G侧孔导管平均最大注射压力高于22G端孔导管。
- 两组腹主动脉、腹腔干、肠系膜上动脉及胰腺实质强化无显著差异。
核心结论
- 24G侧孔导管安全且适用于以3 mL/s流率注射碘浓度300 mg/mL的对比剂,可能有助于某些患者如静脉脆弱细小者的护理。
研究设计与样本
- 前瞻性随机单中心研究,180例患者随机分配至24G侧孔导管组或22G端孔导管组,主要终点为MDCT静脉注射对比剂的安全性(非劣效分析,下限95%CI大于-10%非劣效界值),次要终点为注射压力和对比增强;安全性分析共174例。
诊断标准与关键数值
- 总体外渗率1.1%(2/174例):24G侧孔组1例(1.2%),22G端孔组1例(1.1%);组间差异0.1%(95%CI -3.17%至3.28%,非劣效P=1)。
临床应用与阅片要点
- 对于静脉脆弱细小的患者,作者认为24G侧孔导管可能有助于护理。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: To compare the 24-gauge side-holes catheter and conventional 22-gauge end-hole catheter in terms of safety, injection pressure, and contrast enhancement on multi-detector computed tomography (MDCT).
MATERIALS & METHODS: In a randomized single-center study, 180 patients were randomized to either the 24-gauge side-holes catheter or the 22-gauge end-hole catheter groups. The primary endpoint was safety during intravenous administration of contrast material for MDCT, using a non-inferiority analysis (lower limit 95% CI greater than -10% non-inferiority margin for the group difference). The secondary endpoints were injection pressure and contrast enhancement.
RESULTS: A total of 174 patients were analyzed for safety during intravenous contrast material administration for MDCT. The overall extravasation rate was 1.1% (2/174 patients); 1 (1.2%) minor episode occurred in the 24-gauge side-holes catheter group and 1 (1.1%) in the 22-gauge end-hole catheter group (difference: 0.1%, 95% CI: -3.17% to 3.28%, non-inferiority P=1). The mean maximum pressure was higher with the 24-gauge side-holes catheter than with the 22-gauge end-hole catheter (8.16±0.95kg/cm2 vs. 4.79±0.63kg/cm2, P<0.001). The mean contrast enhancement of the abdominal aorta, celiac artery, superior mesenteric artery, and pancreatic parenchyma in the two groups were not significantly different.
CONCLUSION: In conclusion, our study showed that the 24-gauge side-holes catheter is safe and suitable for delivering iodine with a concentration of 300mg/mL at a flow-rate of 3mL/s, and it may contribute to the care of some patients, such as patients who have fragile and small veins. (Trial registration: UMIN000023727).
引用
- Tamura A, Kato K, Kamata M, et al. Selection of peripheral intravenous catheters with 24-gauge side-holes versus those with 22-gauge end-hole for MDCT: A prospective randomized study. Eur J Radiol. 2016. PMID: 28065379. doi:10.1016/j.ejrad.2016.12.005
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。