光谱探测器CT虚拟平扫图像:与真实平扫CT衰减值的比较

Spectral detector CT-derived virtual non-contrast images: comparison of attenuation values with unenhanced CT.
PubMed 文献 · Abdom Radiol (NY) · 2017-01-01

前瞻性多中心研究纳入46例患者,在原型双层光谱探测器CT上获得84对平扫与虚拟平扫(VNC)图像,比较肝脏、脾脏、肾脏、腰大肌、腹主动脉和皮下脂肪的衰减值;除皮下脂肪外,VNC与平扫衰减值在<10 HU阈值下等效,且各期相间无显著差异;VNC图像质量在动脉期和肾图期优良率分别为84%和85%,尿路期仅40%。

要点

  • VNC与平扫衰减值在除脂肪外所有组织中于<10 HU阈值下等效。
  • 皮下脂肪VNC高估HU,含脂肪病变的衰减阈值可能需调整。
  • 尿路期VNC图像质量优良率仅40%,低于动脉期和肾图期。

核心结论

  • 双层光谱探测器CT生成的VNC图像在除皮下脂肪外的所有评估组织中衰减值与平扫相似,各期相间无显著差异;作者认为需进一步研究以确定临床常用病理衰减阈值是否应调整,尤其对含脂肪病变。

研究设计与样本

  • HIPAA合规、IRB批准的前瞻性多中心研究,46例患者于2013年10月至2015年11月在原型双层光谱探测器CT上接受平扫和增强扫描,获得84对平扫与VNC图像(动脉期25对、门静脉/肾图期39对、尿路期20对);由三位阅片者之一测量肝脏、脾脏、肾脏、腰大肌、腹主动脉和皮下脂肪的平均衰减值;采用等效性检验判断平扫与VNC平均差异是否小于5、10、15 HU;VNC图像质量采用5分制评估。

诊断标准与关键数值

  • 平扫与VNC平均衰减差异<15 HU占全部测量的92.6%,<10 HU占75.2%,<5 HU占44.4%;采用<10 HU差异阈值时,除脂肪外所有组织平扫与VNC衰减值等效(p<0.05);各期相间无显著差异;皮下脂肪VNC相对于平扫高估HU;VNC图像质量评为优秀或良好者动脉期84%、肾图期85%、尿路期40%。

临床应用与阅片要点

  • 作者认为需进一步研究以确定临床常用病理衰减阈值是否应调整,尤其对含脂肪病变。

局限性

  • 作者结论指出需进一步研究以确定临床常用病理衰减阈值是否应调整,尤其对含脂肪病变。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To assess virtual non-contrast (VNC) images obtained on a detection-based spectral detector CT scanner and determine how attenuation on VNC images derived from various phases of enhanced CT compare to those obtained from true unenhanced images.

METHODS: In this HIPAA compliant, IRB approved prospective multi-institutional study, 46 patients underwent pre- and post-contrast imaging on a prototype dual-layer spectral detector CT between October 2013 and November 2015, yielding 84 unenhanced and VNC pairs (25 arterial, 39 portal venous/nephrographic, 20 urographic). Mean attenuation was measured by one of three readers in the liver, spleen, kidneys, psoas muscle, abdominal aorta, and subcutaneous fat. Equivalence testing was used to determine if the mean difference between unenhanced and VNC attenuation was less than 5, 10, or 15 HU. VNC image quality was assessed on a 5 point scale.

RESULTS: Mean difference between unenhanced and VNC attenuation was <15 HU in 92.6%, <10 HU in 75.2%, and <5 HU in 44.4% of all measurements. Unenhanced and VNC attenuation were equivalent in all tissues except fat using a threshold of <10 HU difference (p < 0.05). No significant variation was seen between phases. In fat, VNC overestimated the HU relative to unenhanced images. VNC image quality was rated as excellent or good in 84% of arterial phase and 85% of nephrographic phase cases, but only 40% of urographic phase.

CONCLUSION: VNC images derived from novel dual layer spectral detector CT demonstrate attenuation values similar to unenhanced images in all tissues evaluated except for subcutaneous fat. Further study is needed to determine if attenuation thresholds currently used clinically for common pathology should be adjusted, particularly for lesions containing fat.

来源:Abdom Radiol (NY),查看原文。版权归原作者所有。

引用

  1. Ananthakrishnan L, Rajiah P, Ahn R, et al. Spectral detector CT-derived virtual non-contrast images: comparison of attenuation values with unenhanced CT. Abdom Radiol (NY). 2017. PMID: 28084546. doi:10.1007/s00261-016-1036-9

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。