肝脏磁共振弹性成像检查的技术失败:一项大型单中心研究的经验
该回顾性单中心研究评估了691例连续患者共781次肝脏MR弹性成像检查,比较3.0 T与1.5 T梯度回波序列的技术失败率。1.5 T失败率为3.5%(12/338),3.0 T为15.3%(68/443)。多因素分析显示体重指数、肝铁沉积、大量腹水和使用3.0 T是与技术失败独立相关的因素。
要点
- 1.5 T肝脏MR弹性成像技术失败率3.5%,3.0 T为15.3%,3.0 T失败率明显更高。
- 体重指数、肝铁沉积、大量腹水和使用3.0 T是多因素分析中与失败独立相关的因素。
- 单因素分析中肝硬化与酒精性肝病也与失败显著相关,但多因素分析后不再显著。
核心结论
- 采用梯度回波脉冲序列的肝脏MR弹性成像技术失败率在1.5 T较低,但在3.0 T明显更高;大量腹水、铁沉积和高体重指数是2D梯度回波序列肝脏MR弹性成像失败的额外独立相关因素。
研究设计与样本
- 回顾性单中心研究,纳入2013年6月至2014年8月间691例连续患者的781次MR弹性成像检查,平均年龄58岁,男性434例(62.8%)。
诊断标准与关键数值
- 技术失败定义为无像素值置信指数高于95%和/或未见明显剪切波成像;1.5 T失败率为3.5%(12/338),3.0 T失败率为15.3%(68/443);单因素分析中体重指数、肝铁沉积、大量腹水、使用3.0 T、肝硬化及酒精性肝病均与失败显著相关(P < .004),多因素分析中仅体重指数、肝铁沉积、大量腹水和使用3.0 T与失败显著相关(P < .004)。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Purpose To assess the determinants of technical failure of magnetic resonance (MR) elastography of the liver in a large single-center study. Materials and Methods This retrospective study was approved by the institutional review board. Seven hundred eighty-one MR elastography examinations performed in 691 consecutive patients (mean age, 58 years; male patients, 434 [62.8%]) in a single center between June 2013 and August 2014 were retrospectively evaluated. MR elastography was performed at 3.0 T (n = 443) or 1.5 T (n = 338) by using a gradient-recalled-echo pulse sequence. MR elastography and anatomic image analysis were performed by two observers. Additional observers measured liver T2* and fat fraction. Technical failure was defined as no pixel value with a confidence index higher than 95% and/or no apparent shear waves imaged. Logistic regression analysis was performed to assess potential predictive factors of technical failure of MR elastography. Results The technical failure rate of MR elastography at 1.5 T was 3.5% (12 of 338), while it was higher, 15.3% (68 of 443), at 3.0 T. On the basis of univariate analysis, body mass index, liver iron deposition, massive ascites, use of 3.0 T, presence of cirrhosis, and alcoholic liver disease were all significantly associated with failure of MR elastography (P < .004); but on the basis of multivariable analysis, only body mass index, liver iron deposition, massive ascites, and use of 3.0 T were significantly associated with failure of MR elastography (P < .004). Conclusion The technical failure rate of MR elastography with a gradient-recalled-echo pulse sequence was low at 1.5 T but substantially higher at 3.0 T. Massive ascites, iron deposition, and high body mass index were additional independent factors associated with failure of MR elastography of the liver with a two-dimensional gradient-recalled-echo pulse sequence. © RSNA, 2017.
引用
- Wagner M, Corcuera-Solano I, Lo G, et al. Technical Failure of MR Elastography Examinations of the Liver: Experience from a Large Single-Center Study. Radiology. 2017. PMID: 28045604. doi:10.1148/radiol.2016160863
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。