使用64排MDCT估算儿童心脏CT血管造影的有效剂量:剂量长度乘积与有效剂量的新转换系数
研究用1岁和10岁儿童体模在64排MDCT上模拟不同心脏扫描模式和管电压,基于ICRP 60和ICRP 103定义计算有效剂量并推导转换系数。ICRP 103定义下1岁和10岁体模的转换系数分别为0.099和0.049 mSv·mGy⁻¹·cm⁻¹,较ICRP 60定义高37%。作者结论提示按ICRP 103和心脏扫描范围估算的儿童心脏CT辐射剂量高于既往使用胸部转换系数的报告值。
要点
- 1岁儿童体模转换系数为0.099 mSv·mGy⁻¹·cm⁻¹,10岁为0.049 mSv·mGy⁻¹·cm⁻¹。
- ICRP 103定义所得转换系数平均比ICRP 60定义高37%。
- 心率变化不影响有效剂量或转换系数。
核心结论
- 基于ICRP 103定义和心脏扫描范围推导的转换系数提示,儿童接受当代64排MDCT心脏CT的辐射剂量高于既往使用旧版胸部转换系数报告的值。
研究设计与样本
- 研究使用代表1岁和10岁儿童的拟人体模,每个体模放置50个金属氧化物半导体场效应晶体管剂量计于器官内,在64排MDCT上采用不同常规临床心脏扫描模式和X射线管电压进行扫描。
诊断标准与关键数值
- 基于ICRP 103定义的有效剂量转换系数:1岁体模为0.099 mSv·mGy⁻¹·cm⁻¹,10岁体模为0.049 mSv·mGy⁻¹·cm⁻¹。
- 上述转换系数平均比基于ICRP 60定义计算的相应转换系数高37%。
- 改变心率未影响有效剂量或转换系数。
局限性
- 作者指出,需为其他当代CT扫描仪和不同年龄范围患者制定更多最新的儿童心脏CT转换系数。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: The purpose of this study is to determine the conversion factors that enable accurate estimation of the effective dose (ED) used for cardiac 64-MDCT angiography performed for children.
MATERIALS AND METHODS: Anthropomorphic phantoms representative of 1- and 10-year-old children, with 50 metal oxide semiconductor field-effect transistor dosimeters placed in organs, underwent scanning performed using a 64-MDCT scanner with different routine clinical cardiac scan modes and x-ray tube potentials. Organ doses were used to calculate the ED on the basis of weighting factors published in 1991 in International Commission on Radiological Protection (ICRP) publication 60 and in 2007 in ICRP publication 103. The EDs and the scanner-reported dose-length products were used to determine conversion factors for each scan mode. The effect of infant heart rate on the ED and the conversion factors was also assessed.
RESULTS: The mean conversion factors calculated using the current definition of ED that appeared in ICRP publication 103 were as follows: 0.099 mSv · mGy-1 · cm-1, for the 1-year-old phantom, and 0.049 mSv · mGy-1 · cm-1, for the 10-year-old phantom. These conversion factors were a mean of 37% higher than the corresponding conversion factors calculated using the older definition of ED that appeared in ICRP publication 60. Varying the heart rate did not influence the ED or the conversion factors.
CONCLUSION: Conversion factors determined using the definition of ED in ICRP publication 103 and cardiac, rather than chest, scan coverage suggest that the radiation doses that children receive from cardiac CT performed using a contemporary 64-MDCT scanner are higher than the radiation doses previously reported when older chest conversion factors were used. Additional up-to-date pediatric cardiac CT conversion factors are required for use with other contemporary CT scanners and patients of different age ranges.
引用
- Trattner S, Chelliah A, Prinsen P, et al. Estimating Effective Dose of Radiation From Pediatric Cardiac CT Angiography Using a 64-MDCT Scanner: New Conversion Factors Relating Dose-Length Product to Effective Dose. AJR Am J Roentgenol. 2017. PMID: 28095022. doi:10.2214/AJR.15.15908
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。