同一患者中碘剂量减少70%的双能CT主动脉造影与标准碘剂量单能CT主动脉造影的前瞻性比较
在21例既往接受标准碘剂量单能CT主动脉造影的患者中,前瞻性比较碘剂量减少70%的双能CT主动脉造影。结果显示双能CT 50 keV与单能CT 120 kVp的主动脉衰减、CNR、SNR相似,主观血管强化评分较低但可接受。
要点
- 碘剂量减少70%的双能CT主动脉造影在50 keV重建时主动脉衰减与标准碘剂量单能CT 120 kVp相似(338 HU vs 350 HU,p=0.547)。
- 双能CT 50 keV的主观血管强化评分为中等至良好(2.54-2.93),低于单能CT的良好至优秀(3.33-3.69,p<0.0001)。
- 双能CT 50 keV与单能CT 120 kVp的SNR和CNR无显著差异(p>0.05)。
核心结论
- 碘剂量减少70%的双能CT主动脉造影与标准碘剂量单能CT主动脉造影相比,可能获得相似的主动脉衰减、CNR、SNR,主观图像评分较低但可接受。
研究设计与样本
- 前瞻性研究,21例既往接受标准碘剂量单能CT主动脉造影的患者,随后接受碘剂量减少70%的双能CT主动脉造影。
- 比较单能CT 120 kVp图像与双能CT 50 keV和77 keV重建图像,测量主动脉8个位置的噪声和衰减,并由阅片者按4分法主观评分血管强化。
关键影像表现/征象
- 双能CT 50 keV的平均主动脉衰减为338±57 HU,与单能CT 120 kVp的350±67 HU相似(p=0.547),77 keV较低为152±23 HU。
- 图像噪声在50 keV最大(12±5 HU),77 keV最低(7±2 HU,p>0.001)。
诊断标准与关键数值
- 主观血管强化评分采用4分法,单能CT评分为3.33-3.69(良好至优秀),双能CT 50 keV评分为2.54-2.93(中等至良好,p<0.0001)。
临床应用与阅片要点
- 双能CT主动脉造影采用碘剂量减少70%时,50 keV重建可提供与单能CT 120 kVp相似的主动脉衰减、SNR和CNR,主观评分虽较低但可接受。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: To compare dual-energy computed tomography (DECT) aortography using a 70% reduced iodine dose to single-energy CT (SECT) aortography using a standard iodine dose in the same patient.
METHODS: Twenty-one patients with a prior SECT aortography using standard iodine dose had DECT aortography using 70% reduced iodine dose. Section 120 kVp images were compared to DECT images reconstructed at both 50 and 77 keV. Reviewers measured image noise and attenuation in the aorta at eight locations from proximal to distal and subjectively scored vascular enhancement on a four-point scale. Signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were calculated. The volume CT dose index (CTDIvol) for each exam was recorded.
RESULTS: Mean iodine dose was 50 g for SECT and 15 g for DECT (70% reduction). Mean aortic attenuation was similar for section 120 kVp (350 ± 67 HU) and DECT 50 keV (338 ± 57 HU, p = 0.547) but was lower at 77 keV (152 ± 23 HU). Measured image noise was greatest at 50 keV (12 ± 5 HU) and was lowest at 77 keV (7 ± 2 HU, p > 0.001). There was no difference in SNR or CNR between 120 kVp and 50 keV (p > 0.05). Mean subjective vascular enhancement scores for SECT were between good and excellent (3.33-3.69), and for DECT at 50 keV were between moderate and good (2.54-2.93, p < 0.0001). CTDIvol was 13.6 mGy for SECT and 13.1 mGy for DECT (p = 0.637).
CONCLUSION: 70% Reduced iodine DECT aortography may result in similar aortic attenuation, CNR, SNR, and lower although acceptable subjective image scores when compared to standard iodine SECT aortography in the same patient.
引用
- Shuman WP, O'Malley RB, Busey JM, et al. Prospective comparison of dual-energy CT aortography using 70% reduced iodine dose versus single-energy CT aortography using standard iodine dose in the same patient. Abdom Radiol (NY). 2017. PMID: 28084544. doi:10.1007/s00261-016-1041-z
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。