经导管主动脉瓣置换术前心电图门控心脏CT血管造影主动脉瓣环的自动与手动测量:与三维经食管超声心动图的比较

Automated and Manual Measurements of the Aortic Annulus with ECG-Gated Cardiac CT Angiography Prior to Transcatheter Aortic Valve Replacement: Comparison with 3D-Transesophageal Echocardiography.
PubMed 文献 · Acad Radiol · 2017-01-24

回顾性分析74例TAVR术前30天内行三维经食管超声心动图与心电图门控心脏CT血管造影的患者,比较瓣环短轴、长轴及面积测量。心脏CT收缩期与舒张期测量高度相关但收缩期测值更大,舒张期面积与三维经食管超声更接近;半自动与手动测量高度相关但面积偏大,个体差异可能影响瓣膜尺寸选择。

要点

  • 心脏CT舒张期(80% R-R间期)瓣环面积与三维经食管超声收缩中期测值几乎一致,收缩期(40% R-R间期)则高估约28 mm²。
  • 半自动软件测得的瓣环面积较手动描记大约11–30 mm²,需目视核查轮廓。
  • 心脏CT与三维经食管超声及手动与半自动测量间的个体面积差异可能足以改变人工瓣膜尺寸选择。

核心结论

  • 心脏CT各测量值与三维经食管超声高度相关,舒张期心脏CT测值更接近标准收缩中期三维经食管超声测值;半自动与手动测量高度相关,但半自动轮廓应目视核查,因个体面积差异可能改变瓣膜尺寸选择。

研究设计与样本

  • 回顾性队列研究,纳入74例在30天内行三维经食管超声和心电图门控心脏CT血管造影用于TAVR术前规划的患者;三维经食管超声于收缩中期测量,心脏CT于收缩末期(R-R间期的40%)和舒张中期(R-R间期的80%)测量,瓣环面积分别用手动描记和半自动软件独立测量。

诊断标准与关键数值

  • 心脏CT收缩期与舒张期短轴直径(r=0.91)、长轴直径(r=0.92)、瓣环面积(r=0.96)高度相关,收缩期测值显著更大(P<0.001),以短轴直径最明显;三维经食管超声与心脏CT短轴直径(r=0.84-0.90)、长轴直径(r=0.77-0.79)、瓣环面积(r=0.89-0.90)相关性良好;与三维经食管超声相比,收缩期心脏CT瓣环面积高估28 mm²(P<0.008),舒张期心脏CT与三维经食管超声几乎一致;半自动与手动心脏CT测量在收缩期(r=0.94)和舒张期(r=0.93)高度相关,但半自动瓣环面积较手动描记大约11-30 mm²;Bland-Altman分析95%一致性界限显示,心脏CT与三维经食管超声间瓣环面积个体差异为-100.9至99.6 mm²,手动与半自动心脏CT测量间为-105.9至45.2 mm²,可能足以改变主动脉人工瓣膜尺寸选择。

临床应用与阅片要点

  • 心脏CT评估主动脉瓣环时舒张期测量更接近标准收缩中期三维经食管超声测值;半自动软件得出的瓣环轮廓应目视核查,因个体面积估计差异可能改变人工瓣膜尺寸选择。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

RATIONALE AND OBJECTIVES: Multimodality evaluation of the aortic annulus is generally advocated to plan for transcatheter aortic valve replacement (TAVR). We compared aortic annular measurements by cardiac computed tomography angiography (cCTA) to three-dimensional transesophageal echocardiography (3D-TEE), and also evaluated the use of semi-automated software for cCTA annular measurements.

MATERIALS AND METHODS: A retrospective cohort of 74 patients underwent 3D-TEE and electrocardiogram-gated cCTA of the heart within 30 days for TAVR planning. 3D-TEE measurements were obtained during mid-systole; cCTA measurements were obtained during late-systole (40% of R-R interval) and mid-diastole (80% of R-R interval). Annular area was measured independently by manual planimetry and with semi-automated software.

RESULTS: cCTA measurements in systole and diastole were highly correlated for short-axis diameter (r = 0.91), long-axis diameter (r = 0.92), and annular area (r = 0.96), although systolic measurements were significantly larger (P < 0.001), most notably for the short-axis diameter. Good correlation was observed between 3D-TEE and cCTA for short-axis diameter (r = 0.84-0.90), long-axis diameter (r = 0.77-0.79), and annular area (r = 0.89-0.90). As compared to 3D-TEE, annular area is overmeasured by 28 mm2 on systolic phase cCTA (P < 0.008), but nearly identical with 3D-TEE on diastolic phase cCTA. Semi-automated and manual cCTA annulus measurements were highly correlated in systole (r = 0.94) and diastole (r = 0.93), although the semi-automated annular area measured 11-30 mm2 greater than manual planimetry. Of note, the 95% limits of agreement in our Bland-Altman analysis suggest that the variability in annular area estimates for individual patients between cCTA and 3D-TEE (-100.9 to 99.6 mm2), as well as the variability between manual and automated measurements with cCTA (-105.9 to 45.2 mm2), may be sufficient to alter size selection for an aortic prosthesis.

CONCLUSIONS: Although all cCTA measurements are highly correlated with measurements by 3D-TEE, diastolic phase cCTA measurements tend to be closer to standard mid-systolic 3D-TEE measurements. Semi-automated measurement of the aortic annulus with cCTA is highly correlated with manual planimetry. Nonetheless, annular contours derived by semi-automated software should be visually inspected, as the variability in area estimates for individual cases between manual and automated measurements may alter the sizing of an aortic prosthesis.

来源:Acad Radiol,查看原文。版权归原作者所有。

引用

  1. Guez D, Boroumand G, Ruggiero NJ, et al. Automated and Manual Measurements of the Aortic Annulus with ECG-Gated Cardiac CT Angiography Prior to Transcatheter Aortic Valve Replacement: Comparison with 3D-Transesophageal Echocardiography. Acad Radiol. 2017. PMID: 28130049. doi:10.1016/j.acra.2016.12.008