比较基于模型的迭代重建与滤波反投影算法对亚实性肺结节软件测量的影响
研究对73个亚实性结节(SSNs)的平扫CT分别用FBP和MBIR重建,由两名放射科医师测量结节及其实性成分的直径、平均衰减、质量和体积。MBIR测得的最长径、体积和质量显著高于FBP(p<0.05),但差异均在观察者内和观察者间测量变异范围内;两种算法间观察者内和观察者间重复性无显著差异(p>0.05);半自动分类一致性(纯磨玻璃 vs 部分实性)在两种算法间无显著差异。
要点
- MBIR测得的SSNs最长径、体积、质量及其实性成分相应值均显著高于FBP(p<0.05)。
- 两种算法间测量差异均在观察者内和观察者间变异范围内,作者认为无临床显著意义。
- 半自动测量可不受重建算法影响,纯磨玻璃与部分实性的分类一致性在两种算法间无显著差异。
核心结论
- SSNs的半自动测量在FBP与MBIR间存在显著差异,但差异在测量变异范围内;两种算法的观察者内和观察者间重复性无显著差异,半自动分类一致性也无显著差异。
研究设计与样本
- 对73个SSNs的平扫CT采用同一扫描方案、分别以FBP和MBIR重建,由两名放射科医师测量结节及实性成分的直径、平均衰减、质量和体积,用Bland-Altman法和Wilcoxon检验评估观察者内/间变异及两种算法间差异。
诊断标准与关键数值
- MBIR较FBP:结节最长径平均差异1.1%(一致性界限-6.4%至8.5%),体积3.2%(-20.9%至27.3%),质量2.9%(-16.9%至22.7%)。
- 实性成分MBIR较FBP:最长径平均差异3.2%(-20.5%至27%),体积6.3%(-51.9%至64.6%),质量6.6%(-50.1%至63.3%)。
- FBP与MBIR间一致性界限均在两种算法观察者内和观察者间变异范围内(直径、体积、质量及实性成分相应值)。
- 观察者内和观察者间变异在FBP与MBIR间无显著差异(p>0.05)。
- SSNs半自动分类(纯磨玻璃 vs 部分实性)一致性在两种算法间无显著差异。
临床应用与阅片要点
- 作者认为SSNs半自动测量可不受重建算法影响进行。
- 作者认为重建算法引起的SSNs半自动测量差异无临床显著意义。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: To evaluate the differences between filtered back projection (FBP) and model-based iterative reconstruction (MBIR) algorithms on semi-automatic measurements in subsolid nodules (SSNs).
METHODS: Unenhanced CT scans of 73 SSNs obtained using the same protocol and reconstructed with both FBP and MBIR algorithms were evaluated by two radiologists. Diameter, mean attenuation, mass and volume of whole nodules and their solid components were measured. Intra- and interobserver variability and differences between FBP and MBIR were then evaluated using Bland-Altman method and Wilcoxon tests.
RESULTS: Longest diameter, volume and mass of nodules and those of their solid components were significantly higher using MBIR (p < 0.05) with mean differences of 1.1% (limits of agreement, -6.4 to 8.5%), 3.2% (-20.9 to 27.3%) and 2.9% (-16.9 to 22.7%) and 3.2% (-20.5 to 27%), 6.3% (-51.9 to 64.6%), 6.6% (-50.1 to 63.3%), respectively. The limits of agreement between FBP and MBIR were within the range of intra- and interobserver variability for both algorithms with respect to the diameter, volume and mass of nodules and their solid components. There were no significant differences in intra- or interobserver variability between FBP and MBIR (p > 0.05).
CONCLUSION: Semi-automatic measurements of SSNs significantly differed between FBP and MBIR; however, the differences were within the range of measurement variability.
KEY POINTS: • Intra- and interobserver reproducibility of measurements did not differ between FBP and MBIR. • Differences in SSNs' semi-automatic measurement induced by reconstruction algorithms were not clinically significant. • Semi-automatic measurement may be conducted regardless of reconstruction algorithm. • SSNs' semi-automated classification agreement (pure vs. part-solid) did not significantly differ between algorithms.
引用
- Cohen JG, Kim H, Park SB, et al. Comparison of the effects of model-based iterative reconstruction and filtered back projection algorithms on software measurements in pulmonary subsolid nodules. Eur Radiol. 2017. PMID: 28058482. doi:10.1007/s00330-016-4716-5
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。