CTDIvol低于1 mGy时采用迭代重建技术的胸部CT评估
研究纳入25例患者,在标准护理胸部CT(8 mGy)后采集0.25、0.4和0.8 mGy三档低剂量胸部CT,分别用SafeCT、ASIR和MBIR重建。182个病灶中,0.25、0.4、0.8 mGy分别漏诊35、24、9个;非钙化肺结节漏诊最多,0.25和0.4 mGy时漏诊<5 mm结节25/112和18/112,0.8 mGy时漏诊<4 mm结节3/112;三种迭代重建技术对肺结节检出性能相似,0.8 mGy时检出97%肺结节;0.25-0.8 mGy时多数纵隔结构显示不充分。
要点
- CTDIvol<1 mGy时非钙化肺结节可被漏诊,且与所用迭代重建技术无关。
- 0.25、0.4和0.8 mGy分别漏诊35、24和9个病灶,0.8 mGy检出肺结节比例最高(97%)。
- 0.25-0.8 mGy时多数纵隔结构显示不充分。
核心结论
- CTDIvol<1 mGy(0.25、0.4和0.8 mGy)时非钙化肺结节可被漏诊,与迭代重建技术无关;0.8 mGy时检出多数肺结节(97%);0.25-0.8 mGy时多数纵隔结构显示不充分。
研究设计与样本
- 经机构审查委员会批准的研究,25例患者签署书面知情同意,在64通道多排CT上先完成标准护理CT(8 mGy),再采集0.25、0.4和0.8 mGy三档低剂量胸部CT,分别用SafeCT、ASIR和MBIR重建;由两名委员会认证的胸部放射科医师先评估最低到最高剂量的低剂量CT系列,随后评估标准护理CT。
关键影像表现/征象
- 182个检出病灶中,0.25、0.4和0.8 mGy分别漏诊35、24和9个病灶。
- 漏诊最多的为非钙化肺结节:0.25 mGy时<5 mm结节漏诊25/112,0.4 mGy时<5 mm结节漏诊18/112,0.8 mGy时<4 mm结节漏诊3/112。
- 0.25、0.4和0.8 mGy时肺结节检出率分别为78%、84%和97%,与所用迭代重建技术无关。
- 0.25-0.8 mGy时多数纵隔结构显示不充分。
诊断标准与关键数值
- 0.25 mGy时肺结节检出率78%,非钙化肺结节<5 mm漏诊25/112。
- 0.4 mGy时肺结节检出率84%,非钙化肺结节<5 mm漏诊18/112。
- 0.8 mGy时肺结节检出率97%,非钙化肺结节<4 mm漏诊3/112。
- ASIR、SafeCT和MBIR在0.25、0.4和0.8 mGy时对肺结节检出性能相似。
临床应用与阅片要点
- 在CTDIvol<1 mGy的胸部CT中,需注意非钙化肺结节可能漏诊,且与迭代重建技术无关。
- 0.8 mGy时肺结节检出率最高(97%),0.25和0.4 mGy时检出率分别为78%和84%。
- 0.25-0.8 mGy时多数纵隔结构显示不充分。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: To assess the image quality of chest CT reconstructed with image-based iterative reconstruction (SafeCT; MedicVision®, Tirat Carmel, Israel), adaptive statistical iterative reconstruction (ASIR; GE Healthcare, Waukesha, WI) and model-based iterative reconstruction (MBIR; GE Healthcare, Waukesha, WI) techniques at CT dose index volume (CTDIvol) <1 mGy.
METHODS: In an institutional review board-approved study, 25 patients gave written informed consent for acquisition of three reduced dose (0.25-, 0.4- and 0.8-mGy) chest CT after standard of care CT (8 mGy) on a 64-channel multidetector CT (MDCT) and reconstructed with SafeCT, ASIR and MBIR. Two board-certified thoracic radiologists evaluated images from the lowest to the highest dose of the reduced dose CT series and subsequently for standard of care CT.
RESULTS: Out of the 182 detected lesions, the missed lesions were 35 at 0.25, 24 at 0.4 and 9 at 0.8 mGy with SafeCT, ASIR and MBIR, respectively. The most missed lesions were non-calcified lung nodules (NCLNs) 25/112 (<5 mm) at 0.25, 18/112 (<5 mm) at 0.4 and 3/112 (<4 mm) at 0.8 mGy. There were 78%, 84% and 97% lung nodules detected at 0.25, 0.4 and 0.8 mGy, respectively regardless of iterative reconstruction techniques (IRTs), Most mediastinum structures were not sufficiently seen at 0.25-0.8 mGy.
CONCLUSION: NCLNs can be missed in chest CT at CTDIvol of <1 mGy (0.25, 0.4 and 0.8 mGy) regardless of IRTs. The most lung nodules (97%) were detected at CTDIvol of 0.8 mGy. The most mediastinum structures were not sufficiently seen at 0.25-0.8 mGy. Advances in knowledge: NCLNs can be missed regardless of IRTs in chest CT at CTDIvol of <1 mGy. The performance of ASIR, SafeCT and MBIR was similar for lung nodule detection at 0.25, 0.4 and 0.8 mGy.
引用
- Padole A, Digumarthy S, Flores E, et al. Assessment of chest CT at CTDIvol less than 1 mGy with iterative reconstruction techniques. Br J Radiol. 2017. PMID: 28055250. doi:10.1259/bjr.20160625