实时前瞻性运动校正(PROMO)在3D T1加权成像脑皮质分割中的效用:针对不配合患者的基于体素形态学分析
研究纳入33例不配合患者(帕金森病11例、痴呆22例),对比有/无PROMO的3D T1WI在VBM分析中的灰质体积与皮质厚度差异。结果显示无PROMO扫描的总灰质体积显著小于有PROMO扫描(p<0.05),且无PROMO扫描全脑VBM分析显示部分区域灰质体积显著减少(FWE校正p<0.05),皮质厚度分析亦显示无PROMO扫描皮质厚度减低(p<0.05)。结论提示PROMO可减少VBM分割中的误分类,但可能无法阻止灰质体积减少。
要点
- 运动伪影是VBM分析的重要问题,PROMO可减少高分辨率3D T1WI的运动伪影。
- 不配合患者中,无PROMO扫描的总灰质体积显著小于有PROMO扫描(p<0.05),源于采集时运动导致的分割问题。
- PROMO可能提高VBM分析的精确性,但可能无法阻止灰质体积减少。
核心结论
- 在不配合患者中,使用PROMO可减少VBM分析分割过程中的误分类,但可能无法阻止灰质体积减少。
研究设计与样本
- 在33例不配合患者(帕金森病11例、痴呆22例)中,分别进行有和无PROMO的高分辨率3D T1加权成像,比较两种扫描的灰质体积和皮质厚度。
关键影像表现/征象
- 无PROMO扫描的总灰质体积显著小于有PROMO扫描(p<0.05),由采集时运动导致的分割问题所致。
- 全脑VBM分析显示,无PROMO扫描在部分区域出现显著灰质体积减少(FWE校正p<0.05)。
- 皮质厚度分析中,无PROMO扫描较有PROMO扫描显示皮质厚度减低(p<0.05)。
临床应用与阅片要点
- 对不配合患者进行VBM分析时,使用PROMO可减少运动伪影导致的分割误分类。
- 作者指出PROMO可能提高VBM分析的精确性,但可能无法阻止灰质体积减少。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: To assess the utility of the motion correction method with prospective motion correction (PROMO) in a voxel-based morphometry (VBM) analysis for 'uncooperative' patient populations.
METHODS: High-resolution 3D T1-weighted imaging both with and without PROMO were performed in 33 uncooperative patients with Parkinson's disease (n = 11) or dementia (n = 22). We compared the grey matter (GM) volumes and cortical thickness between the scans with and without PROMO.
RESULTS: For the mean total GM volume with the VBM analysis, the scan without PROMO showed a significantly smaller volume than that with PROMO (p < 0.05), which was caused by segmentation problems due to motion during acquisition. The whole-brain VBM analysis showed significant GM volume reductions in some regions in the scans without PROMO (familywise error corrected p < 0.05). In the cortical thickness analysis, the scans without PROMO also showed decreased cortical thickness compared to the scan with PROMO (p < 0.05).
CONCLUSION: Our results with the uncooperative patients indicate that the use of PROMO can reduce misclassification during segmentation of the VBM analyses, although it may not prevent GM volume reduction.
KEY POINTS: • Motion artifacts pose significant problems for VBM analyses. • PROMO correction can reduce the motion artifacts in high-resolution 3D T1WI. • The use of PROMO may improve the precision of VBM analyses.