扫描主动脉内支架或机械心脏瓣膜患者时是否应关闭CT自动曝光控制?一项体模研究

Should the automatic exposure control system of CT be disabled when scanning patients with endoaortic stents or mechanical heart valves? A phantom study.
PubMed 文献 · Eur Radiol · 2016-12-12

该体模研究扫描7枚支架和2个瓣膜,共540次扫描,比较自动曝光控制(AEC)开启与关闭及有无假体时的剂量长度乘积(DLP)。AEC开启与关闭差异无统计学意义(p≥0.059),假体对AEC影响可忽略,关闭AEC不推荐。

要点

  • AEC开启与关闭的DLP差异无统计学意义(p≥0.059)。
  • 假体对AEC影响可忽略,仅120 kVp时DLP轻度增加。
  • 不推荐因主动脉内假体或机械心脏瓣膜而关闭AEC。

核心结论

  • 假体对AEC系统的影响可忽略且无临床相关性,关闭AEC不太可能带来获益,不推荐关闭AEC。

研究设计与样本

  • 体模研究,扫描7枚支架和2个瓣膜,管电压80/100/120 kVp,AEC开启/关闭及假体存在/不存在组合,共540次扫描。

诊断标准与关键数值

  • AEC开启与关闭的DLP差异无统计学意义(p≥0.059);无假体时辐射剂量1.8 mGy至2.4 mGy,有假体时1.8 mGy至2.5 mGy(置信水平37-100%)。

临床应用与阅片要点

  • 扫描携带主动脉内支架或机械心脏瓣膜的患者时,不推荐关闭AEC;仅120 kVp时可能观察到可忽略的DLP轻度增加。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To estimate the impact of endoaortic stents/mechanical heart valves on the output of an automatic exposure control (AEC) system and CT radiation dose.

METHODS: In this phantom study, seven stents and two valves were scanned with varying tube voltage (80/100/120 kVp), AEC activation (enabled/disabled) and prosthesis (present/absent), for a total of 540 scans. For each prosthesis, the dose-length product (DLP) was compared between scans with the AEC enabled and disabled. Percentage confidence levels for differences due to the prosthesis were calculated.

RESULTS: Differences between results with the AEC enabled and disabled were not statistically significant (p ≥ 0.059). In the comparison with and without the prosthesis, DLP was unchanged at 80 kVp and 100 kVp, while a slight increase was observed at 120 kVp. The radiation dose varied from 1.8 mGy to 2.4 mGy without the prosthesis and from 1.8 mGy to 2.5 mGy with the prosthesis (confidence level 37-100%).

CONCLUSIONS: The effect of the prosthesis on the AEC system was negligible and not clinically relevant. Therefore, disabling the AEC system when scanning these patients is not likely to provide a benefit.

KEY POINTS: • CT-AEC system is not impaired in patients with endoaortic prostheses/heart valves. • Negligible differences may be observed only at 120 kVp. • Disabling the AEC system in these patients is not recommended.

来源:Eur Radiol,查看原文。版权归原作者所有。

引用

  1. Di Leo G, Spadavecchia C, Zanardo M, et al. Should the automatic exposure control system of CT be disabled when scanning patients with endoaortic stents or mechanical heart valves? A phantom study. Eur Radiol. 2016. PMID: 27957643. doi:10.1007/s00330-016-4676-9

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。