青少年头部CT采用100 kVp联合正弦图确认迭代重建算法降低辐射剂量:对灰白质对比度和图像噪声的影响

Radiation dose reduction using 100-kVp and a sinogram-affirmed iterative reconstruction algorithm in adolescent head CT: Impact on grey-white matter contrast and image noise.
PubMed 文献 · Eur Radiol · 2016-12-13

回顾性纳入69例12–17岁青少年头部CT,比较120 kVp(34例)与100 kVp(35例)协议,100 kVp图像分别用滤波反投影和正弦图确认迭代重建。100 kVp联合迭代重建的有效剂量较120 kVp降低46%,幕上灰白质对比度更高、噪声更低,其余主观指标无显著差异。

要点

  • 100 kVp联合正弦图确认迭代重建使青少年头部CT有效剂量较120 kVp降低46%。
  • 该方案幕上灰白质对比度客观值和主观评分均优于120 kVp,图像噪声更低。
  • 除灰白质对比度外,其余主观图像质量指标与120 kVp无显著差异。

核心结论

  • 青少年头部CT采用100 kVp联合正弦图确认迭代重建可在不增加图像噪声的情况下大幅降低辐射剂量并改善灰白质对比度。

研究设计与样本

  • 回顾性研究,纳入69例12–17岁患者,34例采用120 kVp协议、35例采用100 kVp协议;120 kVp图像用滤波反投影重建,100 kVp图像分别用滤波反投影和正弦图确认迭代重建重建;比较有效剂量、灰白质对比度、图像噪声和对比噪声比。

关键影像表现/征象

  • 100 kVp联合迭代重建的幕上灰白质对比度高于120 kVp,图像噪声低于120 kVp(p < 0.001)。
  • 100 kVp联合迭代重建的对比噪声比高于120 kVp(p < 0.001)。
  • 100 kVp联合迭代重建的灰白质对比度主观评分优于120 kVp(p < 0.001)。

诊断标准与关键数值

  • 有效剂量:100 kVp较120 kVp降低46%(p < 0.001)。
  • 主观图像质量采用四点量表评估灰白质对比度、图像噪声、锐利度、伪影和整体图像质量。
  • 100 kVp联合迭代重建与120 kVp在除灰白质对比度外的其他主观指标上无显著差异(p = 0.072–0.966)。

临床应用与阅片要点

  • 青少年头部CT可考虑采用100 kVp联合正弦图确认迭代重建,以在降低辐射剂量的同时维持或改善图像质量。
  • 迭代重建对噪声的抑制在幕上区域尤为明显。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To retrospectively evaluate the image quality and radiation dose of 100-kVp scans with sinogram-affirmed iterative reconstruction (IR) for unenhanced head CT in adolescents.

METHODS: Sixty-nine patients aged 12-17 years underwent head CT under 120- (n = 34) or 100-kVp (n = 35) protocols. The 120-kVp images were reconstructed with filtered back-projection (FBP), 100-kVp images with FBP (100-kVp-F) and sinogram-affirmed IR (100-kVp-S). We compared the effective dose (ED), grey-white matter (GM-WM) contrast, image noise, and contrast-to-noise ratio (CNR) between protocols in supratentorial (ST) and posterior fossa (PS). We also assessed GM-WM contrast, image noise, sharpness, artifacts, and overall image quality on a four-point scale.

RESULTS: ED was 46% lower with 100- than 120-kVp (p < 0.001). GM-WM contrast was higher, and image noise was lower, on 100-kVp-S than 120-kVp at ST (p < 0.001). CNR of 100-kVp-S was higher than of 120-kVp (p < 0.001). GM-WM contrast of 100-kVp-S was subjectively rated as better than of 120-kVp (p < 0.001). There were no significant differences in the other criteria between 100-kVp-S and 120-kVp (p = 0.072-0.966).

CONCLUSIONS: The 100-kVp with sinogram-affirmed IR facilitated dramatic radiation reduction and better GM-WM contrast without increasing image noise in adolescent head CT.

KEY POINTS: • 100-kVp head CT provides 46% radiation dose reduction compared with 120-kVp. • 100-kVp scanning improves subjective and objective GM-WM contrast. • Sinogram-affirmed IR decreases head CT image noise, especially in supratentorial region. • 100-kVp protocol with sinogram-affirmed IR is suited for adolescent head CT.

来源:Eur Radiol,查看原文。版权归原作者所有。

引用

  1. Nagayama Y, Nakaura T, Tsuji A, et al. Radiation dose reduction using 100-kVp and a sinogram-affirmed iterative reconstruction algorithm in adolescent head CT: Impact on grey-white matter contrast and image noise. Eur Radiol. 2016. PMID: 27966043. doi:10.1007/s00330-016-4679-6

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。