低能量CT成像对阳性口服对比剂浓度选择的影响
研究纳入100例接受阳性口服对比剂增强CT的患者,比较120 kVp单能CT与100 kVp或40-70 keV双能CT重建图像上胃肠道管腔的衰减差异,并通过体模实验评估不同稀释浓度对比剂在低能量扫描下的衰减表现。结果显示低能量成像可提高管腔衰减,允许将口服对比剂浓度降低25%-75%。
要点
- 低能量CT(100 kVp或40-60 keV)较120 kVp显著提高阳性口服对比剂胃肠道管腔衰减。
- 低kVp和低keV图像上,25%和50%稀释对比剂溶液可提供与120 kVp标准浓度相似或更高的衰减。
- 低能量CT成像允许将阳性口服对比剂浓度降低25%-75%。
核心结论
- 低能量CT成像增加阳性口服对比剂充盈的胃肠道管腔衰减,允许将口服对比剂浓度降低25%-75%。
研究设计与样本
- 纳入100例患者(38男62女,年龄62±11岁,BMI 26±5),A组50例为120 kVp单能CT后行100 kVp采集,B组50例为快速kV切换单源双能CT的40-70 keV重建;另用胃肠道体模对三种标准剂量口服对比剂及其75%、50%、25%稀释液行单能CT(120、100、80 kVp)和双能CT(80/140 kVp)扫描。
诊断标准与关键数值
- 100 kVp图像管腔衰减328 HU、40-60 keV图像410-924 HU,均高于120 kVp的298 HU(p<0.05);体模中管腔衰减与X射线能量呈负相关,低kVp图像最高增加至527 HU、低keV图像最高增加至999 HU(p<0.05);25%和50%稀释口服对比剂溶液在低kVp和低keV下分别提供与120 kVp标准浓度相似或更高的衰减。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: To determine to what extent low-energy CT imaging affects attenuation of gastrointestinal tract (GIT) opacified with positive oral contrast media (OCM). Second, to establish optimal OCM concentrations for low-energy diagnostic CT exams.
METHODS: One hundred patients (38 men and 62 women; age 62 ± 11 years; BMI 26 ± 5) with positive OCM-enhanced 120-kVp single-energy CT (SECT), and follow-up 100-kVp acquisitions (group A; n = 50), or 40-70-keV reconstructions from rapid kV switching-single-source dual-energy CT (ssDECT) (group B; n = 50) were included. Luminal attenuation from different GIT segments was compared between exams. Standard dose of three OCM and diluted solutions (75%, 50%, and 25% concentrations) were introduced serially in a gastrointestinal phantom and scanned using SECT (120, 100, and 80 kVp) and DECT (80/140 kVp) acquisitions on a ssDECT scanner. Luminal attenuation was obtained on SECT and DECT images (40-70 keV), and compared to 120-kVp scans with standard OCM concentrations.
RESULTS: Luminal attenuation was higher on 100-kVp (328 HU) and on 40-60-keV images (410-924 HU) in comparison to 120-kVp scans (298 HU) in groups A and B (p < 0.05). Phantom: There was an inverse correlation between luminal attenuation and X-ray energy, increasing up to 527 HU on low-kVp and 999 HU on low-keV images (p < 0.05). 25% and 50% diluted OCM solutions provided similar or higher attenuation than 120 kVp, at low kVp and keV, respectively.
CONCLUSIONS: Low-energy CT imaging increases the attenuation of GIT opacified with positive OCM, permitting reduction of 25%-75% OCM concentration.
引用
- Patino M, Murcia DJ, Iamurri AP, et al. Impact of low-energy CT imaging on selection of positive oral contrast media concentration. Abdom Radiol (NY). 2017. PMID: 27933477. doi:10.1007/s00261-016-0993-3
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。