文献速递腹部技术与安全CT

光谱光子计数CT:双对比剂K边缘结肠成像的初步体验

Spectral Photon-counting CT: Initial Experience with Dual-Contrast Agent K-Edge Colonography.
PubMed 文献 · Radiology · 2016-12-02

研究在定制结肠体模中评估光谱光子计数CT利用钆的K边缘区分钆基与碘基对比剂的可行性;常规CT上两种对比剂吸收值均约110 HU,而对比剂图清晰区分分布,钆仅见于息肉、碘位于结肠腔,定量浓度与实际配制混合物吻合良好。

要点

  • 在体模研究中,常规CT无法区分碘与钆对比剂,两者吸收值均约110 HU。
  • 光谱光子计数CT通过多能箱信息行物质分解,可分别生成碘图与钆图。
  • 钆仅见于模拟息肉的胶囊、碘位于结肠腔,定量浓度与实际混合物吻合良好。

核心结论

  • 采用碘充盈肠腔、钆标记息肉的双对比光谱光子计数CT结肠成像可能有助于区分息肉与标记粪便物质。

研究设计与样本

  • 体模可行性研究:定制结肠体模内充填非离子型碘对比剂,结肠壁上放置代表对比剂增强息肉的钆填充胶囊,用临床前光谱光子计数CT系统扫描获取光谱与常规数据,通过多能箱光谱信息行物质分解生成碘图与钆图,并在肠腔与息肉内行定量测量。

关键影像表现/征象

  • 在常规CT层面上,两种对比剂的吸收值相似,均约为110 HU。
  • 对比剂图清晰区分分布:钆仅见于息肉,碘位于结肠腔。
  • 肠腔与息肉内对比剂浓度的定量测量结果与实际配制的混合物吻合良好。

临床应用与阅片要点

  • 在体模条件下,光谱光子计数CT的物质分解图可区分钆标记息肉与碘充盈肠腔。
  • 常规CT上碘与钆吸收值均约110 HU,单凭常规CT层面无法区分两者。
  • 定量测量肠腔与息肉内碘、钆浓度与实际配制混合物吻合良好。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

Purpose To investigate the feasibility of using spectral photon-counting computed tomography (CT) to differentiate between gadolinium-based and nonionic iodine-based contrast material in a colon phantom by using the characteristic k edge of gadolinium. Materials and Methods A custom-made colon phantom was filled with nonionic iodine-based contrast material, and a gadolinium-filled capsule representing a contrast material-enhanced polyp was positioned on the colon wall. The colon phantom was scanned with a preclinical spectral photon-counting CT system to obtain spectral and conventional data. By fully using the multibin spectral information, material decomposition was performed to generate iodine and gadolinium maps. Quantitative measurements were performed within the lumen and polyp to quantitatively determine the absolute content of iodine and gadolinium. Results In a conventional CT section, absorption values of both contrast agents were similar at approximately 110 HU. Contrast material maps clearly differentiated the distributions, with gadolinium solely in the polyp and iodine in the lumen of the colon. Quantitative measurements of contrast material concentrations in the colon and polyp matched well with those of actual prepared mixtures. Conclusion Dual-contrast spectral photon-counting CT colonography with iodine-filled lumen and gadolinium-tagged polyps may enable ready differentiation between polyps and tagged fecal material. © RSNA, 2016.

来源:Radiology,查看原文。版权归原作者所有。

引用

  1. Muenzel D, Bar-Ness D, Roessl E, et al. Spectral Photon-counting CT: Initial Experience with Dual-Contrast Agent K-Edge Colonography. Radiology. 2016. PMID: 27918709. doi:10.1148/radiol.2016160890

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。