压力标准化乳腺X线摄影不影响稳定病灶的可见性、对比度和锐利度

Pressure-standardised mammography does not affect visibility, contrast and sharpness of stable lesions.
PubMed 文献 · Eur J Radiol · 2016-11-28

研究比较2009年18 daN目标压力方案与2014/2015年10 kPa压力标准化方案下188个稳定病灶的乳腺X线图像,4名放射科医师盲法评价。结果显示压力标准化方案压迫力更轻、乳房稍厚、剂量更低,可见性、对比度和锐利度的非劣效性均成立,非首选图像中仅5幅被认为质量不足。

要点

  • 10 kPa压力标准化方案对稳定病灶的可见性、对比度和锐利度非劣效于18 daN目标压力方案。
  • 压力标准化方案压迫力平均降低17%(头尾位)至29%(内外斜位),乳房厚度平均增加2.4%(1.4 mm)。
  • 60幅非首选图像中仅5幅被多位观察者认为质量不足,其中4幅来自18 daN方案、1幅来自10 kPa方案。

核心结论

  • 与疼痛耐受限制的18 daN目标压力压迫相比,10 kPa压力标准化压迫的减痛乳腺X线摄影对稳定病灶的可见性、对比度和锐利度均为非劣效。

研究设计与样本

  • 放射科医师盲法比较研究,4名有经验的放射科医师比较2009年18 daN目标压力方案与2014/2015年10 kPa(75 mmHg)压力标准化方案下可见的188个稳定病灶和结构。

诊断标准与关键数值

  • 非劣效性判定标准为4名阅片者平均的「新方案更好」比例的95%置信区间下限超过非劣效界值0.463;可见性下限0.479、对比度下限0.473、锐利度下限0.488、偏好下限0.486,均超过非劣效界值。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

INTRODUCTION: A recent technological development allows pressure-standardised mammography by personalizing the compression force to the breast size and firmness. The technique has been shown to reduce pain and compression variability between consecutive exams, but also results in a slightly thicker compressed breast during exposure. This raises the question whether visibility, contrast and sharpness of lesions are affected?

METHODS: Four experienced radiologists compared 188 stable lesions and structures including (clusters of) calcifications, (oil) cysts and lymph nodes that were visible in mammograms obtained in 2009 with a pain-tolerance limited 18 daN target force compression protocol, and in 2014/2015 obtained with a 10kPa (75mmHg) pressure-standardised compression protocol. Observers were blinded for all DICOM metadata and rated which of the randomly ordered, side by side presented images had better lesion visibility, contrast and sharpness, or whether they saw no difference. They also indicated which overall image they preferred, if any, and whether the non-preferred image was still adequate. Statistical non-inferiority is concluded when the lower limit of the 95% confidence interval of the 4-rater averaged 'new protocol better' proportions exceed the non-inferiority limit of 0.463.

RESULTS: In 2014/2015, the compressions were significantly milder, with on average 17% (mediolateral oblique) to 29% (craniocaudal) lower forces. Breasts remained on average 2.4% (1.4mm) thicker. Dose was significantly lower (6.5%), which is explained by glandular atrophy. The 95% confidence interval lower limits are 0.479 for visibility, 0.473 for contrast, 0.488 for sharpness and 0.486 for preference, all exceeding the non-inferiority limit. Of the 60 non-preferred mammograms, multiple observers found only five to be inadequate: 4 obtained with the force protocol and 1 with the pressure protocol.

CONCLUSION: Pain-reduced mammography with 10kPa pressure-standardised compression has non-inferior visibility, contrast and sharpness for stable lesions compared to pain-tolerance limited 18daN target force compression.

来源:Eur J Radiol,查看原文。版权归原作者所有。

引用

  1. de Groot JE, Hopman IGM, van Lier MGJTB, et al. Pressure-standardised mammography does not affect visibility, contrast and sharpness of stable lesions. Eur J Radiol. 2016. PMID: 28027762. doi:10.1016/j.ejrad.2016.11.030

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。