3.0 T 乳腺扩散加权成像多 b 值图像质量的比较与优化

Comparison and Optimization of 3.0 T Breast Images Quality of Diffusion-Weighted Imaging with Multiple B-Values.
PubMed 文献 · Acad Radiol · 2016-12-07

本研究纳入 96 例合格患者共 112 个病灶及 14 例患者对侧 14 个正常乳腺,在 3.0 T 下用单次激发平面回波序列行 13 个 b 值的扩散加权成像,测量正常乳腺与良恶性病灶的信噪比、信号强度比和对比噪声比。结果显示 b 值 600–1200 s/mm2 图像质量较高、1000–1200 s/mm2 最佳,信号强度比在 b=1200 s/mm2 时鉴别良恶性病灶的诊断效能最高。

要点

  • b 值 600–1200 s/mm2 时乳腺扩散加权成像对比噪声比更高,1000–1200 s/mm2 最佳。
  • 正常乳腺及良恶性病灶信噪比随 b 值升高而下降,呈负相关。
  • 信号强度比在 b=1200 s/mm2 时鉴别良恶性病灶的诊断效能最高。

核心结论

  • 乳腺扩散加权成像在 b 值 600–1200 s/mm2 时图像质量较高,在 1000–1200 s/mm2 时最佳。
  • 信号强度比在 b=1200 s/mm2 时鉴别良恶性病灶的诊断效能最高。

研究设计与样本

  • 本研究为影像学研究,纳入 96 例合格患者共 112 个病灶及 14 例患者对侧 14 个正常乳腺。
  • 采用单次激发平面回波成像序列行扩散加权成像,共使用 13 个 b 值:0、50、100、100、200、400、600、800、1000、1200、1500、1800、2000 和 2500 s/mm2。

关键影像表现/征象

  • b 值 600–1200 s/mm2 时乳腺扩散加权成像对比噪声比更高。
  • 良性病灶在 b=1000 s/mm2 时对比噪声比最佳,恶性病灶在 b=1200 s/mm2 时对比噪声比最佳。
  • 正常乳腺信号强度和信噪比随 b 值升高而下降,呈负相关(r=-0.945,P<0.01)。
  • 良性病灶平均信噪比呈负相关(r=-0.982,P<0.01),恶性病灶平均信噪比呈负相关(r=-0.947,P<0.01),均随 b 值升高逐渐下降。
  • 良性病灶平均信号强度比随 b 值升高逐渐下降,呈负相关(r=-0.991,P<0.01);恶性病灶平均信号强度比在 b 值 0–1200 s/mm2 之间随 b 值升高逐渐升高,b 值≥1500 s/mm2 时随 b 值升高逐渐下降。

诊断标准与关键数值

  • b 值 600、800、1000 和 1200 s/mm2 时,信号强度比鉴别良恶性病灶的敏感度和特异度随 b 值升高逐渐升高,在 1200 s/mm2 时达峰值。

临床应用与阅片要点

  • 作者结论认为 b 值 600–1200 s/mm2 时图像质量较高(1000–1200 s/mm2 最佳),b=1200 s/mm2 时信号强度比诊断效能最高。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

RATIONALE AND OBJECTIVES: Breast 3.0 T magnetic resonance diffusion-weighted imaging (MR-DWI) of benign and malignant lesions were obtained to measure and calculate the signal-to-noise ratio (SNR), signal intensity ratio (SIR), and contrast-to-noise ratio (CNR) of lesions at different b-values. The variation patterns of SNR and SIR were analyzed with different b-values and the images of DWI were compared at four different b-values with higher image quality. The effect of SIR on the differential diagnostic efficiency of benign and malignant lesions was compared using receiver operating characteristic curves to provide a reference for selecting the optimal b-value.

MATERIALS AND METHODS: A total of 96 qualified patients with 112 lesions and 14 patients with their contralateral 14 normal breasts were included in this study. The single-shot echo planar imaging sequence was used to perform the DWI and a total of 13 b-values were used: 0, 50, 100, 200, 400, 600, 800, 1000, 1200, 1500, 1800, 2000, and 2500 s/mm2. On DWI, the suitable regions of interest were selected. The SNRs of normal breasts (SNRnormal), SNRlesions, SIR, and CNR of benign and malignant lesions were measured on DWI with different b-values and calculated. The variation patterns of SNR, SIR, and CNR values on DWI for normal breasts, benign lesions, and malignant lesions with different b-values were analyzed by using Pearson correlation analysis. The SNR and SIR of benign and malignant lesions with the same b-values were compared using t-tests. The diagnostic efficiencies of SIR with different b-values for benign and malignant lesions were evaluated using receiver operating characteristic curves.

RESULTS: Breast DWI had higher CNR for b-values ranging from 600 to 1200 s/mm2. It had the best CNR at b = 1000 s/mm2 for the benign lesions and at b = 1200 s/mm2 for the malignant lesions. The signal intensity and SNR values of normal breasts decreased with increasing b-values, with a negative correlation (r = -0.945, P < 0.01). The mean SNR values of benign and malignant lesions were negatively correlated (r = -0.982 and -0.947, respectively, and P < 0.01), gradually decreasing with increasing b-values. The mean SIR value of benign lesions gradually decreased with increasing b-values, a negative correlation (r = -0.991, P < 0.01). The mean SIR values of malignant lesions gradually increased with increasing b-values between 0 and 1200 s/mm2, and gradually decreased with increasing b-values ≥ 1500 s/mm2. For b-values of 600, 800, 1000, and 1200 s/mm2, the sensitivity and specificity of SIR in identifying benign and malignant lesions gradually increased with increasing b-values, peaking at 1200 s/mm2.

CONCLUSIONS: Breast DWI had higher image quality for b-values ranging from 600 to 1200 s/mm2, and was best for b-values ranging from 1000 to 1200 s/mm2. The SIR had the highest diagnostic efficiency in differentiating benign and malignant lesions for a b-value of 1200 s/mm2.

来源:Acad Radiol,查看原文。版权归原作者所有。

引用

  1. Han X, Li J, Wang X, et al. Comparison and Optimization of 3.0 T Breast Images Quality of Diffusion-Weighted Imaging with Multiple B-Values. Acad Radiol. 2016. PMID: 27955879. doi:10.1016/j.acra.2016.11.006

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。