体外血液灌注牛肝模型中三维容积彩色多普勒超声血管指数对射频消融区体积的预测

The prediction of radiofrequency ablation zone volume using vascular indices of 3-dimensional volumetric colour Doppler ultrasound in an in vitro blood-perfused bovine liver model.
PubMed 文献 · Br J Radiol · 2016-12-07

在体外氧合血液灌注系统的13个牛肝37个区域获取三维彩色多普勒容积数据并计算血管指数,随后在相同位置行37次射频消融;FI与大体标本消融区体积显著相关(p=0.0047),但仅解释少量方差(Rβ2=0.21),VI和VFI与消融区体积无显著关联(p>0.05),三者均不能用于临床预测消融区体积。

要点

  • FI与消融区体积显著相关(p=0.0047),但仅解释21%的方差。
  • VI和VFI与消融区体积无显著关联(p>0.05)。
  • VI、FI、VFI均不能用于预测肝脏射频消融区体积。

核心结论

  • 尽管FI与消融区体积相关,但不足以解释其变异性,限制了其临床适用性。
  • VI、FI和VFI在预测肝脏射频消融区体积方面均无临床实用价值。

研究设计与样本

  • 体外研究:13个牛肝连接体外氧合血液灌注系统,从37个区域获取三维彩色多普勒容积数据。
  • 使用三维容积分析软件获得VI、FI和VFI,并在相同位置行37次射频消融。
  • 采用以肝脏为随机效应的一般线性混合模型和后退逐步回归分析血管指数与大体标本消融区体积的关系。

诊断标准与关键数值

  • FI与大体标本消融区体积显著相关(p=0.0047),但解释方差少(Rβ2=0.21)。
  • FI每增加1,消融区体积减少0.23 cm³(95%CI:-0.38,-0.08)。
  • VI和VFI与消融区体积均无显著关联(p>0.05)。

临床应用与阅片要点

  • 原文结论认为VI、FI和VFI不能用于预测肝脏射频消融区体积,并建议研究其他多普勒血管指数以供临床使用。

局限性

  • 原文结论指出FI虽与消融区体积相关,但不能充分解释其变异性,限制其临床适用性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: To determine the most reliable predictor of radiofrequency (RF) ablation zone volume among three-dimensional (3D) volumetric colour Doppler vascular indices in an in vitro blood-perfused bovine liver model.

METHODS: 3D colour Doppler volume data of the local hepatic parenchyma were acquired from 37 areas of 13 bovine livers connected to an in vitro oxygenated blood perfusion system. Doppler vascular indices of vascularization index (VI), flow index (FI) and vascularization flow index (VFI) were obtained from the volume data using 3D volume analysis software. 37 RF ablations were performed at the same locations where the ultrasound data were obtained from. The relationship of these vascular indices and the ablation zone volumes measured from gross specimens were analyzed using a general linear mixed model fit with random effect for liver and backward stepwise regression analysis.

RESULTS: FI was significantly associated with ablation zone volumes measured on gross specimens (p = 0.0047), but explained little of the variance (Rβ2 = 0.21). Ablation zone volume decreased by 0.23 cm3 (95% confidence interval: -0.38, -0.08) for every 1 increase in FI. Neither VI nor VFI was significantly associated with ablation zone volumes (p > 0.05).

CONCLUSION: Although FI was associated with ablation zone volumes, it could not sufficiently explain their variability, limiting its clinical applicability. VI, FI and VFI are not clinically useful in the prediction of RF ablation zone volume in the liver. Advances in knowledge: Despite a significant association of FI with ablation zone volumes, VI, FI and VFI cannot be used for their prediction. Different Doppler vascular indices need to be investigated for clinical use.

来源:Br J Radiol,查看原文。版权归原作者所有。

引用

  1. Chang S, Lanctot AC, McCarter MD, et al. The prediction of radiofrequency ablation zone volume using vascular indices of 3-dimensional volumetric colour Doppler ultrasound in an in vitro blood-perfused bovine liver model. Br J Radiol. 2016. PMID: 27925468. doi:10.1259/bjr.20160661

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。