基于原始数据的金属伪影减少算法可改善接受小金属植入物治疗的肝细胞癌患者腹部CT的图像质量
研究纳入58例接受肝细胞癌治疗后带有小金属植入物(3-15 mm)的患者,行腹部CT并用滤波反投影联合或不联合SEMAR算法重建,比较肝脏平均CT值、伪影指数及主观图像质量。结果显示SEMAR组平均CT值和伪影指数显著低于无SEMAR组(P<0.0001),肝脏、胰腺图像质量及血管显示显著改善(P<0.0001),观察者间一致性为0.62≤κ≤0.83。结论认为SEMAR可通过减少金属伪影改善此类患者腹部CT的图像质量。
要点
- SEMAR算法可显著减少小金属植入物(3-15 mm)在腹部CT中产生的金属伪影。
- SEMAR可改善动态增强CT中肝脏的图像质量。
- SEMAR还可提高肝血管解剖结构显示的可信度。
核心结论
- SEMAR可通过减少金属伪影改善带有小金属植入物患者的腹部CT图像质量。
研究设计与样本
- 研究纳入58例接受肝细胞癌治疗后带有小金属植入物(大小3-15 mm)的患者行腹部CT,数据分别用滤波反投影联合与不联合SEMAR算法在轴位和冠状位重建,计算肝脏内平均CT值(HU和SD)及伪影指数,并由两名读者独立以5分视觉评分评估肝脏、胰腺图像质量及血管显示。
关键影像表现/征象
- SEMAR组肝脏平均CT值和伪影指数显著低于无SEMAR组(P<0.0001)。
- SEMAR组肝脏和胰腺图像质量及血管显示视觉评分显著改善(P<0.0001)。
- 两名读者间一致性为0.62≤κ≤0.83,达高度或几乎完美一致。
诊断标准与关键数值
- 以肝脏内平均CT值(HU和SD)及伪影指数(AI)评估金属伪影减少程度。
- 以5分视觉评分评估肝脏和胰腺图像质量及血管显示。
- SEMAR组平均CT值和伪影指数显著低于无SEMAR组(P<0.0001)。
临床应用与阅片要点
- 对带有小金属植入物(3-15 mm)的肝细胞癌治疗后患者行腹部CT时,可考虑使用SEMAR算法减少金属伪影。
- SEMAR可改善动态增强CT中肝脏图像质量及肝血管解剖结构显示的可信度。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: To determine the value of a raw data-based metal artifact reduction (SEMAR) algorithm for image quality improvement in abdominal CT for patients with small metal implants.
METHODS: Fifty-eight patients with small metal implants (3-15 mm in size) who underwent treatment for hepatocellular carcinoma were imaged with CT. CT data were reconstructed by filtered back projection with and without SEMAR algorithm in axial and coronal planes. To evaluate metal artefact reduction, mean CT number (HU and SD) and artefact index (AI) values within the liver were calculated. Two readers independently evaluated image quality of the liver and pancreas and visualization of vasculature using a 5-point visual score. HU and AI values and image quality on images with and without SEMAR were compared using the paired Student's t-test and Wilcoxon signed rank test. Interobserver agreement was evaluated using linear-weighted κ test.
RESULTS: Mean HU and AI on images with SEMAR was significantly lower than those without SEMAR (P < 0.0001). Liver and pancreas image qualities and visualizations of vasculature were significantly improved on CT with SEMAR (P < 0.0001) with substantial or almost perfect agreement (0.62 ≤ κ ≤ 0.83).
CONCLUSIONS: SEMAR can improve image quality in abdominal CT in patients with small metal implants by reducing metallic artefacts.
KEY POINTS: • SEMAR algorithm significantly reduces metallic artefacts from small implants in abdominal CT. • SEMAR can improve image quality of the liver in dynamic CECT. • Confidence visualization of hepatic vascular anatomies can also be improved by SEMAR.