文献速递腹部MR

自由呼吸动态肝脏检查:采用中等欠采样径向三维T1加权梯度回波序列用于屏气能力受限患者

Free-breathing dynamic liver examination using a radial 3D T1-weighted gradient echo sequence with moderate undersampling for patients with limited breath-holding capacity.
PubMed 文献 · Eur J Radiol · 2016-11-03

该研究纳入23例因严重呼吸伪影无法屏气的肿瘤随访患者,比较自由呼吸中等欠采样径向VIBE(us-radial-VIBE)与标准屏气三维梯度回波VIBE的动态肝脏MRI。us-radial-VIBE的SNR显著更低、图像噪声更高,但信号强度无差异,动脉期对比增强比更高;主观评分标准屏气序列略高,但在屏气序列也有严重呼吸伪影的亚组中us-radial-VIBE评分显著更高。作者认为us-radial-VIBE是屏气不能患者有前景的替代方案。

要点

  • us-radial-VIBE的SNR显著低于标准屏气3D GRE VIBE(p<0.0001),图像噪声显著更高(p<0.0001)。
  • us-radial-VIBE动脉期对比增强比显著高于标准屏气序列(p=0.028),时间分辨率准确。
  • us-radial-VIBE几乎均有条纹伪影,但不影响诊断有效性;标准屏气序列呼吸伪影更严重。

核心结论

  • 自由呼吸动态肝脏成像使用us-radial-VIBE可提供准确的时间分辨率、低运动伪影敏感性和良好图像质量,是屏气不能患者有前景的替代方案。

研究设计与样本

  • 23例连续患者在肿瘤随访中接受自由呼吸us-radial-VIBE动态肝脏MR检查,均因计划或平扫序列严重呼吸伪影而符合自由呼吸方案;将us-radial-VIBE与患者上次分期肝脏MRI的标准屏气3D GRE VIBE比较,客观评估信号强度、图像噪声、SNR和对比增强比,并由四名读者独立、回顾性、盲法对代表性图像质量参数及典型伪影评分。

关键影像表现/征象

  • us-radial-VIBE几乎全部采集均可见条纹伪影,但不影响诊断有效性。
  • 标准屏气3D GRE VIBE中呼吸伪影总体更严重。

诊断标准与关键数值

  • us-radial-VIBE的SNR显著低于标准屏气3D GRE VIBE(p<0.0001),图像噪声显著更高(p<0.0001)。
  • 信号强度在us-radial-VIBE与标准屏气3D GRE VIBE间无差异(p=0.62)。
  • 动脉期对比增强比us-radial-VIBE显著高于标准屏气3D GRE VIBE(p=0.028)。
  • 主观图像质量参数标准屏气3D GRE VIBE总体略高评分;在参考屏气3D GRE VIBE也有严重呼吸伪影的较小亚组中,us-radial-VIBE图像质量评分显著更高。

临床应用与阅片要点

  • 对屏气能力受限、计划或平扫序列已见严重呼吸伪影的患者,可考虑自由呼吸us-radial-VIBE动态肝脏检查。
  • 阅片时注意us-radial-VIBE几乎均存在条纹伪影,但原文指出其不影响诊断有效性;标准屏气序列呼吸伪影更严重。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To compare free-breathing radial VIBE with moderate undersampling (us-radial-VIBE) with a standard breathhold T1-weighted volumetric interpolated sequence (3D GRE VIBE) in patients unable to suspend respiration during dynamic liver examination.

MATERIAL AND METHODS: 23 consecutive patients underwent dynamic liver MR examination using the free-breathing us-radial-VIBE sequence as part of their oncologic follow-up. All patients were eligible for the free-breathing protocol due to severe respiratory artifacts at the planning or precontrast sequences. The us-radial-VIBE acquisitions were compared to the patientś last staging liver MRI including a standard breathhold 3D GRE VIBE. For an objective image evaluation, signal intensity (SI), image noise (IN), signal-to-noise ratio (SNR) and contrast-enhancement ratio (CER) were compared. Representative image quality parameters, including typical artifacts were independently, retrospectively and blindly scored by four readers.

RESULTS: Us-radial-VIBE had significant lower SNR (p<0.0001) and higher IN (p<0.0001), whereas SI did not differ (p=0.62). Temporal resolution assessed with CER in the arterial phase showed higher values for us-radial-VIBE (p=0.028). Subjective image quality parameters received generally slightly higher scores for 3D GRE VIBE. In a smaller subgroup comprising patients with severe respiratory artifacts also at reference breathhold 3D GRE VIBE examination, us-radial-VIBE showed significantly higher image quality scores. Furthermore, there were generally more severe respiratory artifacts in 3D GRE VIBE, whereas streaking was characteristic in almost all us-radial-VIBE acquisitions but did not affect diagnostic validity.

CONCLUSION: Free-breathing dynamic liver imaging using us-radial-VIBE delivers accurate temporal resolution, low motion artifact susceptibility and good image quality and represents a promising alternative in patients unable to suspend respiration.

来源:Eur J Radiol,查看原文。版权归原作者所有。