混合迭代重建算法降低剂量CT检测富血供肝脏病变的诊断效能:体模研究
研究使用30个含或不含模拟富血供病变的肝脏体模,在320层CT上以对照剂量(40 mAs)和降低剂量(30、20 mAs)扫描,对照剂量图像用滤波反投影(FBP)重建,降低剂量图像用FBP和混合迭代重建(IR)算法重建。结果显示降低剂量混合IR图像客观噪声更低、CNR更高(P < .05),但JAFROC1品质因子低于对照剂量FBP图像,仅20 mAs图像与对照剂量FBP图像差异对两位阅片者均有统计学意义(P < .01)。结论认为大幅降低辐射剂量会损害富血供肝脏病变的可检测性,尽管混合IR算法可保持客观图像噪声和CNR。
要点
- 降低剂量(30、20 mAs)混合IR图像客观噪声显著低于对照剂量(40 mAs)FBP图像(P < .05)。
- 降低剂量混合IR图像CNR显著高于对照剂量FBP图像(P < .05)。
- 20 mAs混合IR图像JAFROC1品质因子显著低于对照剂量FBP图像(两位阅片者P < .01)。
核心结论
- 大幅降低辐射剂量会损害富血供肝脏病变的可检测性,尽管混合IR算法可保持客观图像噪声和CNR。
研究设计与样本
- 体模研究,30个含或不含模拟富血供病变的肝脏体模,320层CT扫描,对照剂量40 mAs与降低剂量30、20 mAs,对照剂量用FBP重建,降低剂量用FBP和混合IR重建,2位盲法阅片者独立判读,行JAFROC分析。
诊断标准与关键数值
- 降低剂量(30和20 mAs)混合IR图像客观噪声显著低于对照剂量(40 mAs)FBP图像(P < .05)。
- 降低剂量(30和20 mAs)混合IR图像CNR显著高于对照剂量(40 mAs)FBP图像(P < .05)。
- 20 mAs混合IR图像JAFROC1品质因子显著低于对照剂量FBP图像(两位阅片者P < .01)。
- 30 mAs混合IR图像JAFROC1品质因子低于对照剂量FBP图像,但差异对两位阅片者均未达统计学显著。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: To investigate the diagnostic performance of reduced-dose CT with a hybrid iterative reconstruction (IR) algorithm for the detection of hypervascular liver lesions.
METHODS: Thirty liver phantoms with or without simulated hypervascular lesions were scanned with a 320-slice CT scanner with control-dose (40 mAs) and reduced-dose (30 and 20 mAs) settings. Control-dose images were reconstructed with filtered back projection (FBP), and reduced-dose images were reconstructed with FBP and a hybrid IR algorithm. Objective image noise and the lesion to liver contrast-to-noise ratio (CNR) were evaluated quantitatively. Images were interpreted independently by 2 blinded radiologists, and jackknife alternative free-response receiver-operating characteristic (JAFROC) analysis was performed.
RESULTS: Hybrid IR images with reduced-dose settings (both 30 and 20 mAs) yielded significantly lower objective image noise and higher CNR than control-dose FBP images (P < .05). However, hybrid IR images with reduced-dose settings had lower JAFROC1 figure of merit than control-dose FBP images, although only the difference between 20 mAs images and control-dose FBP images was significant for both readers (P < .01).
CONCLUSIONS: An aggressive reduction of the radiation dose would impair the detectability of hypervascular liver lesions, although objective image noise and CNR would be preserved by a hybrid IR algorithm.
KEY POINTS: • A half-dose scan with a hybrid iterative reconstruction preserves objective image quality. • A hybrid iterative reconstruction algorithm does not improve diagnostic performance. • An aggressive dose reduction would impair the detectability of low-contrast lesions.
引用
- Nakamoto A, Tanaka Y, Juri H, et al. Diagnostic performance of reduced-dose CT with a hybrid iterative reconstruction algorithm for the detection of hypervascular liver lesions: a phantom study. Eur Radiol. 2016. PMID: 27957640. doi:10.1007/s00330-016-4687-6