文献速递腹部技术与安全CT

混合迭代重建对平扫肝脏CT的影响

Impact of hybrid iterative reconstruction on unenhanced liver CT.
PubMed 文献 · Br J Radiol · 2016-12-20

研究纳入30例肝细胞癌(HCC)患者行平扫CT,比较滤波反投影(FBP)与混合迭代重建(HIR)弱(Level 1)、中(Level 4)、强(Level 7)三档的图像。定量显示强档HIR病灶衰减高于FBP,三档HIR肝实质衰减均低于FBP,各档HIR噪声均小于FBP;定性显示病灶显示清晰度与定性诊断在FBP与各档HIR间无显著差异。

要点

  • 强档HIR(Level 7)病灶衰减高于FBP(p=0.0005),但病灶显示清晰度与定性诊断无显著差异。
  • 三档HIR肝实质衰减均低于FBP(p≤0.0002),阅片时需注意肝实质CT值变化。
  • 各档HIR图像噪声均显著小于FBP(p<0.0001),但噪声降低未转化为病灶显示改善。

核心结论

  • HIR与FBP虽在病灶及肝实质衰减上存在显著差异,但HIR对病灶显示清晰度和定性诊断无显著影响。

研究设计与样本

  • 30例肝细胞癌(HCC)患者行平扫CT,图像分别以FBP及HIR弱(Level 1)、中(Level 4)、强(Level 7)三档重建(iDose4;Philips Medical Systems,Cleveland,OH),定量比较采用双侧配对t检验,定性比较采用Wilcoxon符号秩检验。

诊断标准与关键数值

  • 强档HIR病灶衰减显著高于FBP(p=0.0005);三档HIR肝实质衰减均显著低于FBP(所有比较p≤0.0002);各档HIR图像噪声均显著小于FBP(任意比较p<0.0001);FBP与各档HIR间病灶显示清晰度及定性诊断无显著差异(所有比较p≥0.0819)。

临床应用与阅片要点

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: To clarify the impact of hybrid iterative reconstruction (HIR) and filtered back projection (FBP) on unenhanced liver CT.

METHODS: 30 patients with hepatocellular carcinoma (HCC) underwent unenhanced CT. The images were reconstructed with FBP and weak (Level 1), mild (Level 4) and strong (Level 7) levels of HIR (iDose4; Philips Medical Systems, Cleveland, OH). Quantitatively, attenuations of the HCC (with the largest lesion in each case), hepatic parenchyma (the average of four segments) and image noise (standard deviation of the attenuations in hepatic parenchyma) were compared between the four kinds of reconstruction using the two-tailed paired t-test. Qualitatively, liver lesion conspicuity and characterization were also compared using the Wilcoxon signed-rank test.

RESULTS: Attenuation of the liver lesion with the strong level of HIR was significantly higher than that with FBP (p = 0.0005). Attenuations of hepatic parenchyma with all three HIR levels were significantly lower than that with FBP (p ≤ 0.0002 in all comparisons). Image noise with each of the three HIR levels was significantly smaller than that with FBP (p < 0.0001 in any comparison). There was no significant difference in lesion conspicuity and characterization between FBP and each HIR level (p ≥ 0.0819 in all comparisons).

CONCLUSION: Although attenuations of the liver lesion and hepatic parenchyma were significantly different between HIR and FBP, HIR had no significant effect on lesion conspicuity and characterization. Advances in knowledge: Attenuations of liver lesions and hepatic parenchyma differ significantly between HIR and FBP images.

来源:Br J Radiol,查看原文。版权归原作者所有。

引用

  1. Kondo M, Nishie A, Fujita N, et al. Impact of hybrid iterative reconstruction on unenhanced liver CT. Br J Radiol. 2016. PMID: 27993096. doi:10.1259/bjr.20150670

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。