T1 mapping心血管磁共振成像检测心肌炎:切片方向对诊断性能的影响
本研究纳入48例临床诊断心肌炎患者和13例健康对照,在1.5T行CMR T1 mapping,比较短轴与长轴切片方向对诊断性能的影响。结果显示短轴切片的 native T1 和 ECV-BOPTA 诊断性能优于长轴切片,长轴切片似乎低估心肌炎患者 native T1;疑似心肌炎的 T1 mapping 可仅用单个中间心室短轴切片而不显著损失诊断性能。
要点
- 短轴切片 native T1 和 ECV-BOPTA 诊断心肌炎的 AUC 高于长轴切片。
- 心肌炎患者 3LAX 的 native T1 低于 3SAX,长轴切片似乎低估 native T1。
- 疑似心肌炎可仅用单个中间心室短轴切片行 T1 mapping,诊断性能无显著损失。
核心结论
- 短轴切片的 native T1 和 ECV-BOPTA 对心肌炎的诊断性能优于长轴切片,因长轴切片似乎低估心肌炎患者的 native 心肌 T1;疑似心肌炎的 T1 mapping 可限制于单个中间心室短轴切片而不显著损失诊断性能。
研究设计与样本
- 研究纳入48例临床定义的心肌炎患者和13例健康对照,在1.5T行CMR,采用改良 Look-Locker 反转恢复(MOLLI)序列在注射0.075mmol/kg 钆布醇(Gadolinium-BOPTA)前及后15min行 T1 mapping,分别采集3个短轴和3个长轴切片,使用 OsiriX 专用插件计算 native T1、增强后 T1 和 ECV-BOPTA 图。
诊断标准与关键数值
- native 心肌 T1 鉴别心肌炎与健康对照的 AUC:3SAX 0.85(p<0.0001)、1SAX 0.85(p<0.0001)、3LAX 0.76(p=0.0002)、4CH 0.70(p=0.0075)、3CH 0.72(p=0.0020)、2CH 0.75(p=0.0003)。
- ECV-BOPTA 鉴别心肌炎与健康对照的 AUC:3SAX 0.83(p<0.0001)、1SAX 0.82(p<0.0001)、3LAX 0.77(p=0.0005)、4CH 0.71(p=0.0079)、3CH 0.69(p=0.0371)、2CH 0.75(p=0.0006)。
- 心肌炎患者 native T1:3LAX 1081ms(1037-1131ms)vs 3SAX 1107ms(1069-1143ms),p=0.0022。
- 健康对照 native T1:3LAX 1026ms(1009-1059ms)vs 3SAX 1039ms(1023-1055ms),p=0.2719。
临床应用与阅片要点
- 作者结论认为,短轴切片 native T1 和 ECV-BOPTA 对心肌炎的诊断性能优于长轴切片,因长轴切片似乎低估心肌炎患者 native 心肌 T1。
- 作者结论认为,疑似心肌炎的 T1 mapping 可限制于单个中间心室短轴切片而不显著损失诊断性能。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND: T1 mapping is a promising diagnostic tool to improve the diagnostic accuracy of cardiovascular magnetic resonance (CMR) in patients with suspected myocarditis. However, there are currently no data on the potential influence of slice orientation on the diagnostic performance of CMR. Thus, we compared the diagnostic performance of global myocardial T1 and extracellular volume (ECV) values to differentiate patients with myocarditis from healthy individuals between different slice orientations.
METHODS: This study included 48 patients with clinically defined myocarditis and 13 healthy controls who underwent CMR at 1.5T. A modified Look-Locker inversion-recovery (MOLLI) sequence was used for T1 mapping before and 15min after administration of 0.075mmol/kg Gadolinium-BOPTA. T1 mapping was performed on three short and on three long axes slices, respectively. Native T1, post-contrast T1 and extracellular volume (ECV) -BOPTA maps were calculated using a dedicated plug-in written for the OsiriX software and compared between the mean value of three short-axes slices (3SAX), the central short-axis (1SAX), the mean value of three long-axes slices (3LAX), the four-chamber view (4CH), the three-chamber view (3CH) and the two-chamber view (2CH).
RESULTS: There were significantly lower native T1 values on 3LAX (1081ms (1037-1131ms)) compared to 3SAX (1107ms (1069-1143ms), p=0.0022) in patients with myocarditis, but not in controls (1026ms (1009-1059ms) vs. 1039ms (1023-1055ms), p=0.2719). The areas under the curve (AUC) to discriminate between myocarditis and healthy controls by native myocardial T1 were 0.85 (p<0.0001) on 3SAX, 0.85 (p<0.0001) on 1SAX, 0.76 (p=0.0002) on 3LAX, 0.70 (p=0.0075) on 4CH, 0.72 (p=0.0020) on 3CH and 0.75 (p=0.0003) on 2CH. The AUCs for ECV-BOPTA were 0.83 (p<0.0001) on 3 SAX, 0.82 (p<0.0001) on 1SAX, 0.77 (p=0.0005) on 3LAX, 0.71 (p=0.0079) on 4CH, 0.69 (p=0.0371) on 3CH and 0.75 (p=0.0006) on 2CH.
CONCLUSION: Native T1 and ECV-BOPTA on short axes slices provide a better diagnostic performance in myocarditis than long axes slices since long axes slices seem to underestimate native myocardial T1 in myocarditis. T1 mapping in suspected myocarditis can be restricted to a single mid-ventricular short-axis slice without a significant loss in diagnostic performance.