先天性心脏病婴儿模型迭代重建CT血管成像的图像质量:与滤波反投影及混合迭代重建的比较

Image quality of ct angiography using model-based iterative reconstruction in infants with congenital heart disease: Comparison with filtered back projection and hybrid iterative reconstruction.
PubMed 文献 · Eur J Radiol · 2016-10-17

纳入51例先天性心脏病婴儿,行256层多排CT血管成像(80kVp、前瞻性心电触发),比较滤波反投影、混合迭代重建(iDose4)与迭代模型重建(IMR)的图像质量、冠状动脉显示率及诊断准确率。IMR的噪声显著低于FBP和iDose4(P<0.01),SNR、CNR及主观图像质量评分显著更高(P<0.01)。iDose4和IMR显示的冠状动脉节段总数均显著多于FBP(P=0.002和P=0.025),但iDose4与IMR间无显著差异(P=0.397)。三种算法的诊断准确率无显著差异(χ2=0.343,P=0.842)。

要点

  • IMR较FBP和iDose4显著降低图像噪声并提高SNR、CNR及主观评分(均P<0.01)。
  • iDose4和IMR显示的冠状动脉节段总数均显著多于FBP(P=0.002和P=0.025)。
  • 三种重建算法的诊断准确率无显著差异(χ2=0.343,P=0.842)。

核心结论

  • 在先天性心脏病婴儿心脏CTA中,IMR较FBP和iDose4显著提高客观与主观图像质量,但较iDose4未提高诊断准确率或冠状动脉显示。

研究设计与样本

  • 回顾性纳入2014年12月至2015年3月在本机构行心脏CTA的51例先天性心脏病婴儿,采用256层多排CT血管成像、前瞻性心电触发、80kVp管电压,比较FBP、iDose4和IMR三种重建算法。

关键影像表现/征象

  • IMR较FBP和iDose4噪声显著降低(P<0.01),SNR和CNR显著升高(P<0.01)。
  • IMR主观图像质量评分显著高于FBP和iDose4(P<0.01)。
  • iDose4和IMR显示的冠状动脉节段总数均显著多于FBP(P=0.002和P=0.025),iDose4与IMR间无显著差异(P=0.397)。

诊断标准与关键数值

  • 平均有效辐射剂量为0.61±0.32 mSv。
  • 三种算法诊断准确率无显著差异(χ2=0.343,P=0.842)。

临床应用与阅片要点

  • 在婴儿先天性心脏病心脏CTA中,IMR可显著改善图像噪声、SNR、CNR及主观图像质量。
  • IMR较iDose4未提高诊断准确率或冠状动脉显示,iDose4与IMR显示的冠状动脉节段总数无显著差异(P=0.397)。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To compare the image quality, rate of coronary artery visualization and diagnostic accuracy of 256-slice multi-detector computed tomography angiography (CTA) with prospective electrocardiographic (ECG) triggering at a tube voltage of 80kVp between 3 reconstruction algorithms (filtered back projection (FBP), hybrid iterative reconstruction (iDose4) and iterative model reconstruction (IMR)) in infants with congenital heart disease (CHD).

METHODS: Fifty-one infants with CHD who underwent cardiac CTA in our institution between December 2014 and March 2015 were included. The effective radiation doses were calculated. Imaging data were reconstructed using the FBP, iDose4 and IMR algorithms. Parameters of objective image quality (noise, signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR)); subjective image quality (overall image quality, image noise and margin sharpness); coronary artery visibility; and diagnostic accuracy for the three algorithms were measured and compared.

RESULTS: The mean effective radiation dose was 0.61±0.32 mSv. Compared to FBP and iDose4, IMR yielded significantly lower noise (P<0.01), higher SNR and CNR values (P<0.01), and a greater subjective image quality score (P<0.01). The total number of coronary segments visualized was significantly higher for both iDose4 and IMR than for FBP (P=0.002 and P=0.025, respectively), but there was no significant difference in this parameter between iDose4 and IMR (P=0.397). There was no significant difference in the diagnostic accuracy between the FBP, iDose4 and IMR algorithms (χ2=0.343, P=0.842).

CONCLUSIONS: For infants with CHD undergoing cardiac CTA, the IMR reconstruction algorithm provided significantly increased objective and subjective image quality compared with the FBP and iDose4 algorithms. However, IMR did not improve the diagnostic accuracy or coronary artery visualization compared with iDose4.

来源:Eur J Radiol,查看原文。版权归原作者所有。