文献速递心血管CT

利用体重和身高依赖的扫描触发阈值改善冠状动脉CT血管造影的图像质量

Improving Image Quality of Coronary Computed Tomography Angiography Using Patient Weight and Height-Dependent Scan Trigger Threshold.
PubMed 文献 · Acad Radiol · 2016-12-09

本研究基于100例按体重指数设定触发阈值的CCTA数据建立触发阈值与身高、体重的回归方程并生成查询表,随后将100例连续患者分为个体化阈值组(n=50)与常规150 HU对照组(n=50)。个体化组左主干、右冠近段及主动脉强化分别平均提高12.30%、12.94%、10.6%,相应对比噪声比提高26.03%、27.08%、23.17%,主动脉与左心室对比提高633.1%,右心室与下腔静脉强化平均降低22.7%与32.7%,两组图像噪声无显著差异(P > .05),个体化组主观图像质量评分更高。

要点

  • 回归方程:触发阈值K(HU)=0.811×身高H(cm)+1.917×体重BW(kg)-99.341。
  • 个体化组左主干、右冠近段及主动脉强化较对照组分别平均提高12.30%、12.94%、10.6%。
  • 个体化组右心室与下腔静脉强化较对照组分别平均降低22.7%与32.7%,两组图像噪声无显著差异(P > .05)。

核心结论

  • 使用个体化体重和身高依赖的Smart Prep阈值调整扫描触发时间可显著改善CCTA图像质量。

研究设计与样本

  • 研究分两部分:第一部分分析100例按体重指数依赖Smart Prep序列(触发阈值80 HU至250 HU)采集的择期CCTA数据并建立回归方程与查询表;第二部分将100例连续患者分为个体化组(n=50,体重和身高依赖阈值)与对照组(n=50,常规恒定阈值150 HU),比较冠状动脉、主动脉、左右心室及下腔静脉强化和主观图像质量评分。

关键影像表现/征象

  • 个体化组左主干冠状动脉强化较对照组平均提高12.30%,近段右冠状动脉平均提高12.94%,主动脉平均提高10.6%。
  • 个体化组右心室强化较对照组平均降低22.7%,下腔静脉平均降低32.7%。
  • 两组图像噪声无显著差异(P > .05),个体化组主观图像质量评分高于对照组。

诊断标准与关键数值

  • Smart Prep触发阈值(K,HU)、身高(H,cm)与体重(BW,kg)的回归方程为K=0.811×H+1.917×BW-99.341。
  • 第一部分按体重指数设定触发阈值范围80 HU至250 HU,第二部分对照组采用常规恒定阈值150 HU。
  • 个体化组左主干、近段右冠状动脉及主动脉的对比噪声比分别较对照组提高26.03%、27.08%、23.17%,主动脉与左心室对比提高633.1%。

临床应用与阅片要点

  • 可根据身高和体重查询表设定个体化Smart Prep触发阈值,以替代常规恒定150 HU阈值。
  • 个体化阈值在提高冠状动脉与主动脉强化的同时降低右心室和下腔静脉强化,有利于减少右心系统高浓度对比剂干扰。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

RATIONALE AND OBJECTIVES: We aim to improve the image quality of coronary computed tomography angiography (CCTA) by using personalized weight and height-dependent scan trigger threshold.

MATERIALS AND METHODS: This study was divided into two parts. First, we performed and analyzed the 100 scheduled CCTA data, which were acquired by using body mass index-dependent Smart Prep sequence (trigger threshold ranged from 80 Hu to 250 Hu based on body mass index). By identifying the cases of high quality image, a linear regression equation was established to determine the correlation among the Smart Prep threshold, height, and body weight. Furthermore, a quick search table was generated for weight and height-dependent Smart Prep threshold in CCTA scan. Second, to evaluate the effectiveness of the new individual threshold method, an additional 100 consecutive patients were divided into two groups: individualized group (n = 50) with weight and height-dependent threshold and control group (n = 50) with the conventional constant threshold of 150 HU. Image quality was compared between the two groups by measuring the enhancement in coronary artery, aorta, left and right ventricle, and inferior vena cava. By visual inspection, image quality scores were performed to compare between the two groups.

RESULTS: Regression equation between Smart Prep threshold (K, Hu), height (H, cm), and body weight (BW, kg) was K = 0.811 × H + 1.917 × BW - 99.341. When compared to the control group, the individualized group presented an average overall increase of 12.30% in enhancement in left main coronary artery, 12.94% in proximal right coronary artery, and 10.6% in aorta. Correspondingly, the contrast-to-noise ratios increased by 26.03%, 27.08%, and 23.17%, respectively, and by 633.1% in contrast between aorta and left ventricle. Meanwhile, the individualized group showed an average overall decrease of 22.7% in enhancement of right ventricle and 32.7% in inferior vena cava. There was no significant difference of the image noise between the two groups (P > .05). By visual inspection, the image quality score of the individualized group was higher than that of the control group.

CONCLUSION: Using personalized weight and height-dependent Smart Prep threshold to adjust scan trigger time can significantly improve the image quality of CCTA.

来源:Acad Radiol,查看原文。版权归原作者所有。

引用

  1. Kang D, Hua H, Peng N, et al. Improving Image Quality of Coronary Computed Tomography Angiography Using Patient Weight and Height-Dependent Scan Trigger Threshold. Acad Radiol. 2016. PMID: 27940232. doi:10.1016/j.acra.2016.11.013

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。