文献速递心血管CT

高心率或不规则心率患者采用70 kV管电压行回顾性心电门控冠状动脉CT血管成像的辐射剂量水平

Radiation Dose Levels of Retrospectively ECG-Gated Coronary CT Angiography Using 70-kVp Tube Voltage in Patients with High or Irregular Heart Rates.
PubMed 文献 · Acad Radiol · 2016-10-28

研究纳入154例高心率或不规则心率患者,前瞻性分别采用第三代双源CT 70 kV(103例)或第二代双源CT 100 kV(51例)行回顾性心电门控冠状动脉CT血管成像。70 kV组平均有效辐射剂量4.5 mSv,低于100 kV组的8.4 mSv(P < 0.05),降低52%。采用全期相重建算法时所有冠状动脉节段均可诊断评估,而最佳舒张期重建有71%患者存在运动伪影,最佳收缩期重建降至37%。

要点

  • 70 kV回顾性心电门控冠状动脉CT血管成像平均有效剂量4.5 mSv,较100 kV的8.4 mSv降低52%。
  • 最佳舒张期重建时71%患者存在运动伪影,最佳收缩期重建降至37%。
  • 采用全期相重建算法时两种方案所有冠状动脉节段均可诊断评估。

核心结论

  • 70 kV回顾性心电门控冠状动脉CT血管成像使辐射剂量降低52%,采用全期相重建算法可对所有冠状动脉节段进行狭窄的诊断评估,而单用最佳舒张期或最佳收缩期重建则不行。

研究设计与样本

  • 经机构审查委员会批准的前瞻性研究,纳入154例高心率或不规则心率患者(中位年龄54岁,98例男性),70 kV组103例采用第三代双源CT,100 kV组51例采用第二代双源CT,均行无管电流调制的回顾性心电门控冠状动脉CT血管成像。图像按R-R间期10%间隔重建为最佳舒张期、最佳收缩期和全期相三种方式,评估客观与主观图像质量及运动伪影。

关键影像表现/征象

  • 最佳舒张期重建时71%(110例)患者在一个或多个冠状动脉节段出现运动伪影。
  • 最佳收缩期重建时出现运动伪影的患者数降至37%(57例)。
  • 采用全期相重建时,两种冠状动脉CT血管成像方案的所有血管和冠状动脉节段均可评估。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

RATIONALE AND OBJECTIVES: Despite ongoing technical refinements, coronary computed tomography angiography (cCTA) remains challenging in its diagnostic value by electrocardiographic (ECG) misregistration and motion artifacts, which commonly occur in patients with atrial fibrillation and high or irregular heart rates. The aim of this study was to evaluate the radiation dose and the number of inconclusive coronary segments at cCTA using retrospective ECG gating at 100 and 70 kV.

MATERIALS AND METHODS: With institutional review board approval, 154 patients (median age 54 years, 98 men) with high or irregular heart rate prospectively underwent retrospectively ECG-gated cCTA without tube current modulation on a third-generation dual-source computed tomography (DSCT) system at 70 kV (n = 103) or on a second-generation DSCT system at 100 kV (n = 51). Images were reconstructed in best diastolic phase (BDP), best systolic phase (BSP), and in all phases (APs) at 10% intervals across the R-R cycle. Objective and subjective image qualities were evaluated as well as the presence of motion artifacts with the three different reconstruction approaches.

RESULTS: The mean heart rate was 93 ± 16 bpm. The mean effective radiation dose was 4.5 mSv for 70 kV compared to 8.4 mSv for 100 kV (P < 0.05). At BDP reconstruction, 71% (n = 110) of the patients showed motion artifacts in one or more coronary segments. At BSP reconstruction, the number of patients with motion artifacts decreased to 37% (n = 57). In contrast, if images were reconstructed with the AP approach, all vessels and coronary segments were evaluable with both cCTA protocols.

CONCLUSIONS: Retrospectively ECG-gated cCTA at 70 kV results in 52% decreased radiation dose. Further using the AP algorithm allowed for diagnostic evaluation of all coronary segments for stenosis, in contrast to BDP or BSP phase alone.