文献速递儿科影像心血管CT

低浓度碘对比剂联合低管电压与低管电流在复杂先天性心脏病患儿心脏CT中的对比剂剂量、辐射剂量与图像质量的前瞻性评估

A prospective evaluation of contrast and radiation dose and image quality in cardiac CT in children with complex congenital heart disease using low-concentration iodinated contrast agent and low tube voltage and current.
PubMed 文献 · Br J Radiol · 2016-12-07

前瞻性随机研究纳入110例先天性心脏病患儿,比较低浓度对比剂联合低管电流方案(A组,n=45,270 mgI/ml、120 mA)与常规方案(B组,n=65,370 mgI/ml、160 mA),两组均用80 kVp及70%自适应统计迭代重建。A组碘负荷与有效剂量均显著低于B组,而图像噪声、CT值、CNR、SNR及主观图像质量两组相似,诊断准确率亦相近。

要点

  • A组碘负荷3976±747 mgI、有效剂量0.60±0.08 mSv,均显著低于B组(p<0.001)。
  • 两组图像噪声、CT值、CNR、SNR及主观评分相似(均p>0.05),观察者间一致性好。
  • A组心外与心内缺损诊断准确率分别为96%和92%,B组为95%和93%。

核心结论

  • 与标准剂量方案相比,采用80 kVp管电压、120 mA管电流和270 mgI/ml低浓度碘对比剂可使碘负荷降低30%、辐射剂量降低22%,同时保持相当的图像质量和诊断准确率。

研究设计与样本

  • 前瞻性随机研究,纳入110例先天性心脏病患儿,随机分为A组(n=45,120 mA管电流、270 mgI/ml对比剂Visipaque)和B组(n=65,常规160 mA、370 mgI/ml对比剂Iopamiro);两组均用80 kVp管电压并以70%自适应统计迭代重建算法重建。

关键影像表现/征象

  • 测量主动脉弓处CT值与噪声并计算SNR和CNR,两组间噪声、CT值、CNR、SNR均相似(均p>0.05)。
  • 采用五点量表主观评价图像质量,两组主观图像质量相似(p>0.05),两名观察者间一致性好。

诊断标准与关键数值

  • A组碘负荷3976±747 mgI vs B组5763±1018 mgI,A组降低30%(p<0.001)。
  • A组有效剂量0.60±0.08 mSv vs B组0.77±0.10 mSv,A组降低22%(p<0.001)。
  • 与手术结果对照,A组心外与心内缺损诊断准确率分别为96%和92%,B组分别为95%和93%。

临床应用与阅片要点

  • 作者结论认为,该低管电压、低管电流联合低浓度碘对比剂方案可大幅减少辐射和对比剂对儿童的不良影响,并可有效用于复杂先天性心脏病的检查。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: To the assess image quality, contrast dose and radiation dose in cardiac CT in children with congenital heart disease (CHD) using low-concentration iodinated contrast agent and low tube voltage and current in comparison with standard dose protocol.

METHODS: 110 patients with CHD were randomized to 1 of the 2 scan protocols: Group A (n = 45) with 120 mA tube current and contrast agent of 270 mgI/ml in concentration (Visipaque™; GE Healthcare Ireland, Co., Cork, UK); and Group B (n = 65) with the conventional 160 mA and 370 mgI/ml concentration contrast (Iopamiro®; Shanghai Bracco Sine Pharmaceutical Corp Ltd, Shanghai, China). Both groups used 80 kVp tube voltage and were reconstructed with 70% adaptive statistical iterative reconstruction algorithm. The CT value and noise in aortic arch were measured and the signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were calculated. A five-point scale was used to subjectively evaluate image quality. Contrast and radiation dose were recorded.

RESULTS: There was no difference in age and weight between the two groups (all p > 0.05). The iodine load and radiation dose in Group A were statistically lower (3976 ± 747 mgI vs 5763 ± 1018 mgI in iodine load and 0.60 ± 0.08 mSv vs 0.77 ± 0.10 mSv in effective dose; p < 0.001). However, image noise, CT value, CNR, SNR and subjective image quality for the two groups were similar (all p > 0.05), and with good agreement between the two observers. Comparing the surgery results, the diagnostic accuracy for extracardiac and intracardiac defects for Group A was 96% and 92%, respectively, while the corresponding numbers for Group B were 95% and 93%.

CONCLUSION: Compared with the standard dose protocol, the use of low tube voltage (80 kVp), low tube current (120 mA) and low-concentration iodinated contrast agent (270 mgI/ml) enables a reduction of 30% in iodine load and 22% in radiation dose while maintaining compatible image quality and diagnostic accuracy. Advances in knowledge: The new cardiac CT scanning protocol can largely reduce the adverse effects of radiation and contrast media to children. Meanwhile, it also can be used effectively to examine complex CHD.

来源:Br J Radiol,查看原文。版权归原作者所有。

引用

  1. Hou QR, Gao W, Sun AM, et al. A prospective evaluation of contrast and radiation dose and image quality in cardiac CT in children with complex congenital heart disease using low-concentration iodinated contrast agent and low tube voltage and current. Br J Radiol. 2016. PMID: 27925466. doi:10.1259/bjr.20160669