采用自动管电压选择的CT血管造影用于经导管主动脉瓣置换术规划:图像质量与辐射暴露
本研究评估第三代双源CT用于经导管主动脉瓣置换术(TAVR)规划CT血管造影(CTA)的图像质量与准确性,纳入125例患者,采用回顾性心电门控冠状动脉CTA和前瞻性心电触发主动脉髂动脉CTA两部分方案,应用自动管电压选择和高级迭代重建。结果显示图像质量良好至优秀者CCTA占97.6%、主动脉髂动脉CTA占100%;管电压>100kV较70-100kV客观图像质量下降且有效剂量增加;39例接受TAVR者瓣环直径均在推荐范围内,无严重心脏或血管并发症。
要点
- 第三代双源CT TAVR规划CTA中,CCTA图像质量良好至优秀者占97.6%,主动脉髂动脉CTA占100%。
- 管电压>100kV较70-100kV客观图像质量下降,且有效剂量随管电压从70kV升至>100kV持续增加。
- 39例接受TAVR者瓣环直径均在推荐范围内,无严重心脏或血管并发症。
核心结论
- 第三代双源CT用于TAVR规划CTA可提供诊断级图像质量,有助于可靠评估TAVR装置和输送选项,同时降低辐射剂量。
研究设计与样本
- 评估125例在第三代双源CT上行TAVR规划CTA的患者,采用两部分方案:回顾性心电门控冠状动脉CTA和前瞻性心电触发主动脉髂动脉CTA,使用60mL对比剂,应用自动管电压选择和高级迭代重建。
关键影像表现/征象
- CCTA图像质量主观评分为良好至优秀者占97.6%,主动脉髂动脉CTA占100%。
- 管电压>100kV的CTA研究较70-100kV客观图像质量下降(SNR所有p≤0.0459;CNR所有p≤0.0462)。
- 平均有效剂量从70kV至>100kV持续增加(CCTA:4.5±1.7mSv至13.6±2.9mSv,所有p≤0.0233;主动脉髂动脉CTA:2.4±0.9mSv至6.8±2.7mSv,所有p≤0.0414)。
诊断标准与关键数值
- 图像质量主观分析采用五分制量表,CCTA良好至优秀者占97.6%,主动脉髂动脉CTA占100%。
- 39例接受TAVR者瓣环直径均在推荐范围内。
临床应用与阅片要点
- TAVR规划CTA采用两部分方案:回顾性心电门控冠状动脉CTA和前瞻性心电触发主动脉髂动脉CTA,使用60mL对比剂。
- 应用自动管电压选择和高级迭代重建,管电压>100kV时客观图像质量下降且有效剂量增加。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: To assess image quality and accuracy of CT angiography (CTA) for transcatheter aortic valve replacement (TAVR) planning performed with 3rd generation dual-source CT (DSCT).
MATERIAL AND METHODS: We evaluated 125 patients who underwent TAVR-planning CTA on 3rd generation DSCT. A two-part protocol was performed including retrospectively ECG-gated coronary CTA (CCTA) and prospectively ECG-triggered aortoiliac CTA using 60mL of contrast medium. Automated tube voltage selection and advanced iterative reconstruction were applied. Effective dose (ED), signal-to-noise (SNR) and contrast-to-noise ratios (CNR) were calculated. Five-point scales were used for subjective image quality analysis. In patients who underwent TAVR, sizing parameters were obtained.
RESULTS: Image quality was rated good to excellent in 97.6% of CCTA and 100% of aortoiliac CTAs. CTA studies at >100kV showed decreased objective image quality compared to 70-100kV (SNR, all p≤0.0459; CNR, all p≤0.0462). Mean ED increased continuously from 70 to >100kV (CCTA: 4.5±1.7mSv-13.6±2.9mSv, all p≤0.0233; aortoiliac CTA: 2.4±0.9mSv-6.8±2.7mSv, all p≤0.0414). In 39 patients TAVR was performed and annulus diameter was within the recommended range in all patients. No severe cardiac or vascular complications were noted.
CONCLUSION: 3rd generation DSCT provides diagnostic image quality in TAVR-planning CTA and facilitates reliable assessment of TAVR device and delivery option while reducing radiation dose.