文献速递心血管胸部CT

迭代重建对冠状动脉钙化积分的影响——体模与临床研究

The Influence of Iterative Reconstruction on Coronary Artery Calcium Scoring-Phantom and Clinical Studies.
PubMed 文献 · Acad Radiol · 2016-11-29

本研究对30例连续患者的冠状动脉钙化扫描分别用滤波反投影(FBP)、混合迭代重建(HIR)和迭代模型重建(IMR)重建,比较图像噪声与钙化积分。IMR图像噪声最低,HIR和IMR的钙化积分分别较FBP降低4.2%和8.9%,三种重建间平均钙化积分无显著差异,IMR的观察者间一致性最佳。

要点

  • IMR图像噪声为9.3±1.3 HU,显著低于HIR的29.6±4.8 HU和FBP的48.0±7.9 HU(P<.01)。
  • HIR和IMR的钙化积分分别较FBP降低4.2%和8.9%,但三种重建间平均钙化积分无显著差异。
  • IMR可减少图像噪声和开花伪影,降低测得的钙化积分,并提高测量可重复性。

核心结论

  • 对于冠状动脉钙化积分,IMR可降低图像噪声和开花伪影,从而降低测得的钙化积分,并可减小测量变异性、获得稳定可重复的测量。

研究设计与样本

  • 对30例连续患者(18男12女,年龄70.1±12.2岁)的冠状动脉钙化扫描分别用FBP、HIR和IMR重建,由两名放射科医师独立使用半自动软件测量钙化积分并评估观察者间一致性。

关键影像表现/征象

  • FBP、HIR和IMR图像的平均噪声分别为48.0±7.9、29.6±4.8和9.3±1.3 HU,所有重建组合间差异显著(P<.01)。
  • HIR和IMR的钙化积分分别较FBP降低4.2%和8.9%。
  • 三种重建间平均钙化积分无显著差异。

诊断标准与关键数值

  • 观察者间相关性:FBP r²=0.96,HIR r²=0.99,IMR r²=0.99,均为优。
  • Bland-Altman一致性以IMR最佳,其次为HIR,FBP最差。

临床应用与阅片要点

  • IMR可减少图像噪声和开花伪影,使钙化积分测量更稳定可重复。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

RATIONALE AND OBJECTIVES: We compared the effect of iterative model reconstruction (IMR), filtered back projection (FBP), and hybrid iterative reconstruction (HIR) on coronary artery calcium (CAC) scoring.

MATERIALS AND METHODS: CAC scans of 30 consecutive patients (18 men and 12 women, age 70.1 ± 12.2 years) were reconstructed with FBP, HIR, and IMR, and the image noise was measured on all images. Two radiologists independently measured the CAC scores using semiautomated software, and interobserver agreement was evaluated. Statistical analysis included the Spearman correlation coefficient and Bland-Altman analysis.

RESULTS: The mean image noise on FBP, HIR, and IMR images was 48.0 ± 7.9, 29.6 ± 4.8, and 9.3 ± 1.3 Hounsfield units, respectively. The difference among all reconstruction combinations was significant (P < .01). The CAC score on HIR and IMR scans was 4.2% and 8.9% lower, respectively, than the CAC score on FBP images. There was no significant difference in the mean CAC score among the three reconstructions. The interobserver correlation was excellent for all three reconstructions (r2 = 0.96 FBP, 0.99 HIR, 0.99 IMR); the best Bland-Altman measure of agreement was with IMR, followed by HIR and FBP.

CONCLUSION: For CAC scoring, IMR can reduce the image noise and blooming artifacts, and consequently lowers the measured CAC score. IMR can lessen measurement variability and yield stable, reproducible measurements.

来源:Acad Radiol,查看原文。版权归原作者所有。

引用

  1. Oda S, Utsunomiya D, Nakaura T, et al. The Influence of Iterative Reconstruction on Coronary Artery Calcium Scoring-Phantom and Clinical Studies. Acad Radiol. 2016. PMID: 27913107. doi:10.1016/j.acra.2016.11.003

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。