快速千伏切换双能CT钙化抑制冠状动脉CT血管成像的诊断效能
研究纳入67例疑似或已知冠状动脉疾病患者,采用单探测器单源双能CT快速千伏切换(80与140 kVp)行冠状动脉CT血管成像,以侵入性冠状动脉造影为参考标准,评估239个钙化节段的钙化抑制物质密度图像与虚拟单能图像的诊断效能。物质密度图像对显著狭窄的阳性预测值显著高于虚拟单能图像(75% vs 56%,P<0.0001),诊断效能更优。
要点
- 钙化抑制物质密度图像对钙化节段显著狭窄的阳性预测值为75%,高于虚拟单能图像的56%。
- 物质密度图像与虚拟单能图像对显著狭窄的敏感度分别为88%和91%,特异度分别为88%和71%。
- 双能CT钙化抑制物质密度成像可减少钙化病灶的假阳性诊断。
核心结论
- 钙化抑制物质密度图像提高了钙化冠状动脉病灶的阳性预测值和诊断效能。
研究设计与样本
- 单探测器单源双能CT快速千伏切换(80与140 kVp)研究,纳入67例疑似或已知冠状动脉疾病患者,评估239个钙化节段,以侵入性冠状动脉造影为参考标准,比较物质密度图像与虚拟单能图像的诊断效能。
诊断标准与关键数值
- 物质密度图像检测显著狭窄的敏感度88%、特异度88%、阳性预测值75%、阴性预测值95%、准确率88%;虚拟单能图像分别为敏感度91%、特异度71%、阳性预测值56%、阴性预测值95%、准确率77%;阳性预测值物质密度图像显著高于虚拟单能图像(P<0.0001)。
临床应用与阅片要点
- 钙化抑制物质密度图像可减少钙化冠状动脉病灶的假阳性诊断,有助于提高钙化节段狭窄评估的阳性预测值。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: Multi-detector-row computed tomography angiography (MDCTA) plays an important role in the assessment of patients with suspected coronary artery disease. However, MDCTA tends to overestimate stenosis in calcified coronary artery lesions. The aim of our study was to evaluate the diagnostic performance of calcification-suppressed material density (MD) images produced by using a single-detector single-source dual-energy computed tomography (ssDECT).
METHODS: We enrolled 67 patients with suspected or known coronary artery disease who underwent ssDECT with rapid kilovolt-switching (80 and 140 kVp). Coronary artery stenosis was evaluated on the basis of MD images and virtual monochromatic (VM) images. The diagnostic performance of the two methods for detecting coronary artery disease was compared with that of invasive coronary angiography as a reference standard.
RESULTS: We evaluated 239 calcified segments. In all the segments, the sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV) and accuracy for detecting significant stenosis were respectively 88%, 88%, 75%, 95% and 88% for the MD images, 91%, 71%, 56%, 95% and 77% for the VM images. PPV was significantly higher on the MD images than on the VM images (P < 0.0001).
CONCLUSIONS: Calcification-suppressed MD images improved PPV and diagnostic performance for calcified coronary artery lesions.
KEY POINTS: • Computed tomography angiography tends to overestimate stenosis in calcified coronary artery. • Dual-energy CT enables us to suppress calcification of coronary artery lesions. • Calcification-suppressed material density imaging reduces false-positive diagnosis of calcified lesion.
引用
- Yunaga H, Ohta Y, Kaetsu Y, et al. Diagnostic performance of calcification-suppressed coronary CT angiography using rapid kilovolt-switching dual-energy CT. Eur Radiol. 2016. PMID: 27966040. doi:10.1007/s00330-016-4675-x
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。