文献速递胸部CT

迭代重建算法对肺癌筛查超低剂量CT中肺结节计算机辅助检测软件效果的影响

Effects of Iterative Reconstruction Algorithms on Computer-assisted Detection (CAD) Software for Lung Nodules in Ultra-low-dose CT for Lung Cancer Screening.
PubMed 文献 · Acad Radiol · 2016-12-13

研究纳入85名接受低剂量CT及超低剂量CT扫描的高危人群受试者,比较滤波反投影与三维自适应迭代剂量降低及基于模型的迭代算法对肺结节计算机辅助检测软件性能的影响。超低剂量CT采用迭代重建后计算机辅助检测敏感度高于滤波反投影,其中基于模型的迭代算法敏感度接近低剂量CT滤波反投影。

要点

  • 超低剂量CT采用迭代重建可改善肺结节计算机辅助检测软件敏感度。
  • 超低剂量CT基于模型的迭代重建敏感度接近低剂量CT滤波反投影。
  • 超低剂量CT滤波反投影下计算机辅助检测敏感度明显下降。

核心结论

  • 在超低剂量CT中,使用迭代重建算法可提高计算机辅助检测软件对肺结节的检测性能,其中基于模型的迭代重建方案性能几乎等同于低剂量CT滤波反投影。

研究设计与样本

  • 研究纳入85名接受低剂量CT和超低剂量CT扫描的高危人群受试者,低剂量CT以滤波反投影重建,超低剂量CT分别以滤波反投影、三维自适应迭代剂量降低和基于模型的迭代重建重建,并比较计算机辅助检测软件性能。

诊断标准与关键数值

  • 在每例3.0个假阳性条件下,计算机辅助检测软件对全部结节的敏感度:低剂量CT滤波反投影78.7%,超低剂量CT滤波反投影9.3%,超低剂量CT三维自适应迭代剂量降低69.4%,超低剂量CT基于模型的迭代重建77.8%;超低剂量CT三维自适应迭代剂量降低和基于模型的迭代重建敏感度显著高于超低剂量CT滤波反投影(P < .001)。

临床应用与阅片要点

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

RATIONALE AND OBJECTIVES: This study aimed to evaluate the effects of iterative reconstruction (IR) algorithms on computer-assisted detection (CAD) software for lung nodules in ultra-low-dose computed tomography (ULD-CT) for lung cancer screening.

MATERIALS AND METHODS: We selected 85 subjects who underwent both a low-dose CT (LD-CT) scan and an additional ULD-CT scan in our lung cancer screening program for high-risk populations. The LD-CT scans were reconstructed with filtered back projection (FBP; LD-FBP). The ULD-CT scans were reconstructed with FBP (ULD-FBP), adaptive iterative dose reduction 3D (AIDR 3D; ULD-AIDR 3D), and forward projected model-based IR solution (FIRST; ULD-FIRST). CAD software for lung nodules was applied to each image dataset, and the performance of the CAD software was compared among the different IR algorithms.

RESULTS: The mean volume CT dose indexes were 3.02 mGy (LD-CT) and 0.30 mGy (ULD-CT). For overall nodules, the sensitivities of CAD software at 3.0 false positives per case were 78.7% (LD-FBP), 9.3% (ULD-FBP), 69.4% (ULD-AIDR 3D), and 77.8% (ULD-FIRST). Statistical analysis showed that the sensitivities of ULD-AIDR 3D and ULD-FIRST were significantly higher than that of ULD-FBP (P < .001).

CONCLUSIONS: The performance of CAD software in ULD-CT was improved by using IR algorithms. In particular, the performance of CAD in ULD-FIRST was almost equivalent to that in LD-FBP.

来源:Acad Radiol,查看原文。版权归原作者所有。

引用

  1. Nomura Y, Higaki T, Fujita M, et al. Effects of Iterative Reconstruction Algorithms on Computer-assisted Detection (CAD) Software for Lung Nodules in Ultra-low-dose CT for Lung Cancer Screening. Acad Radiol. 2016. PMID: 27986507. doi:10.1016/j.acra.2016.09.023

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。