临床呼吸运动校正软件(重建、配准与平均-RRA)用于18F-FDG-PET-CT:体模验证、实际意义与患者评估
研究在体模和患者数据上验证PET-CT系统的呼吸运动校正算法(RRA),体模需4分钟获得稳定定量,SUVmax恢复在最小球体(≤2.1 ml)最明显;患者28个肺病灶中RRA与未门控SUVmax无差异(7.0±7.8 vs 6.9±7.8,p=0.23),但SUV恢复与病灶位移显著相关,位移>1.2 mm时定量改善真实。
要点
- 体模中RRA法需4分钟采集时间获得稳定定量,参考相位和相位数不影响定量。
- 患者RRA与未门控SUVmax无统计学差异(7.0±7.8 vs 6.9±7.8,p=0.23)。
- SUV恢复与病灶位移显著相关,位移>1.2 mm的病灶定量改善真实。
核心结论
- 体模结果显示RRA法有前景,对病灶定量有真实影响;患者研究显示未门控与RRA数据间SUV有升高趋势、体积有降低趋势,且RRA较未门控的定量恢复与病灶位移显著相关,提示RRA方法应保留用于小病灶或位移大于1.2 cm的淋巴结患者。
研究设计与样本
- 体模验证与患者评估研究,在Discovery GE710 PET-CT系统上采集3个含5个18F-FDG填充球体的体模(球体置于水或充满18F-FDG的Styrofoam®罐中以产生不同对比度和衰减环境),球体做2 cm纵向呼吸运动;比较呼吸门控(RRA)与未门控PET图像及无运动静态参考图像,并评估28个肺病灶的定量改善。
关键影像表现/征象
- 体模中RRA法定量在静态采集相似,而未门控图像不同。
- SUVmax恢复在最小球体(≤2.1 ml)最明显。
- 患者RRA与未门控采集的SUVmax与全身图像SUVmax有统计学差异(p=0.05),但RRA与未门控之间无差异(平均SUVmax 7.0±7.8 vs 6.9±7.8,p=0.23)。
- SUV恢复与病灶位移呈统计学显著相关,位移>1.2 mm的病灶SUV定量改善真实。
诊断标准与关键数值
- 体模中RRA法需4分钟获得稳定定量。
- RRA法定量不受参考相位和相位数影响。
- 患者中RRA与未门控SUVmax无统计学差异(p=0.23),但与全身图像SUVmax有统计学差异(p=0.05)。
- SUV恢复与病灶位移显著相关,位移>1.2 mm时SUV定量改善真实。
临床应用与阅片要点
- 体模结果提示RRA法对病灶定量有真实影响,但患者研究显示RRA与未门控SUVmax无统计学差异。
- 作者指出RRA方法应保留用于小病灶或位移大于1.2 cm的淋巴结患者。
局限性
- 患者研究显示未门控与RRA数据间SUV有升高趋势、体积有降低趋势。
- 作者结论指出RRA方法应保留用于小病灶或位移大于1.2 cm的淋巴结患者。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: On fluorine-18 fludeoxyglucose (18F-FDG) positron emission tomography (PET) CT of pulmonary or hepatic lesions, standard uptake value (SUV) is often underestimated due to patient breathing. The aim of this study is to validate, on phantom and patient data, a motion correction algorithm [reconstruct, register and averaged (RRA)] implemented on a PET-CT system.
METHODS: Three phantoms containing five spheres filled with 18F-FDG and suspended in a water or Styrofoam®18F-FDG-filled tank to create different contrasts and attenuation environment were acquired on a Discovery GE710. The spheres were animated with a 2-cm longitudinal respiratory-based movement. Respiratory-gated (RRA) and ungated PET images were compared with static reference images (without movement). The optimal acquisition time, number of phases and the best phase within the respiratory cycle were investigated. The impact of irregular motion was also investigated. Quantification impact was computed on each sphere. Quantification improvement on 28 lung lesions was also investigated.
RESULTS: Phantoms: 4 min was required to obtain a stable quantification with the RRA method. The reference phase and the number of phases used for RRA did not affect the quantification which was similar on static acquisitions but different on ungated images. The results showed that the maximum standard uptake value (SUVmax) restoration is majored for the smallest spheres (≤2.1 ml).
PATIENTS: SUVmax on RRA and ungated acquisitions were statistically different to the SUVmax on whole-body images (p = 0.05) but not different from each other (mean SUVmax: 7.0 ± 7.8 vs 6.9 ± 7.8, p = 0.23 on RRA and ungated images, respectively). We observed a statistically significant correlation between SUV restoration and lesion displacement, with a real SUV quantitation improvement for lesion with movement >1.2 mm.
CONCLUSION: According to the results obtained using phantoms, RRA method is promising, showing a real impact on the lesion quantification on phantom data. With regard to the patient study, our results showed a trend towards an increase in the SUVs and a decrease in the volume between the ungated and RRA data. We also noticed a statistically significant correlation between the quantitative restoration obtained with RRA compared with ungated data and lesion displacement, indicating that the RRA approach should be reserved to patients with small lesions or nodes moving with a displacement larger than 1.2 cm. Advances in knowledge: This article investigates the performances of motion correction software recently introduced in PET. The conclusion revealed that such respiratory motion correction approach shows a real impact on the lesion quantification but must be reserved to the patient for whom lesion displacement was confirmed and high enough to clearly impact lesion evaluation.