研究纳入302个甲状腺结节,在细针穿刺或手术前进行常规超声和虚拟触诊组织成像与定量(VTIQ)检查,以病理或细胞学结合随访为参考标准,评估VTIQ剪切波速度指标预测甲状腺恶性的诊断效能;结果显示VTIQ各指标均优于常规超声特征,以平均剪切波速度2.60 m/s为阈值时敏感度为84.6%、阴性预测值为94.3%,且操作者间和操作者内重复性优良。

要点

  • VTIQ各剪切波速度指标预测甲状腺恶性的效能均优于常规超声特征(均P < .05)。
  • 以平均剪切波速度2.60 m/s为阈值,鉴别良恶性结节的敏感度为84.6%、阴性预测值为94.3%。
  • 结节大于10 mm时SWSmean的AUC为0.891,高于10 mm及以下结节的0.725(均P < .05)。

核心结论

  • VTIQ是预测甲状腺恶性有用且可重复的工具。

研究设计与样本

  • 共302个甲状腺结节在细针穿刺或手术前接受常规超声和VTIQ检查,以组织病理学或细胞学结果结合随访为参考标准。

诊断标准与关键数值

  • 以平均剪切波速度2.60 m/s为阈值,VTIQ鉴别良恶性结节的敏感度为84.6%、阴性预测值为94.3%。

临床应用与阅片要点

  • VTIQ各剪切波速度指标(SWSmin、SWSmax、SWSmean)预测甲状腺恶性的效能均优于常规超声特征(均P < .05);结节大于10 mm时SWSmax(AUC 0.862 vs 0.717)、SWSmin(AUC 0.866 vs 0.717)、SWSmean(AUC 0.891 vs 0.725)的AUC均高于10 mm及以下结节(均P < .05)。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To investigate the diagnostic performance of a 2-dimensional shear wave elastographic technique (Virtual Touch tissue imaging and quantification [VTIQ]; Siemens Medical Solutions, Mountain View, CA) for predicting thyroid malignancy.

METHODS: A total of 302 thyroid nodules underwent conventional sonography and VTIQ before fine-needle aspiration examination or surgery. Compared with histopathologic or cytologic results in combination with follow-up, the diagnostic performance of various shear wave speed (SWS) indices (minimum [SWSmin ], maximum [SWSmax ], and mean [SWSmean ]) on VTIQ as well as conventional sonographic features for predicting thyroid malignancy was evaluated in all of the nodules.

RESULTS: Sixty-five malignant and 237 benign thyroid nodules were histopathologically or cytologically confirmed. All SWS indices on VTIQ were lower in benign nodules than thyroid malignancy (all P < .001). For discrimination between malignant and benign nodules, all VTIQ SWS indices were better than conventional sonographic features, such as a solid component, a taller-than-wide shape, microcalcification, a poorly defined margin and hypoechogenicity, in predicting thyroid malignancy (all P < .05). By applying a cutoff SWSmean value of 2.60 m/s, VTIQ achieved sensitivity and negative predictive values of 84.6% and 94.3%, respectively, for differentiating nodules. The areas under the receiver operating characteristic curve of SWSmax (0.862 versus 0.717), SWSmin (0.866 versus 0.717), and SWSmean (0.891 versus 0.725) for nodules larger than 10 mm were higher than those for nodules of 10 mm or smaller (all P < .05). Interoperator and intraoperator reproducibility was proven to be excellent, with all interclass correlation coefficient values higher than 0.80 (range, 0.813-0.905) CONCLUSIONS: Virtual Touch tissue imaging and quantification is a useful and reproducible tool for predicting thyroid malignancy.