Dynatron Solaris®超声仪:3 MHz、1.0 W/cm²条件下升温速度慢于教科书推荐
研究在30名健康志愿者内侧腓肠肌三个深度置入热电偶,用Dynatron Solaris 708超声仪以3 MHz、1.0 W/cm²连续治疗20分钟,测量组织升温速率。结果显示仅1.0 cm深度升温速率与教科书推荐相似,更深组织升温较慢,提示使用Omnisound教科书参数在Dynatron Solaris 708上可能无法达到目标组织温度。
要点
- Dynatron Solaris 708在3 MHz、1.0 W/cm²下,1.0 cm深度6分钟升温4.22°C(0.70°C/min),为唯一与教科书推荐相似的深度
- 1.75 cm深度10分钟升温3.93°C(0.39°C/min),2.5 cm深度20分钟升温3.60°C(0.18°C/min),均慢于教科书推荐
- 临床使用Omnisound教科书参数时,在深度大于1.0 cm处可能无法达到目标组织温度
核心结论
- Dynatron Solaris 708超声仪在3 MHz、1.0 W/cm²下各深度组织升温速率不同,仅1.0 cm深度与教科书推荐相似;使用Omnisound参数在深度大于1.0 cm时无法达到目标组织温度,可能影响临床医师、教育者和州认证考试。
研究设计与样本
- 3×13重复测量设计,30名健康志愿者(女11、男19,年龄21.30±1.95岁,脂肪厚度0.54±0.15 cm),在内侧腓肠肌1.0、1.75、2.5 cm深度置入热电偶,用Dynatron Solaris 708以3 MHz、1.0 W/cm²连续超声治疗20分钟。
诊断标准与关键数值
- 1.0 cm深度:6分钟升温4.22±1.58°C,速率0.70°C/min
- 1.75 cm深度:10分钟升温3.93±1.94°C,速率0.39°C/min
- 2.5 cm深度:20分钟升温3.60±1.86°C,速率0.18°C/min
- 深度主效应F2,52=29.76,P<0.001;时间主效应F12,312=181.59,P<.001;时间与深度交互F24,624=15.49,P<.001
临床应用与阅片要点
- 原文结论指出,临床医师使用Omnisound教科书参数在Dynatron Solaris 708上、深度大于1.0 cm时无法达到组织温度目标,可能影响临床医师、教育者和州认证考试。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
CONTEXT: Therapeutic ultrasound clinical parameters are provided in many modality textbooks based on research performed with the Omnisound brand. Literature exists to support variability in heating rates with different manufacturers. It is unknown if the Dynatron Solaris heats at rates consistent with textbook recommendations.
OBJECTIVE: Determine the rate of tissue-temperature increases in the medial triceps surae with the Dynatron Solaris® 708 ultrasound unit.
DESIGN: 3 × 13 repeated measures. Independent variables were tissue depth (1.0, 1.75, and 2.5 cm) and time (13 time periods throughout the treatment).
SETTING: Research laboratory.
PARTICIPANTS: 30 healthy volunteers (female = 11, male = 19; age 21.30 ± 1.95 y; adipose thickness = 0.54 ± 0.15 cm).
INTERVENTION: Three thermocouples were inserted into the medial triceps surae at 1.0, 1.75, and 2.5-cm depths. A continuous 3-MHz, 1.0-W/cm2 for 20 minutes ultrasound treatment was performed with a Dynatron Solaris 708 machine.
MAIN OUTCOME MEASURES: Intramuscular tissue-temperature increases at each depth throughout the 20-min treatment.
RESULTS: There was a significant main effect of depth (F2,52 = 29.76, P < 0.001) and time (F12,312 = 181.59, P < .001) and a significant interaction between times and depths (F24,624 = 15.49, P < .001). The 1.0-cm depth increased 4.22 ± 1.58°C in 6 min (0.70°C/min rate), the 1.75-cm depth increased 3.93 ± 1.94°C in 10 min (0.39°C/min rate), and the 2.5-cm depth increased 3.60 ± 1.86°C in 20 min (0.18°C/min rate).
CONCLUSIONS: The rate of tissue-temperature increase varied at each depth and the 1.0-cm depth was the only rate similar to textbooks. Clinicians will not reach tissue-temperature goals using Omnisound textbook parameters on the Dynatron Solaris 708 at depths greater than 1.0 cm, which may affect clinicians, educators, and state-certification exams.