正弦图确认迭代重建的CT图像质量体模研究——是否存在收益递减点?
该体模研究在ACR CT认证体模上扫描并重建FBP与SAFIRE(强度1、3、5),发现SAFIRE相对FBP的SNR与峰值SNR改善随CTDI降低而减小(CTDI>1.42 mGy时约70%,0.25 mGy时25%),CNR改善不随剂量变化,低对比与高对比分辨率无显著差异。
要点
- SAFIRE相对FBP的SNR改善随CTDI降低而减小,CTDI 1.42 mGy以上约70%,0.25 mGy时约25%。
- SAFIRE的CNR改善不随剂量变化,低对比与高对比分辨率与FBP无显著差异。
- CTDI低于某一阈值时,SAFIRE相对FBP的增量改善下降。
核心结论
- 在临床使用的CTDI范围内,应用SAFIRE后SNR改善并不均等,低于某一CTDI阈值时SAFIRE相对FBP的增量改善下降。
研究设计与样本
- 体模研究,获机构审查委员会豁免,使用ACR CT认证体模,以递增kV步级和有效管电流区间扫描,采集图像分别用FBP及SAFIRE强度1、3、5重建。
诊断标准与关键数值
- SAFIRE相对FBP的SNR与峰值SNR改善百分比从CTDI>1.42 mGy时的约70%降至CTDI 0.25 mGy时的25%。
- SAFIRE的CNR改善不随剂量变化。
- SAFIRE图像的低对比分辨率与高对比分辨率与FBP相比无显著差异。
临床应用与阅片要点
- 本研究中心的小儿实践中观察到,与FBP相比,SAFIRE生成的图像质量改善有限,尤其在CTDIvol接近或低于1 mGy时。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND: In our pediatric practice, we have observed qualitatively limited improvement in the image quality of images generated with sinogram affirmed iterative reconstruction (SAFIRE) compared to series generated with filtered back projection (FBP), particularly in cases near or below a CT dose index volume (CTDIvol) of 1-mGy.
OBJECTIVE: To determine whether the image quality advantage of SAFIRE remains constant across clinically used CT dose levels in an American College of Radiology (ACR) CT accreditation phantom including the lower dose range used in pediatric imaging.
MATERIALS AND METHODS: An exemption from institutional review board approval was obtained for this phantom-based study. An ACR quality phantom was scanned in incremental kV steps and effective tube current intervals. Acquisitions were reconstructed with FBP and SAFIRE strengths of 1, 3 and 5. Image quality measures were calculated including signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), low-contrast resolution and high-contrast resolution. Peak SNR was also calculated. Descriptive and nonparametric statistics were used to compare these image quality metrics while normalizing to CT dose index (CTDI).
RESULTS: The percent improvement in SNR and peak SNR of SAFIRE reconstructions compared to FBP decreased from about 70% for image sets acquired above a 1.42 mGy CTDI to 25% at a 0.25 mGy CTDI. CNR improvement with SAFIRE did not vary with dose. No significant difference was seen in the low-contrast resolution or high-contrast resolution of SAFIRE images compared to FBP.
CONCLUSION: SNR did not improve equally after applying SAFIRE across a spectrum clinically used CTDIs. Below a threshold CTDI, the incremental improvement of SAFIRE compared to FBP decreased.
引用
- Infante JC, Liu Y, Rigsby CK, et al. CT image quality in sinogram affirmed iterative reconstruction phantom study - is there a point of diminishing returns? Pediatr Radiol. 2016. PMID: 27891546. doi:10.1007/s00247-016-3745-1
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。