文献速递神经介入DSA

使用手术弹性蛋白酶诱导的复杂梭形动脉瘤模型测试支架与血流导向装置

Testing Stenting and Flow Diversion Using a Surgical Elastase-Induced Complex Fusiform Aneurysm Model.
PubMed 文献 · AJNR Am J Neuroradiol · 2016-11-24

研究在16只兔中尝试用手术弹性蛋白酶输注联合球囊扩张创建复杂梭形动脉瘤,另15只创建标准囊状颈动脉瘤,随机分配至血流导向装置、高孔隙率支架、双高孔隙率支架或对照,3个月行血管造影和病理分析;复杂梭形和标准囊状动脉瘤创建成功率分别为12/16和13/15,血流导向装置闭塞所有囊状(4/4)和复杂梭形(4/4)动脉瘤,双高孔隙率支架闭塞3/3囊状但0/2复杂梭形,单高孔隙率支架闭塞1/3囊状和0/3复杂梭形。

要点

  • 复杂梭形动脉瘤可由手术改良弹性蛋白酶法在兔中创建,成功率12/16。
  • 血流导向装置闭塞所有囊状(4/4)和复杂梭形(4/4)动脉瘤。
  • 双高孔隙率支架闭塞3/3囊状但0/2复杂梭形动脉瘤。

核心结论

  • 通过手术改良弹性蛋白酶法可在兔中创建具有挑战性的梭形动脉瘤。

研究设计与样本

  • 动物实验:16只兔尝试创建复杂梭形动脉瘤,15只创建标准囊状颈动脉瘤,随机分配至血流导向装置(n=8)、高孔隙率支架(n=6)、双高孔隙率支架(n=5)和对照(n=6),3个月行血管造影和病理分析。

关键影像表现/征象

  • 未治疗的两种动脉瘤病理表现相同:瘤壁缺乏弹性层和平滑肌细胞,管腔衬以厚度不一的新内膜。
  • 动脉瘤闭塞时装置的新内膜覆盖完整,而渗漏始终与动脉瘤残留相关。

诊断标准与关键数值

  • 复杂梭形动脉瘤创建成功12/16,标准囊状动脉瘤创建成功13/15。
  • 血流导向装置闭塞囊状动脉瘤4/4、复杂梭形动脉瘤4/4。
  • 双高孔隙率支架闭塞囊状动脉瘤3/3、复杂梭形动脉瘤0/2。
  • 单高孔隙率支架闭塞囊状动脉瘤1/3、复杂梭形动脉瘤0/3。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND AND PURPOSE: Rabbit elastase-induced saccular aneurysms have been commonly used for preclinical testing of endovascular devices, including flow diverters. However, all tested devices have been shown to be capable of aneurysm occlusion with this model. We aimed to create a more challenging model to test and discriminate among neurovascular devices of varying efficacies.

MATERIALS AND METHODS: With a surgical approach that included elastase infusion and balloon dilation, we attempted the creation of complex fusiform aneurysms in 16 rabbits, with standard saccular carotid aneurysms created in 15 other animals. Aneurysms were randomly allocated to one of the following treatments: flow diversion (n = 8), high-porosity stent (n = 6), double high-porosity stent (n = 5), and control (n = 6). Angiographic assessment and pathologic analyses were performed at 3 months.

RESULTS: Creation of complex fusiform and standard saccular aneurysms was successful in 12/16 and 13/15 attempts, respectively. All saccular (n = 4) or complex fusiform (n = 4) aneurysms treated with flow diverters were successfully occluded. Three of 3 saccular compared with 0/2 complex fusiform aneurysms were occluded by double high-porosity stents. One of 3 saccular and 0/3 complex fusiform aneurysms were occluded by a single high-porosity stent. Both aneurysm types shared the same pathologic findings when untreated: The aneurysm wall lacked an elastic layer and smooth muscle cells, while the lumen was lined with neointima of varying thickness. Neointimal coverage of the devices was complete when aneurysms were occluded, while leaks were always associated with aneurysm remnants.

CONCLUSIONS: Challenging fusiform aneurysms can be created in rabbits by using a surgical modification of the elastase method.

来源:AJNR Am J Neuroradiol,查看原文。版权归原作者所有。

引用

  1. Fahed R, Darsaut TE, Salazkin I, et al. Testing Stenting and Flow Diversion Using a Surgical Elastase-Induced Complex Fusiform Aneurysm Model. AJNR Am J Neuroradiol. 2016. PMID: 27884881. doi:10.3174/ajnr.A5018

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。