文献速递乳腺技术与安全乳腺X线

在全视野数字乳腺X线摄影、数字乳腺断层合成及两种数字X线乳腺摄影系统生成的人工合成二维图像中模拟微钙化检出率的比较

Comparison of the Detection Rate of Simulated Microcalcifications in Full-Field Digital Mammography, Digital Breast Tomosynthesis, and Synthetically Reconstructed 2-Dimensional Images Performed With 2 Different Digital X-ray Mammography Systems.
PubMed 文献 · Invest Radiol · 2017-01-01

本研究在体模模型中比较全视野数字乳腺X线摄影、人工合成二维图像与数字乳腺断层合成对模拟微钙化的检出率,使用两种数字X线乳腺摄影系统。结果显示数字乳腺断层合成对微钙化数量的正确评估率最高,全视野数字乳腺X线摄影优于人工合成二维图像,作者认为人工合成二维图像目前能否替代全视野数字乳腺X线摄影用于临床检查存疑,需进一步研究评估其临床影响。

要点

  • 数字乳腺断层合成对模拟微钙化数量的正确评估率为59.5%,高于全视野数字乳腺X线摄影的55.3%和人工合成二维图像的50.9%。
  • 全视野数字乳腺X线摄影对微钙化数量的正确评估率高于人工合成二维图像,平均差异4%(95%CI 2%-7%,P<0.001)。
  • 对100-399μm最小微钙化数量的正确评估率:全视野数字乳腺X线摄影25.2%、人工合成二维图像14.2%、数字乳腺断层合成28.3%。

核心结论

  • 在采用拟人化体模模型的体外环境中,标准全视野数字X线乳腺摄影对模拟微钙化的检出优于人工合成二维图像。
  • 作者认为,鉴于上述结果,人工合成二维图像目前能否在临床检查中替代全视野数字乳腺X线摄影存疑。
  • 需进一步研究评估上述体外结果的临床影响。

研究设计与样本

  • 研究类型为采用拟人化体模模型的体外比较研究。
  • 将不同数量(0至>39)、大小(直径100-800μm)和形状(圆形与不均匀)的模拟微钙化随机散布于50个胶片体模上,每个体模分为4个象限。
  • 使用两种数字X线乳腺摄影系统(SenoClaire,GE Healthcare;Selenia Dimensions,Hologic)配合拟人化散射体和自动曝光控制对上述50个胶片体模分别拍摄全视野数字乳腺X线摄影和数字乳腺断层合成图像,曝光条件与临床环境相似,人工合成二维图像在两套系统上自动生成。
  • 由6名放射科医师使用结构化问卷以随机顺序独立判读所有全视野数字乳腺X线摄影、人工合成二维图像和数字乳腺断层合成图像。

关键影像表现/征象

  • 模拟微钙化数量分类的正确评估率:全视野数字乳腺X线摄影55.3%、人工合成二维图像50.9%、数字乳腺断层合成59.5%(每位阅片者每台设备A和B各200个象限汇总)。
  • 全视野数字乳腺X线摄影对微钙化数量的正确评估率高于人工合成二维图像,平均差异4%(95%CI 2%-7%,P<0.001)。
  • 数字乳腺断层合成对微钙化数量的正确评估率高于全视野数字乳腺X线摄影(平均差异4%,95%CI 2%-7%,P=0.002)和人工合成二维图像(平均差异9%,95%CI 6%-11%,P<0.001),该趋势在所有亚组分析中一致。
  • 模拟微钙化数量被低估比被高估更常见。
  • 模拟微钙化大小分类的正确评估率:全视野数字乳腺X线摄影45.4%、人工合成二维图像39.9%、数字乳腺断层合成43.6%(每位阅片者每台成像设备各200个象限汇总)。

诊断标准与关键数值

  • 模拟微钙化数量分类正确评估率:全视野数字乳腺X线摄影55.3%、人工合成二维图像50.9%、数字乳腺断层合成59.5%。
  • 最小微钙化(100-399μm)数量正确评估率:全视野数字乳腺X线摄影25.2%、人工合成二维图像14.2%、数字乳腺断层合成28.3%。
  • 模拟微钙化大小分类正确评估率:全视野数字乳腺X线摄影45.4%、人工合成二维图像39.9%、数字乳腺断层合成43.6%。

临床应用与阅片要点

  • 作者认为,鉴于本体外结果,人工合成二维图像目前能否在临床检查中替代全视野数字乳腺X线摄影存疑。
  • 需进一步研究评估上述体外结果的临床影响。

局限性

  • 研究为采用拟人化体模模型的体外环境,作者指出需进一步研究评估该体外结果的临床影响。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: The aim of this study was to compare the microcalcification detectability in an anthropomorphic phantom model regarding number, size, and shape in full-field digital mammography (FFDM), synthetically reconstructed 2-dimensional (Synthetic-2D) images, and digital breast tomosynthesis (DBT) performed with 2 different x-ray mammography systems.

MATERIALS AND METHODS: Simulated microcalcifications of different numbers (0 to >39), sizes (diameter, 100-800 μm), and shapes (round vs heterogeneous) were scattered by random distribution on 50 film phantoms each divided in 4 quadrants. The FFDM and DBT x-rays were taken from each of these 50 films with both x-ray mammography systems (SenoClaire; GE Healthcare, Selenia Dimensions, Hologic) using an anthropomorphic scattering body and automatic exposure control. The resulting exposure factors were similar to a clinical setting. The synthetically reconstructed 2D images were generated automatically on both systems. All FFDM, Synthetic-2D, and DBT images were interpreted in randomized order and independently of each other by 6 radiologists using a structured questionnaire.

RESULTS: The number categories of simulated microcalcifications were correctly evaluated in 55.3% of instances (quadrant by reader) in FFDM, 50.9% in the Synthetic-2D views, and 59.5% in DBT, summarized for 200 quadrants per reader for each Device A and B, respectively. Full-field digital mammography was superior to Synthetic-2D (mean difference, 4%; 95% confidence interval [CI], 2%-7%; P < 0.001), and DBT was superior to both FFDM (mean difference, 4%; 95% CI, 2%-7%; P = 0.002) and Synthetic-2D (mean difference, 9%; 95% CI, 6%-11%; P < 0.001). This trend was consistent in all subgroup analyses. The number of the smallest microcalcifications (100-399 μm) was correctly evaluated in 25.2% of the FFDM, in 14.2% for Synthetic-2D, and in 28.3% of the DBT images. Underestimations of the number of simulated microcalcifications were more common than overestimations. Regarding the size categories of simulated microcalcifications, the rates of correct assessments were in 45.4% of instances in FFDM, 39.9% in the Synthetic-2D views, and 43.6% in DBT, summarized for 200 quadrants per reader and both imaging devices.

CONCLUSIONS: In the presented in vitro environment using an anthropomorphic phantom model, standard full-field digital x-ray mammography was superior to synthetically reconstructed 2-dimensional images in the detection of simulated microcalcifications. In view of these results, it is questionable whether Synthetic-2D images can replace FFDM in clinical examinations at the present time. Further investigations are needed to assess the clinical impact of the in vitro results.

来源:Invest Radiol,查看原文。版权归原作者所有。

引用

  1. Peters S, Hellmich M, Stork A, et al. Comparison of the Detection Rate of Simulated Microcalcifications in Full-Field Digital Mammography, Digital Breast Tomosynthesis, and Synthetically Reconstructed 2-Dimensional Images Performed With 2 Different Digital X-ray Mammography Systems. Invest Radiol. 2017. PMID: 27861206. doi:10.1097/RLI.0000000000000334

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。