新型高时空分辨率与标准动态对比增强乳腺MRI的图像质量比较
本研究纳入30例在标准动态对比增强乳腺MRI上显示局灶性强化(29例良性、1例恶性)的患者,于另一日接受新型DISCO研究性DCE乳腺MRI。结果显示DISCO与标准MRI图像质量评分差异均小于0.8分(10分制),两者均达诊断图像质量,且DISCO有效时间分辨率快6倍(27秒 vs 168秒),对比后时间点20个 vs 3个。
要点
- DISCO与标准DCE乳腺MRI图像质量评分差异均小于0.8分(10分制),均达诊断质量。
- DISCO有效时间分辨率27秒,较标准MRI的168秒快6倍,对比后时间点20个 vs 3个。
- 两者空间分辨率匹配,均为0.8×0.8×1.6 mm。
核心结论
- DISCO提供与标准DCE乳腺MRI相当的图像质量,同时有效时间分辨率快6倍且保持相同高空间分辨率。
研究设计与样本
- 患者自身对照研究,30例在标准MRI上显示局灶性强化(29例良性、1例恶性)的患者于另一日接受研究性DISCO DCE乳腺MRI;由3名放射科医师盲法评分图像质量;空间分辨率均为0.8×0.8×1.6 mm。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: Currently, dynamic contrast-enhanced (DCE) breast magnetic resonance imaging (MRI) prioritizes spatial resolution over temporal resolution given the limitations of acquisition techniques. The purpose of our intrapatient study was to assess the ability of a novel high spatial and high temporal resolution DCE breast MRI method to maintain image quality compared with the clinical standard-of-care (SOC) MRI.
MATERIALS AND METHODS: Thirty patients, each demonstrating a focal area of enhancement (29 benign, 1 cancer) on their SOC MRI, consented to undergo a research DCE breast MRI on a second date. For the research DCE MRI, a method (DIfferential Subsampling with Cartesian Ordering [DISCO]) using pseudorandom k-space sampling, view sharing reconstruction, 2-point Dixon fat-water separation, and parallel imaging was used to produce images with an effective temporal resolution 6 times faster than the SOC MRI (27 vs 168 seconds, respectively). Both the SOC and DISCO MRI scans were acquired with matching spatial resolutions of 0.8 × 0.8 × 1.6 mm. Image quality (distortion/artifacts, resolution, fat suppression, lesion conspicuity, perceived signal-to-noise ratio, and overall image quality) was scored by 3 radiologists in a blinded reader study.
RESULTS: Differences in image quality scores between the DISCO and SOC images were all less than 0.8 on a 10-point scale, and both methods were assessed as providing diagnostic image quality in all cases. DISCO images with the same high spatial resolution, but 6 times the effective temporal resolution as the SOC MRI scans, were produced, yielding 20 postcontrast time points with DISCO compared with 3 for the SOC MRI, over the same total time interval.
CONCLUSIONS: DISCO provided comparable image quality compared with the SOC MRI, while also providing 6 times faster effective temporal resolution and the same high spatial resolution.
引用
- Morrison CK, Henze Bancroft LC, DeMartini WB, et al. Novel High Spatiotemporal Resolution Versus Standard-of-Care Dynamic Contrast-Enhanced Breast MRI: Comparison of Image Quality. Invest Radiol. 2017. PMID: 27898602. doi:10.1097/RLI.0000000000000329